Bipolar resistivity profiling of 3D tissue culture.

We describe a new low-cost technique for continuous monitoring of the thickness of biofilms and tissue cultures, and we demonstrate the advantage of using electrodes of different dimensions to probe different depths of a sample. We have used electric impedance spectroscopy to monitor keratinocyte stem cells (YF29) growing on an array of Ti/Pt coplanar microelectrodes. The thickness of the sample was reconstructed by fitting the measurements to theoretical curves. We have developed an algorithm for the rapid calculation of the resistance through a multilayered sample. This algorithm is based on conformal mapping and the serial partial capacitance technique. The validity of the technique was tested by measuring the sedimentation rate of an alumina powder. Sample thicknesses between 10 and 80 microm could be measured with a resolution of a few microns using the device.

[1]  Hagen Thielecke,et al.  3D-biohybrid systems: applications in drug screening. , 2002, Trends in biotechnology.

[2]  B Wolf,et al.  Monitoring of cellular behaviour by impedance measurements on interdigitated electrode structures. , 1997, Biosensors & bioelectronics.

[3]  Brian H. Brown,et al.  Bipolar and tetrapolar transfer impedance measurements from volume conductor , 2000 .

[4]  V. Hanna,et al.  Extension of the application of conformal mapping techniques to coplanar lines with finite dimensions , 1980 .

[5]  P. Jacobs,et al.  Design optimisation of planar electrolytic conductivity sensors , 2006, Medical and Biological Engineering and Computing.

[6]  Ulrich Egert,et al.  Biological application of microelectrode arrays in drug discovery and basic research , 2003, Analytical and bioanalytical chemistry.

[7]  Omowunmi A. Sadik,et al.  Impedance Spectroscopy: A Powerful Tool for Rapid Biomolecular Screening and Cell Culture Monitoring , 2005 .

[8]  Jenny Zhu,et al.  Label-free, real-time monitoring of IgE-mediated mast cell activation on microelectronic cell sensor arrays. , 2004, Journal of immunological methods.

[9]  S. Gawad,et al.  Micromachined impedance spectroscopy flow cytometer for cell analysis and particle sizing. , 2001, Lab on a chip.

[10]  W. Kenan,et al.  Impedance Spectroscopy: Emphasizing Solid Materials and Systems , 1987 .

[11]  I. Giaever,et al.  Micromotion of mammalian cells measured electrically. , 1991, Proceedings of the National Academy of Sciences of the United States of America.

[12]  Y. Barrandon,et al.  Epithelial sodium channel in human epidermal keratinocytes: expression of its subunits and relation to sodium transport and differentiation. , 1999, Journal of cell science.

[13]  Arnaud Bertsch,et al.  Resistivity probing of multi-layered tissue phantoms using microelectrodes. , 2004, Physiological measurement.

[14]  H. Galla,et al.  Impedance analysis of epithelial and endothelial cell monolayers cultured on gold surfaces. , 1996, Journal of biochemical and biophysical methods.

[15]  R. Mansel,et al.  Antagonistic effect of NK4 on HGF/SF induced changes in the transendothelial resistance (TER) and paracellular permeability of human vascular endothelial cells , 2002, Journal of cellular physiology.

[16]  A. Apparao,et al.  Depth Of Investigation In Direct Current Methods , 1971 .

[17]  Roland Schinzinger,et al.  Conformal Mapping: Methods and Applications , 1991 .

[18]  D.M. Wilson,et al.  Sensor technologies for monitoring metabolic activity in single cells-part II: nonoptical methods and applications , 2004, IEEE Sensors Journal.

[19]  G. Ghione,et al.  Revisiting the partial-capacitance approach to the analysis of coplanar transmission lines on multilayered substrates , 2003 .

[20]  Mirko Lehmann,et al.  New insights into the nanometer-scaled cell-surface interspace by cell-sensor measurements. , 2005, Experimental cell research.

[21]  Philippe Renaud,et al.  Comment on "AC frequency characteristics of coplanar impedance sensors as design parameters" by Jongin Hong, Dae Sung Yoon, Sung Kwan Kim, Tae Song Kim, Sanghyo Kim, Eugene Y. Pak and Kwangsoo No, Lab Chip, 2005, 5, 270. , 2005, Lab on a chip.

[22]  D. Stenger,et al.  Development and Application of Cell-Based Biosensors , 1999, Annals of Biomedical Engineering.

[23]  D. Jans,et al.  Transepithelial capacitance decrease reveals closure of lateral interspace in A6 epithelia , 1999, Pflügers Archiv.

[24]  W. Hilberg,et al.  From Approximations to Exact Relations for Characteristic Impedances , 1969 .

[25]  P. Bergveld,et al.  Theoretical and experimental determination of cell constants of planar-interdigitated electrolyte conductivity sensors , 1995 .

[26]  J. W. Schultze,et al.  Optimization of passivation layers for corrosion protection of silicon-based microelectrode arrays , 2000 .

[27]  Yuji Murakami,et al.  Application of on-chip cell cultures for the detection of allergic response. , 2004, Biosensors & bioelectronics.

[28]  B. Kasemo,et al.  Simultaneous frequency and dissipation factor QCM measurements of biomolecular adsorption and cell adhesion. , 1997, Faraday discussions.

[29]  K. Urabe,et al.  Reciprocal regulation of permeability through a cultured keratinocyte sheet by IFN-γ and IL-4 , 2004 .

[30]  Stephanus Buettgenbach,et al.  Monolithic fabrication of wireless miniaturized quartz crystal microbalance (QCM-R) arrays and their application for biochemical sensors , 2003 .

[31]  Robert Plonsey,et al.  Electrode guarding in electrical impedance measurements of physiological systems—a critique , 2006, Medical and Biological Engineering and Computing.

[32]  B Wolf,et al.  Functional cellular assays with multiparametric silicon sensor chips. , 2003, Lab on a chip.

[33]  Ursula van Rienen,et al.  Choosing electrodes for deep brain stimulation experiments–electrochemical considerations , 2005, Journal of Neuroscience Methods.