SIC: an intracerebral radiosensitive probe for in vivo neuropharmacology investigations in small laboratory animals: first prototype design, characterization and in vivo evaluation

Several high-resolution tomographs dedicated to small animal imaging are presently under development. However, these devices suffer from two major drawbacks: their cost is high and require animal anaesthesia and immobilization thus restricting the field of investigations. Therefore, the authors have proposed SIC (French acronym for "Sonde IntraCerebrale") a low cost beta microprobe able to count the radioactivity concentration locally in awake, freely moving animals. Beside a good sensitivity, this tool offers the advantage of a high temporal resolution, a major asset for in vivo measurements of tracer kinetics. Furthermore, it could be easily coupled with existing techniques such as electrophysiology, microdialysis, voltametry and magnetic resonance imaging thus providing simultaneously complementary information on the cerebral mechanisms. The authors present here the design, manufacture and first evaluations of a first prototype. A first in vitro evaluation using a simple beaker filled with an homogenous aqueous solution of [/sup 18/F]-fluorodeoxyglucose (FDG) confirmed the authors' first simulation results in terms of sensitivity and detection volume. Then, further experimentation carried out in vivo on anaesthetized rats bearing two implanted probes in the brain have validated the use of SIC to measure locally, with a high temporal resolution, the concentration of radiolabeled molecules in living rodents.

[1]  D. Ingram,et al.  In vivo imaging of adenovirus‐mediated over‐expression of dopamine D2 receptors in rat striatum by positron emission tomography , 2000, Neuroreport.

[2]  Roger Lecomte,et al.  Initial results from the Sherbrooke avalanche photodiode positron tomograph , 1996 .

[3]  Sibylle Ziegler,et al.  A small animal PET prototype based on LSO crystals read out by avalanche photodiodes , 1997 .

[4]  M. Reivich,et al.  THE [14C]DEOXYGLUCOSE METHOD FOR THE MEASUREMENT OF LOCAL CEREBRAL GLUCOSE UTILIZATION: THEORY, PROCEDURE, AND NORMAL VALUES IN THE CONSCIOUS AND ANESTHETIZED ALBINO RAT 1 , 1977, Journal of neurochemistry.

[5]  S. Brown,et al.  Cerenkov radiation and its applications , 1955 .

[6]  A. P. Jeavons,et al.  A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals , 1998 .

[7]  L. Sokoloff,et al.  Effects of anesthesia on functional activation of cerebral blood flow and metabolism , 2001, Proceedings of the National Academy of Sciences of the United States of America.

[8]  A. Del Guerra,et al.  High spatial resolution small animal YAP-PET , 1998 .

[9]  Yves Charon,et al.  An original emission tomograph for in vivo brain imaging of small animals , 1996 .

[10]  Y. Blumenfeld,et al.  Must: a silicon strip detector array for radioactive beam experiments , 1999 .

[11]  D. Weber,et al.  Ultra-high-resolution imaging of small animals: Implications for preclinical and research studies , 1999, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[12]  Simon R. Cherry,et al.  In vivo imaging of neuronal activation and plasticity in the rat brain by high resolution positron emission tomography (microPET) , 2000, Nature Biotechnology.

[13]  S. Cherry,et al.  Performance evaluation of microPET: a high-resolution lutetium oxyorthosilicate PET scanner for animal imaging. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[14]  Yves Charon,et al.  SIC, an intracerebral radiosensitive probe for in vivo neuropharmacology investigations in small laboratory animals: theoretical considerations and practical characteristics , 2000 .

[15]  H. Atkins,et al.  Pinhole SPECT: an approach to in vivo high resolution SPECT imaging in small laboratory animals. , 1994, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[16]  Thomas R. Mackie,et al.  Cerenkov light generated in optical fibres and other light pipes irradiated by electron beams , 1992 .

[17]  Hiroyuki Okada,et al.  A high resolution animal PET scanner using compact PS-PMT detectors , 1996 .

[18]  J. Takeuchi,et al.  THE LATEST VACUUM PHOTODETECTOR , 1997 .