Vascular Dynamics of a Shape Memory Polymer Foam Aneurysm Treatment Technique

The vascular dynamics of a shape memory polymer foam aneurysm treatment technique are assessed through the simulated treatment of a generic basilar aneurysm using coupled fluid dynamics and heat transfer calculations. The shape memory polymer foam, which expands to fill the aneurysm when heated, is modeled at three discrete stages of the treatment procedure. To estimate an upper bound for the maximum amount of thermal damage due to foam heating, a steady velocity is specified through the basilar artery, corresponding to a minimum physiological flow velocity over a cardiac cycle. During expansion, the foam alters the flow patterns within the aneurysm by shielding the aneurysm dome from a confined jet that issues from the basilar artery. The time scales for thermal damage to the artery walls and surrounding blood flow are computed from the temperature field. The flow through the post-treatment bifurcation is comprised of two counter-rotating vortex tubes that are located beneath the aneurysm neck and extend downstream into the outlet arteries. Beneath the aneurysm neck, a marked increase in the wall shear stress is observed due to the close proximity of the counter-rotating vortex tubes to the artery wall.

[1]  D. L. Fry Acute Vascular Endothelial Changes Associated with Increased Blood Velocity Gradients , 1968, Circulation research.

[2]  Ashleyj . Welch,et al.  Optical-Thermal Response of Laser-Irradiated Tissue , 1995 .

[3]  F. Crea,et al.  In vitro effects of argon laser radiation on blood: quantitative and morphologic analysis. , 1985, Journal of the American College of Cardiology.

[4]  Giles R Cokelet,et al.  The rheology of human blood , 1963 .

[5]  C. Cognard,et al.  Intracranial berry aneurysms: angiographic and clinical results after endovascular treatment. , 1998, Radiology.

[6]  D. Kallmes,et al.  In vitro proliferation and adhesion of basic fibroblast growth factor-producing fibroblasts on platinum coils. , 1998, Radiology.

[7]  Sharon Thomsen,et al.  Rate Process Analysis of Thermal Damage , 1995 .

[8]  W. H. Austin,et al.  VISCOSITY OF NORMAL HUMAN BLOOD UNDER NORMOTHERMIC AND HYPOTHERMIC CONDITIONS. , 1964, Journal of applied physiology.

[9]  J. Barbee,et al.  The effect of temperature on the relative viscosity of human blood. , 1973, Biorheology.

[10]  Ward Small,et al.  Prototype Fabrication and Preliminary In Vitro Testing of a Shape Memory Endovascular Thrombectomy Device , 2007, IEEE Transactions on Biomedical Engineering.

[11]  D. Liepsch,et al.  Haemodynamic stress in lateral saccular aneurysms , 1987, Acta Neurochirurgica.

[12]  J. Humphrey,et al.  Elastodynamics and Arterial Wall Stress , 2002, Annals of Biomedical Engineering.

[13]  Serbinenko Fa Balloon catheterization and occlusion of major cerebral vessels , 1974 .

[14]  D. Kallmes,et al.  Platinum coil-mediated implantation of growth factor-secreting endovascular tissue grafts: an in vivo study. , 1998, Radiology.

[15]  B. Cantwell,et al.  Topology of fine-scale motions in turbulent channel flow , 1996, Journal of Fluid Mechanics.

[16]  P. Hantson,et al.  Prospective analysis of aneurysm treatment in a series of 103 consecutive patients when endovascular embolization is considered the first option. , 2000, Journal of neurosurgery.

[17]  C. Kleinstreuer,et al.  Fluid-structure interaction effects on sac-blood pressure and wall stress in a stented aneurysm. , 2005, Journal of biomechanical engineering.

[18]  Kristina J. Liu,et al.  Report of two cases , 1995 .

[19]  C. Mehringer,et al.  Therapeutic embolization for vascular trauma of the head and neck. , 1983, AJNR. American journal of neuroradiology.

[20]  C. Strother,et al.  Platinum coil coatings to increase thrombogenicity: a preliminary study in rabbits. , 1993, AJNR. American journal of neuroradiology.

[21]  Murat Aksoy,et al.  Time‐resolved 3D quantitative flow MRI of the major intracranial vessels: Initial experience and comparative evaluation at 1.5T and 3.0T in combination with parallel imaging , 2007, Magnetic resonance in medicine.

[22]  David I. Kaneshiro,et al.  Evolution of endovascular therapy for aneurysm treatment. Historical overview. , 2005, Neurosurgical focus.

[23]  F. Nicoud,et al.  Hemodynamic changes induced by stenting in elastic arteries , 2022 .

