Feasibility of free breathing Lung MRI for Radiotherapy using non-Cartesian k-space acquisition schemes.

OBJECTIVE To test a free-breathing MRI protocol for anatomical and functional assessment during lung cancer radiotherapy by assessing two non-Cartesian acquisition schemes based on T1 weighted 3D gradient recall echo sequence: (i) stack-of stars (StarVIBE) and (ii) spiral (SpiralVIBE) trajectories. METHODS MR images on five healthy volunteers were acquired on a wide bore 3T scanner (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany). Anatomical image quality was assessed on: (1) free breathing (StarVIBE), (2) the standard clinical sequence (volumetric interpolated breath-hold examination, VIBE) acquired in a 20 second (s) compliant breath-hold and (3) 20 s non-compliant breath-hold. For functional assessment, StarVIBE and the current standard breath-hold time-resolved angiography with stochastic trajectories (TWIST) sequence were run as multiphase acquisitions to replicate dynamic contrast enhancement (DCE) in one healthy volunteer. The potential application of the SpiralVIBE sequence for lung parenchymal imaging was assessed on one healthy volunteer. Ten patients with lung cancer were subsequently imaged with the StarVIBE and SpiralVIBE sequences for anatomical and structural assessment. For functional assessment, free-breathing StarVIBE DCE protocol was compared with breath-hold TWIST sequences on four prior lung cancer patients with similar tumour locations. Image quality was evaluated independently and blinded to sequence information by an experienced thoracic radiologist. RESULTS For anatomical assessment, the compliant breath-hold VIBE sequence was better than free-breathing StarVIBE. However, in the presence of a non-compliant breath-hold, StarVIBE was superior. For functional assessment, StarVIBE outperformed the standard sequence and was shown to provide robust DCE data in the presence of motion. The ultrashort echo of the SpiralVIBE sequence enabled visualisation of lung parenchyma. CONCLUSION The two non-Cartesian acquisition sequences, StarVIBE and SpiralVIBE, provide a free-breathing imaging protocol of the lung with sufficient image quality to permit anatomical, structural and functional assessment during radiotherapy. Advances in knowledge: Novel application of non-Cartesian MRI sequences for lung cancer imaging for radiotherapy. Illustration of SpiralVIBE UTE sequence as a promising sequence for lung structural imaging during lung radiotherapy.

[1]  J. Wong,et al.  The use of active breathing control (ABC) to reduce margin for breathing motion. , 1999, International journal of radiation oncology, biology, physics.

[2]  R. Munden,et al.  Effects of radiation therapy on the lung: radiologic appearances and differential diagnosis. , 2004, Radiographics : a review publication of the Radiological Society of North America, Inc.

[3]  Yongxian Qian,et al.  Acquisition‐weighted stack of spirals for fast high‐resolution three‐dimensional ultra‐short echo time MR imaging , 2008, Magnetic resonance in medicine.

[4]  Adam Johansson,et al.  CT substitute derived from MRI sequences with ultrashort echo time. , 2011, Medical physics.

[5]  The role of dynamic, contrast-enhanced MRI in differentiating lung tumor subtypes. , 2011, Clinical imaging.

[6]  M. Puderbach,et al.  MRI of the lung (2/3). Why … when … how? , 2012, Insights into Imaging.

[7]  M. Puderbach,et al.  MRI of the lung (3/3)—current applications and future perspectives , 2012, Insights into Imaging.

[8]  Rasmus Larsen,et al.  Three-dimensional MRI-linac intra-fraction guidance using multiple orthogonal cine-MRI planes , 2013, Physics in medicine and biology.

[9]  E. Merkle,et al.  Contrast-enhanced free-breathing 3D T1-weighted gradient-echo sequence for hepatobiliary MRI in patients with breath-holding difficulties , 2013, European Radiology.

[10]  Kevin M. Johnson,et al.  Optimized 3D ultrashort echo time pulmonary MRI , 2013, Magnetic resonance in medicine.

[11]  Hersh Chandarana,et al.  Free-breathing contrast-enhanced T1-weighted gradient-echo imaging with radial k-space sampling for paediatric abdominopelvic MRI , 2014, European Radiology.

[12]  David Atkinson,et al.  Respiratory motion correction in dynamic MRI using robust data decomposition registration - Application to DCE-MRI , 2014, Medical Image Anal..

[13]  Daniel K. Sodickson,et al.  Towards Routine Clinical Use of Radial Stack-of-Stars 3D Gradient-Echo Sequences for Reducing Motion Sensitivity , 2014 .

[14]  David Grodzki,et al.  Quiet Submillimeter MR Imaging of the Lung Is Feasible with a PETRA Sequence at 1.5 T. , 2015, Radiology.

[15]  H. Chandarana,et al.  Recent Advances in MR Hardware and Software. , 2015, Radiologic clinics of North America.

[16]  Esther G C Troost,et al.  Emerging Role of MRI for Radiation Treatment Planning in Lung Cancer , 2016, Technology in cancer research & treatment.

[17]  N. deSouza,et al.  Mechanism and non-mechanism based imaging biomarkers for assessing biological response to treatment in non-small cell lung cancer. , 2016, European journal of cancer.

[18]  D. Moses,et al.  Magnetic resonance imaging in lung: a review of its potential for radiotherapy. , 2016, The British journal of radiology.

[19]  Shivani Kumar,et al.  MRI in radiotherapy for lung cancer: A free-breathing protocol at 3T. , 2017, Practical radiation oncology.

[20]  T. Nyholm,et al.  A review of substitute CT generation for MRI-only radiation therapy , 2017, Radiation oncology.

[21]  Martin J. Graves,et al.  MRI from Picture to Proton , 2017 .