The Biomechanics of a Validated Finite Element Model of Stress Shielding in a Novel Hybrid Total Knee Replacement

This study proposes a novel hybrid total knee replacement (TKR) design to improve stress transfer to bone in the distal femur and, thereby, reduce stress shielding and consequent bone loss. Three-dimensional finite element (FE) models were developed for a standard and a hybrid TKR and validated experimentally. The Duracon knee system (Stryker Canada) was the standard TKR used for the FE models and for the experimental tests. The FE hybrid device was identical to the standard TKR, except that it had an interposing layer of carbon fibre-reinforced polyamide 12 lining the back of the metallic femoral component. A series of experimental surface strain measurements were then taken to validate the FE model of the standard TKR at 3000 N of axial compression and at 0° of knee flexion. Comparison of surface strain values from FE analysis with experiments demonstrated good agreement, yielding a high Pearson correlation coefficient of R2 = 0.94. Under a 3000 N axial load and knee flexion angles simulating full stance (0°), heel strike (20°), and toe off (60°) during normal walking gait, the FE model showed considerable changes in maximum Von Mises stress in the region most susceptible to stress shielding (i.e. the anterior region, just behind the flange of the femoral implant). Specifically, going from a standard to a hybrid TKR caused an increase in maximum stress of 87.4 per cent (0°; from 0.15 to 0.28 MPa), 68.3 per cent (20°; from 1.02 to 1.71 MPa), and 12.6 per cent (60°; from 2.96 to 3.33 MPa). This can potentially decrease stress shielding and subsequent bone loss and knee implant loosening. This is the first report to propose and biomechanically to assess a novel hybrid TKR design that uses a layer of carbon fibre-reinforced polyamide 12 to reduce stress shielding.

[1]  R. Zdero,et al.  Cortical screw purchase in synthetic and human femurs. , 2009, Journal of biomechanical engineering.

[2]  J. Simões,et al.  Relationship of design features of stemmed tibial knee prosthesis with stress shielding and end-of-stem pain , 2009 .

[3]  M. Hardisty,et al.  The biomechanics of the T2 femoral nailing system: A comparison of synthetic femurs with finite element analysis , 2009, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[4]  S. Kurtz,et al.  Younger Patient Demand for Primary and Revision Joint Replacement: National Projections from 2010 to 2030 , 2009 .

[5]  T. Goswami,et al.  Knee implants – Review of models and biomechanics , 2009 .

[6]  H. Pang,et al.  Does stress shielding occur with the use of long-stem prosthesis in total knee arthroplasty? , 2009, Knee Surgery, Sports Traumatology, Arthroscopy.

[7]  J. Ferreira,et al.  Corrosion aspects of metallic implants — An overview , 2008 .

[8]  R. Zdero,et al.  Cancellous bone screw purchase: A comparison of synthetic femurs, human femurs, and finite element analysis , 2008, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[9]  L. Yahia,et al.  CF/PA12 composite femoral stems: Manufacturing and properties , 2008 .

[10]  L. Yahia,et al.  Performance of CF/PA12 composite femoral stems , 2008, Journal of materials science. Materials in medicine.

[11]  D. Hernández-Vaquero,et al.  Influence of the tibial stem design on bone density after cemented total knee arthroplasty: a prospective seven-year follow-up study , 2008, International Orthopaedics.

[12]  J. Simões,et al.  Finite element and experimental cortex strains of the intact and implanted tibia. , 2007, Journal of biomechanical engineering.

[13]  C. Ranawat,et al.  Review Article: Osteolysis After Total Knee Arthroplasty , 2007 .

[14]  Aurelian Vadean,et al.  Design of a biomimetic polymer-composite hip prosthesis. , 2007, Journal of biomedical materials research. Part A.

[15]  R. Zdero,et al.  Cortical screw pullout strength and effective shear stress in synthetic third generation composite femurs. , 2007, Journal of biomechanical engineering.

[16]  R Zdero,et al.  The biomechanics of human femurs in axial and torsional loading: comparison of finite element analysis, human cadaveric femurs, and synthetic femurs. , 2007, Journal of biomechanical engineering.

[17]  J. Jurvelin,et al.  Periprosthetic femoral bone loss after total knee arthroplasty: 1-year follow-up study of 69 patients. , 2004, The Knee.

[18]  J. K. Spelt,et al.  Finite element analysis of a femoral retrograde intramedullary nail subject to gait loading. , 2004, Medical engineering & physics.

[19]  S. Stulberg Bone loss in revision total knee arthroplasty: graft options and adjuncts. , 2003, The Journal of arthroplasty.

[20]  M Beaugonin,et al.  Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. , 2002, Journal of biomechanics.