[24]  J A Pearce,et al.  Rate process model for arterial tissue thermal damage: Implications on vessel photocoagulation , 1994, Lasers in surgery and medicine.

[25]  Y. Fung,et al.  Biomechanics: Mechanical Properties of Living Tissues , 1981 .

[26]  Y. Dubief,et al.  On coherent-vortex identification in turbulence , 2000 .

[27]  Y Cinar,et al.  Blood viscosity and blood pressure: role of temperature and hyperglycemia. , 2001, American journal of hypertension.

[28]  H. Boecher-Schwarz,et al.  Ex vivo study of the physical effect of coils on pressure and flow dynamics in experimental aneurysms. , 2000, AJNR. American journal of neuroradiology.

[29]  J. Brock,et al.  Embolectomy in a rabbit acute arterial occlusion model using a novel electromechanical extraction device. , 2006, AJNR. American journal of neuroradiology.

[30]  M. S. Chong,et al.  Topology of flow patterns in vortex motions and turbulence , 1994 .

[31]  H Hinghofer-Szalkay,et al.  Volume and density changes of biological fluids with temperature. , 1985, Journal of applied physiology.

[32]  P Regli,et al.  Electromagnetic and heat transfer computations for non-ionizing radiation dosimetry. , 2000, Physics in medicine and biology.

[33]  T. Leipzig,et al.  Deflation of metrizamide-filled balloon used to occlude a carotid-cavernous fistula. Case report. , 1983, Journal of neurosurgery.

[34]  C. Pasqualucci,et al.  Laser surgery in enclosed spaces: A review , 1985, Lasers in surgery and medicine.

[35]  Robert H. Svenson,et al.  Heat-induced changes in optical properties of human whole blood in vitro , 1999, Photonics West - Biomedical Optics.

[36]  P. Canham,et al.  Directional Wall Strength in Saccular Brain Aneurysms from Polarized Light Microscopy , 2000, Annals of Biomedical Engineering.

[37]  S. Thomsen PATHOLOGIC ANALYSIS OF PHOTOTHERMAL AND PHOTOMECHANICAL EFFECTS OF LASER–TISSUE INTERACTIONS , 1991, Photochemistry and photobiology.

[38]  A J Welch,et al.  Pulsed laser-induced thermal damage in whole blood. , 2000, Journal of biomechanical engineering.

[39]  R. Willinsky Detachable coils to treat intracranial aneurysms. , 1999, CMAJ : Canadian Medical Association journal = journal de l'Association medicale canadienne.

[40]  S. Cowin,et al.  Biomechanics: Mechanical Properties of Living Tissues, 2nd ed. , 1994 .

[41]  D D Duncan,et al.  The effect of compliance on wall shear in casts of a human aortic bifurcation. , 1990, Journal of biomechanical engineering.

[42]  G. Bugliarello,et al.  Detailed Characteristics of the Flow of Blood In Vitro , 1963 .

[43]  L. Yahia,et al.  Cold hibernated elastic memory foams for endovascular interventions. , 2003, Biomaterials.

[44]  H. F. Bowman,et al.  Thermal Conductivity and Thermal Diffusivity of Biomaterials: A Simultaneous Measurement Technique , 1977 .

[45]  D. Holdsworth,et al.  An analysis of the geometry of saccular intracranial aneurysms. , 1999, AJNR. American journal of neuroradiology.

[46]  Barbara M. Johnston,et al.  Non-Newtonian blood flow in human right coronary arteries: steady state simulations. , 2004, Journal of biomechanics.

[47]  M. Horowitz,et al.  Does electrothrombosis occur immediately after embolization of an aneurysm with Guglielmi detachable coils? , 1997, AJNR. American journal of neuroradiology.

[48]  A. Fox,et al.  Treatment of 54 traumatic carotid-cavernous fistulas. , 1981, Journal of neurosurgery.

[49]  J. Dion,et al.  Endovascular treatment of intracranial aneurysms with Guglielmi Detachable Coils: emphasis on new techniques , 2000, Journal of Clinical Neuroscience.

[50]  R. Klucznik,et al.  Enlargement of basilar artery aneurysms following balloon occlusion--"water-hammer effect". Report of two cases. , 1991, Journal of neurosurgery.

[51]  H. Handa,et al.  Radiopaque solidifying liquids for releasable balloon technique: a technical note. , 1980, Surgical neurology.

[52]  M. Iwaki,et al.  Development of the biologically active Guglielmi detachable coil for the treatment of cerebral aneurysms. Part II: an experimental study in a swine aneurysm model. , 1999, AJNR. American journal of neuroradiology.

[53]  R. Langer,et al.  Biodegradable, Elastic Shape-Memory Polymers for Potential Biomedical Applications , 2002, Science.

