Performance comparison of two commercial BGO-based PET/CT scanners using NEMA NU 2-2001.

Combined positron emission tomography and computed tomography (PET/CT) scanners play a major role in medicine for in vivo imaging in an increasing number of diseases in oncology, cardiology, neurology, and psychiatry. With the advent of short-lived radioisotopes other than F18 and newer scanners, there is a need to optimize radioisotope activity and acquisition protocols, as well as to compare scanner performances on an objective basis. The Discovery-LS (D-LS) was among the first clinical PET/CT scanners to be developed and has been extensively characterized with older National Electrical Manufacturer Association (NEMA) NU 2-1994 standards. At the time of publication of the latest version of the standards (NU 2-2001) that have been adapted for whole-body imaging under clinical conditions, more recent models from the same manufacturer, i.e., Discovery-ST (D-ST) and Discovery-STE (D-STE), were commercially available. We report on the full characterization both in the two- and three-dimensional acquisition mode of the D-LS according to latest NEMA NU 2-2001 standards (spatial resolution, sensitivity, count rate performance, accuracy of count losses, and random coincidence correction and image quality), as well as a detailed comparison with the newer D-ST widely used and whose characteristics are already published.

[1]  Valentino Bettinardi,et al.  Performance evaluation of the new whole-body PET/CT scanner: Discovery ST , 2004, European Journal of Nuclear Medicine and Molecular Imaging.

[2]  Michael Lassmann,et al.  The impact of PET and SPECT on dosimetry for targeted radionuclide therapy. , 2006, Zeitschrift fur medizinische Physik.

[3]  Michael E Casey,et al.  PET performance measurements using the NEMA NU 2-2001 standard. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[4]  Roberta Matheoud,et al.  Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  Val J Lowe,et al.  NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[6]  Michel Defrise,et al.  Exact and approximate rebinning algorithms for 3-D PET data , 1997, IEEE Transactions on Medical Imaging.

[7]  Klaus Hahn,et al.  Performance evaluation of the PET sub-systems of two PET/CT imagers according to NEMA standard NU 2-2001. , 2006, Zeitschrift fur medizinische Physik.

[8]  O. Mawlawi,et al.  Performance characteristics of a newly developed PET/CT scanner using NEMA standards in 2D and 3D modes. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  T G Turkington,et al.  Performance characteristics of a whole-body PET scanner. , 1994, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.