[21]  J. Liau,et al.  The effect of malalignment on stresses in polyethylene component of total knee prostheses--a finite element analysis. , 2002, Clinical biomechanics.

[22]  T D Brown,et al.  Structural properties of a new design of composite replicate femurs and tibias. , 2001, Journal of biomechanics.

[23]  N. Verdonschot,et al.  Distal femoral bone mineral density after total knee arthroplasty: a comparison with general bone mineral density , 2001, Archives of Orthopaedic and Trauma Surgery.

[24]  J. Liau,et al.  The influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics. , 2001, Clinical biomechanics.

[25]  J T Bryant,et al.  The effect of kinematic conditions on the wear of ultra-high molecular weight polyethylene (UHMWPE) in orthopaedic bearing applications , 2001, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[26]  A. Cappello,et al.  Mechanical validation of whole bone composite tibia models. , 2000, Journal of biomechanics.

[27]  N Verdonschot,et al.  Femoral bone loss in total knee arthroplasty. A review. , 1999, Acta orthopaedica Belgica.

[28]  A. M. Ahmed,et al.  Calculated stress‐shielding in the distal femur after total knee replacement corresponds to the reported location of bone loss , 1996, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[29]  L Cristofolini,et al.  Mechanical validation of whole bone composite femur models. , 1996, Journal of biomechanics.

[30]  J. Bryant,et al.  A quantitative technique for reporting surface degradation patterns of UHMWPE components of retrieved total knee replacements. , 1995, Journal of applied biomaterials : an official journal of the Society for Biomaterials.

[31]  J. Lauritzen,et al.  Changes in bone mineral density of the distal femur following uncemented total knee arthroplasty. , 1995, The Journal of arthroplasty.

[32]  G. Engh,et al.  Osteolysis of the distal femur after total knee arthroplasty. , 1994, The Journal of arthroplasty.

[33]  P. Jalovaara,et al.  Total hip arthroplasty using isoelastic femoral stems. A seven- to nine-year follow-up in 108 patients. , 1994, The Journal of bone and joint surgery. British volume.

[34]  M M Panjabi,et al.  Mechanical behavior of the human lumbar and lumbosacral spine as shown by three-dimensional load-displacement curves. , 1994, The Journal of bone and joint surgery. American volume.

[35]  H. Skinner Isoelasticity and total hip arthroplasty. , 1991, Orthopedics.

[36]  D. Robertson,et al.  Bone loss in the distal anterior femur after total knee arthroplasty. , 1990, Clinical orthopaedics and related research.

[37]  H. Ranu Basic Biomechanics of the Musculoskeletal System , 1989 .

[38]  D. Bartel,et al.  The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. , 1986, The Journal of bone and joint surgery. American volume.

[39]  R. Bombelli,et al.  Cementless Isoelastic RM Total HIP Prosthesis , 1982, Journal of the Royal Society of Medicine.

[40]  Z. Ráliš,et al.  Long-term effects of carbon fibre on soft tissues. , 1982, The Journal of bone and joint surgery. British volume.

[41]  J. Simões,et al.  Strain shielding in proximal tibia of stemmed knee prosthesis: experimental study. , 2008, Journal of biomechanics.

[42]  A. Amirfazli,et al.  Contribution of loading conditions and material properties to stress shielding near the tibial component of total knee replacements. , 2007, Journal of biomechanics.

[43]  C. Rorabeck,et al.  Wear and osteolysis around total knee arthroplasty. , 2007, The Journal of the American Academy of Orthopaedic Surgeons.

[44]  C. Ranawat,et al.  Osteolysis after total knee arthroplasty. , 2007, The Journal of arthroplasty.

[45]  D. Davy,et al.  Orthopaedic Biomechanics: Mechanics and Design in Musculoskeletal Systems , 2006 .

[46]  J. Lonner,et al.  Changes in bone density after cemented total knee arthroplasty: influence of stem design. , 2001, The Journal of arthroplasty.

[47]  P. Layrolle,et al.  Biomimetic Coatings on Orthopedic Implants: a Review , 1999 .

[48]  R. Huiskes,et al.  Stress shielding after total knee replacement may cause bone resorption in the distal femur. , 1997, The Journal of bone and joint surgery. British volume.

[49]  D. Robertson,et al.  Distal loss of femoral bone following total knee arthroplasty. Measurement with visual and computer-processing of roentgenograms and dual-energy x-ray absorptiometry. , 1994, The Journal of bone and joint surgery. American volume.

[50]  Michael Szycher,et al.  High performance biomaterials : a comprehensive guide to medical and pharmaceutical applications , 1992 .

[51]  E. Morscher,et al.  [Total isoelastic hip prosthesis implanted without cement. Initial results]. , 1974, Acta orthopaedica Belgica.