An implantable microphone that can be utilized as part of a totally implantable hearing aid is designed and implemented. The proposed microphone is implanted in the center of the pinna, and designed to ensure the speech frequency range and the appropriate sensitivity. The characteristics of the proposed microphone are evaluated using a finite element analysis (FEA). The microphone is composed of a small electric condenser microphone, titanium case 6.2mm in diameter and 3mm high, and SUS316L vibrating membrane in contact with hypodermic tissue to maintain the sensitivity of the microphone. The microphone components are all made of biocompatible materials, then the assembled microphone is hermetically sealed using a polymer and ceramic. Experiments with the fabricated microphone confirm an operational bandwidth of up to 5kHz without any decline of sensitivity in 6mm of hypodermic tissue.