[54]  M. Bronskill,et al.  Analysis of changes in MR properties of tissues after heat treatment , 1999, Magnetic resonance in medicine.

[55]  H. Handa,et al.  Balloon embolization of a giant aneurysm using a newly developed catheter. , 1979, Surgical neurology.

[56]  J. W. Valvano,et al.  Thermal conductivity and diffusivity of arterial wall and atherosclerotic plaque , 1987 .

[57]  S Chien,et al.  Comparative hemorheology--hematological implications of species differences in blood viscosity. , 1971, Biorheology.

[58]  Frank P. Incropera,et al.  Fundamentals of Heat and Mass Transfer , 1981 .

[59]  J Laudan,et al.  Three-Dimensional Pulsatile Flow Simulation before and after Endovascular Coil Embolization of a Terminal Cerebral Aneurysm , 2001, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[60]  D. Steinman,et al.  Simulation of non-Newtonian blood flow in an end-to-side anastomosis. , 1994, Biorheology.

[61]  D. Steinman,et al.  MESH RESOLUTION REQUIREMENTS FOR THE NUMERICAL SIMULATION OF FLOW THROUGH STENTED ANEURYSMS , 2002 .

[62]  Y. Cho,et al.  Effects of the non-Newtonian viscosity of blood on flows in a diseased arterial vessel. Part 1: Steady flows. , 1991, Biorheology.

[63]  R. Virmani,et al.  Thermal-perfusion balloon coronary angioplasty: in vivo evaluation. , 1993, American heart journal.

[64]  E. Merrill,et al.  Rheology of human blood, near and at zero flow. Effects of temperature and hematocrit level. , 1963, Biophysical journal.

[65]  D M Eckmann,et al.  Hematocrit, Volume Expander, Temperature, and Shear Rate Effects on Blood Viscosity , 2000, Anesthesia and analgesia.

[66]  U. Piomelli,et al.  Turbulent structures in accelerating boundary layers , 2000 .

[67]  Michel E. Mawad,et al.  Guglielmi detachable coil embolization of acute intracranial aneurysm , 1997 .

[68]  J. Dion,et al.  Use of three-dimensional Guglielmi detachable coils in the treatment of wide-necked cerebral aneurysms. , 2000, AJNR. American journal of neuroradiology.

[69]  F. Viñuela,et al.  Endovascular treatment of experimental aneurysms by use of a combination of liquid embolic agents and protective devices. , 2000, AJNR. American journal of neuroradiology.

[70]  Ward Small,et al.  Shape Memory Polymer Stent With Expandable Foam: A New Concept for Endovascular Embolization of Fusiform Aneurysms , 2007, IEEE Transactions on Biomedical Engineering.

[71]  Samuel D. Conte,et al.  Elementary Numerical Analysis , 1980 .

[72]  F. A. Serbinenko Balloon catheterization and occlusion of major cerebral vessels. , 1974, Journal of neurosurgery.

[73]  S. Tamatani,et al.  Histological interaction of cultured endothelial cells and endovascular embolic materials coated with extracellular matrix. , 1997, Journal of neurosurgery.

[74]  G. Duckwiler,et al.  Guglielmi detachable coil embolization of acute intracranial aneurysm: perioperative anatomical and clinical outcome in 403 patients. , 1997, Journal of neurosurgery.

[75]  G. Duckwiler,et al.  Endovascular occlusion of intracranial aneurysms with electrically detachable coils: correlation of aneurysm neck size and treatment results. , 1994, AJNR. American journal of neuroradiology.

[76]  K. Perktold,et al.  Computer simulation of local blood flow and vessel mechanics in a compliant carotid artery bifurcation model. , 1995, Journal of biomechanics.

[77]  A D Hughes,et al.  Blood flow and vessel mechanics in a physiologically realistic model of a human carotid arterial bifurcation. , 2000, Journal of biomechanics.

[78]  T J Flotte,et al.  Er:YAG laser ablation of tissue: Effect of pulse duration and tissue type on thermal damage , 1989, Lasers in surgery and medicine.

[79]  Wim Verkruysse,et al.  Changes in Optical Properties of Human Whole Blood in vitro Due to Slow Heating , 1997, Photochemistry and photobiology.

[80]  F. Viñuela,et al.  Matrix: new bio-absorbable polymeric coils for the treatment of intracranial aneurysms , 2002 .

[81]  H. Steiger,et al.  Hemodynamic stress in lateral saccular aneurysms. , 1987 .

[82]  M. Seebass,et al.  Impact of nonlinear heat transfer on temperature control in regional hyperthermia , 1999, IEEE Transactions on Biomedical Engineering.