Early metallosis-related failure after total knee replacement: a report of 15 cases.

Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.

[1]  P. Pynsent,et al.  The influence of head size and sex on the outcome of Birmingham hip resurfacing. , 2010, The Journal of bone and joint surgery. American volume.

[2]  P. Papagelopoulos,et al.  Wear debris pseudotumor following total knee arthroplasty: a case report , 2009, Journal of medical case reports.

[3]  B Ollivere,et al.  Early clinical failure of the Birmingham metal-on-metal hip resurfacing is associated with metallosis and soft-tissue necrosis. , 2009, The Journal of bone and joint surgery. British volume.

[4]  J. Vogt,et al.  LCS mobile-bearing total knee replacement. A 10-year's follow-up study. , 2009, Orthopaedics & traumatology, surgery & research : OTSR.

[5]  W. Walter,et al.  Osteolysis in third-generation alumina ceramic-on-ceramic hip bearings with severe impingement and titanium metallosis. , 2008, The Journal of arthroplasty.

[6]  R. Gejo,et al.  Severe bone defects and reduced mineralization caused by massive metallosis after total knee arthroplasty: histopathologic and bone morphometric findings , 2007, Modern rheumatology.

[7]  T. Fehring,et al.  Sterilization and wear-related failure in first- and second-generation press-fit condylar total knee arthroplasty. , 2007, Clinical orthopaedics and related research.

[8]  P. Papagelopoulos,et al.  Distal leg wear debris mass from a rotating hinged knee prosthesis. , 2007, The Journal of arthroplasty.

[9]  P. Korovessis,et al.  Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up. , 2006, The Journal of bone and joint surgery. American volume.

[10]  M. Pagnano,et al.  Assuring a painless total hip arthroplasty: a multimodal approach emphasizing peripheral nerve blocks. , 2006, The Journal of arthroplasty.

[11]  W. Bargar,et al.  Early aseptic loosening of a modern acetabular component secondary to a change in manufacturing. , 2004, The Journal of arthroplasty.

[12]  S. Mangaleshkar,et al.  Uncemented rotating-platform total knee arthroplasty: a 4-year to 12-year follow-up. , 2006, The Journal of arthroplasty.

[13]  D. D’Lima,et al.  Outcome of an Acetabular Design with Hydroxyapatite Coating on a Rough Substrate , 2005, Clinical orthopaedics and related research.

[14]  W. Capello,et al.  Arc-deposited Hydroxyapatite-coated Cups: Results at Four to Seven Years , 2005, Clinical orthopaedics and related research.

[15]  R. Oakeshott,et al.  Debridement of wear granulomata in revision total hip arthroplasty: use of a chondrotome shaver blade. , 2004, The Journal of arthroplasty.

[16]  D. W. Roberts,et al.  Factors Influencing Wear and Osteolysis in Press-Fit Condylar Modular Total Knee Replacements , 2004, Clinical orthopaedics and related research.

[17]  A. Greenwald,et al.  Uncemented rotating-platform total knee replacement: a five to twelve-year follow-up study. , 2004, The Journal of bone and joint surgery. American volume.

[18]  Gerard A Engh,et al.  Responsibilities when confronting a defective surgical implant. , 2003, Clinical orthopaedics and related research.

[19]  A. Disch,et al.  Massive metallosis after revision of a fractured ceramic head onto a metal head , 2003, Archives of Orthopaedic and Trauma Surgery.

[20]  F. Buechel Long-Term Followup After Mobile-Bearing Total Knee Replacement , 2002, Clinical orthopaedics and related research.

[21]  F. Buechel,et al.  Twenty-year evaluation of the New Jersey LCS Rotating Platform Knee Replacement. , 2002, The journal of knee surgery.

[22]  F. Buechel,et al.  Twenty-Year Evaluation of Meniscal Bearing and Rotating Platform Knee Replacements , 2001, Clinical orthopaedics and related research.

[23]  W R Walsh,et al.  Morphometric and mechanical evaluation of titanium implant integration: comparison of five surface structures. , 2000, Journal of biomedical materials research.

[24]  D. Howie,et al.  Loosening of matt and polished cemented femoral stems. , 1998, The Journal of bone and joint surgery. British volume.

[25]  J. Arredondo,et al.  Scintigraphic evaluation in total knee failure secondary to severe metallosis. , 1998, The Journal of arthroplasty.

[26]  J. Arredondo,et al.  Midterm results in metal-backed patella revisions for metallosis. , 1997, American journal of orthopedics.

[27]  M. Huo,et al.  Metallosis mimicking infection in a cemented total knee replacement. , 1997, Orthopedics.

[28]  K. Saum,et al.  Impact of gamma sterilization on clinical performance of polyethylene in the knee. , 1996, The Journal of arthroplasty.

[29]  K. Saum,et al.  The Otto Aufranc Award. Impact of gamma sterilization on clinical performance of polyethylene in the hip. , 1995, Clinical orthopaedics and related research.

[30]  R. Ling,et al.  The Use of a Collar and Precoating on Cemented Femoral Stems Is Unnecessary and Detrimental , 1992, Clinical orthopaedics and related research.

[31]  J. Probst,et al.  Metallosis after osteosynthesis. , 1970, Pathologia Europaea.