Projection imaging of photon beams using Čerenkov-excited fluorescence

Full 3D beam profiling and quality assurance (QA) of therapeutic megavoltage linear accelerator (LINAC) x-ray photon beams is not routinely performed due to the slow point-by-point measurement nature of conventional scanning ionization chamber systems. In this study we explore a novel optical-based dose imaging approach using a standard commercial camera, water tank, and fluorescent dye, which when excited by the Čerenkov emission induced by the radiation beam, allows 2D projection imaging in a fast timeframe, potentially leading toward 3D tomographic beam profiling. Detailed analysis was carried out to optimize the imaging parameters in the experimental setup. The results demonstrate that the captured images are linear with delivered dose, independent of dose rate, and comparison of experimentally captured images to a reference dose distribution for a 4 × 4 cm(2) 6 MV x-ray photon beam yielded results with improved accuracy over a previous study which used direct imaging and Monte Carlo calibration of the Čerenkov emission itself. The agreement with the reference dose distribution was within 1-2% in the lateral direction, and ±3% in the depth direction. The study was restricted to single 2D image projection, with the eventual goal of creating full 3D profiles after tomographic reconstruction from multiple projections. Given the increasingly complex advances in radiation therapy, and the increased emphasis on patient-specific treatment plans, further refinement of the technique could prove to be an important tool for fast and robust QA of x-ray photon LINAC beams.

[1]  Lihong V. Wang,et al.  Biomedical Optics: Principles and Imaging , 2007 .

[2]  S Webb,et al.  Radiation dosimetry using polymer gels: methods and applications. , 2000, The British journal of radiology.

[3]  I. Rosenberg,et al.  Radiation Oncology Physics: A Handbook for Teachers and Students , 2008, British Journal of Cancer.

[4]  L J Schreiner,et al.  Review of Fricke gel dosimeters , 2004 .

[5]  J C Gore,et al.  Radiation therapy dosimetry using magnetic resonance imaging of polymer gels. , 1996, Medical physics.

[6]  Louis Archambault,et al.  Transient noise characterization and filtration in CCD cameras exposed to stray radiation from a medical linear accelerator. , 2008, Medical physics.

[7]  H. Ross Measurement of .beta.-emitting nuclides using Cerenkov radiation , 1969 .

[8]  Scott C Davis,et al.  Time-gated Cherenkov emission spectroscopy from linear accelerator irradiation of tissue phantoms. , 2012, Optics letters.

[9]  Harald Paganetti,et al.  Relative biological effectiveness (RBE) values for proton beam therapy. , 2002, International journal of radiation oncology, biology, physics.

[10]  Michael Fenech,et al.  Mitigating the risk of radiation-induced cancers: limitations and paradigms in drug development , 2014, Journal of radiological protection : official journal of the Society for Radiological Protection.

[11]  N Suchowerska,et al.  Plastic scintillation dosimetry: comparison of three solutions for the Cerenkov challenge. , 2011, Physics in medicine and biology.

[12]  J. Janesick,et al.  Scientific Charge-Coupled Devices , 2001 .

[13]  F. H. Attix,et al.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements. , 1992, Physics in medicine and biology.

[14]  Louis Archambault,et al.  A new water-equivalent 2D plastic scintillation detectors array for the dosimetry of megavoltage energy photon beams in radiation therapy. , 2011, Medical physics.

[15]  D. F. Eaton International Union of Pure and Applied Chemistry Organic Chemistry Division Commission on Photochemistry. Reference materials for fluorescence measurement. , 1988, Journal of photochemistry and photobiology. B, Biology.

[16]  I. Tamm,et al.  Coherent visible radiation of fast electrons passing through matter , 1937 .

[17]  A Batalla,et al.  The DOSIMAP, a high spatial resolution tissue equivalent 2D dosimeter for LINAC QA and IMRT verification. , 2009, Medical physics.

[18]  Erin Jackson,et al.  Cerenkov Radiation Energy Transfer (CRET) Imaging: A Novel Method for Optical Imaging of PET Isotopes in Biological Systems , 2010, PloS one.

[19]  Brian W. Pogue,et al.  Projection imaging of photon beams by the Čerenkov effect. , 2012, Medical physics.

[20]  Johan Axelsson,et al.  Cerenkov emission induced by external beam radiation stimulates molecular fluorescence. , 2011, Medical physics.

[21]  R. Mohan,et al.  Exploration of the potential of liquid scintillators for real-time 3D dosimetry of intensity modulated proton beams. , 2009, Medical physics.

[22]  J. Battista,et al.  Optical CT reconstruction of 3D dose distributions using the ferrous-benzoic-xylenol (FBX) gel dosimeter. , 1998, Medical physics.

[23]  Roberto Manduchi,et al.  Bilateral filtering for gray and color images , 1998, Sixth International Conference on Computer Vision (IEEE Cat. No.98CH36271).

[24]  R. Mohan,et al.  Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector system. , 2012, Medical physics.

[25]  P. A. Čerenkov Visible radiation produced by electrons moving in a medium with velocities exceeding that of light , 1937 .

[26]  Louis Archambault,et al.  Technical note: removing the stem effect when performing Ir-192 HDR brachytherapy in vivo dosimetry using plastic scintillation detectors: a relevant and necessary step. , 2011, Medical physics.

[27]  Thomas R. Mackie,et al.  Cerenkov light generated in optical fibres and other light pipes irradiated by electron beams , 1992 .

[28]  J C Gore,et al.  Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeter , 1994, Physics in medicine and biology.

[29]  F. H. Attix,et al.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration. , 1992, Physics in medicine and biology.

[30]  Brian W. Pogue,et al.  Time-gated Cherenkov emission spectroscopy from linear accelerator irradiation of tissue phantoms , 2012 .

[31]  J. Dempsey,et al.  New water equivalent liquid scintillation solutions for 3D dosimetry. , 2000, Medical physics.

[32]  R. E. Kellogg,et al.  Radiationless Intermolecular Energy Transfer. III. Determination of Phosphorescence Efficiencies , 1964 .

[33]  A Batalla,et al.  The DosiMap, a new 2D scintillating dosimeter for IMRT quality assurance: characterization of two Cerenkov discrimination methods. , 2008, Medical physics.

[34]  R. Mohan,et al.  Liquid scintillator for 2D dosimetry for high-energy photon beams. , 2009, Medical physics.