Low contrast imaging with a GaAs pixel digital detector

A digital mammography system based on GaAs pixel detector has been developed by the INFN (Istituto Nazionale of Fisica Nucleare) collaboration MED46. The high atomic number makes the GaAs a very efficient material for low energy X-rays detection (10-30 keV is the typical energy range used in mammography). Low contrast details can be detected with a significant dose reduction to the patient. The system presented in this paper consists of a 4096 pixel matrix built on a 200 /spl mu/m thick Semi Insulating GaAs substrate. The pixel size is 170/spl times/170 /spl mu/m/sup 2/ for a total active area of 1.18 cm/sup 2/. The detector is bump-bonded to a VLSI Front-End chip which implements a single-photon counting architecture. This feature allows to enhance the radiographic contrasts detection with respect to charge integrating devices. The system has been tested by using a standard mammographic tube. Images of mammographic phantoms will be presented and compared with radiographs obtained with traditional film/screen systems. Monte Carlo simulations have been also performed to evaluate the imaging capability of the system. Comparison with simulations and experimental results are shown.

[1]  Simone Stumbo,et al.  MEDIPIX: a VLSI chip for a GaAs pixel detector for digital radiology , 1999 .

[2]  Walter Snoeys,et al.  Performance of a 4096-pixel photon counting chip , 1998, Optics & Photonics.

[3]  J. Alverbro,et al.  X-ray imaging using a 320240 hybrid GaAs pixel detector , 1998 .

[4]  G. Meddeler,et al.  A Readout Chip for a 64 x 64 Pixel Matrix with 15-bit Single Photon Counting* , 1997 .

[5]  M. G. Bisogni,et al.  Experimental study of LEC GaAs detectors for X-ray digital radiography , 1996 .

[6]  Anna Cavallini,et al.  Performance of a new ohmic contact for GaAs particle detectors , 1995 .

[7]  G. A. Carlsson,et al.  Influence on X-ray energy spectrum, contrasting detail and detector on the signal-to-noise ratio (SNR) and detective quantum efficiency (DQE) in projection radiography , 1992 .

[8]  W. R. Nelson,et al.  Evaluation of elemental and compound semiconductors for x-ray digital radiography , 1991 .

[9]  S. Webb The Physics of Medical Imaging , 1990 .

[10]  D. Rogers,et al.  EGS4 code system , 1985 .

[11]  M. G. Bisogni,et al.  IMAGING PERFORMANCE OF A GAAS PIXEL DETECTOR , 1999 .

[12]  Amendolia,et al.  Charge collection properties of GaAs detectors for digital radiography , 1998 .

[13]  Simone Stumbo,et al.  A project for digital mammography based on a GaAs pixel detector and on a self-triggering single photon counting acquisition system , 1997 .

[14]  M Sandborg,et al.  Influence of X-ray energy spectrum, contrasting detail and detector on the signal-to-noise ratio (SNR) and detective quantum efficiency (DQE) in projection radiography , 1992 .

[15]  David W. O. Rogers,et al.  Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transport , 1986 .

[16]  U. G. Dailey Cancer,Facts and Figures about. , 2022, Journal of the National Medical Association.