Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments--a phantom approach.

This paper characterizes and evaluates the potential of three commercial CT iterative reconstruction methods (ASIR™, VEO™ and iDose⁴(™)) for dose reduction and image quality improvement. We measured CT number accuracy, standard deviation (SD), noise power spectrum (NPS) and modulation transfer function (MTF) metrics on Catphan phantom images while five human observers performed four-alternative forced-choice (4AFC) experiments to assess the detectability of low- and high-contrast objects embedded in two pediatric phantoms. Results show that 40% and 100% ASIR as well as iDose⁴ levels 3 and 6 do not affect CT number and strongly decrease image noise with relative SD constant in a large range of dose. However, while ASIR produces a shift of the NPS curve apex, less change is observed with iDose⁴ with respect to FBP methods. With second-generation iterative reconstruction VEO, physical metrics are even further improved: SD decreased to 70.4% at 0.5 mGy and spatial resolution improved to 37% (MTF(50%)). 4AFC experiments show that few improvements in detection task performance are obtained with ASIR and iDose⁴, whereas VEO makes excellent detections possible even at an ultra-low-dose (0.3 mGy), leading to a potential dose reduction of a factor 3 to 7 (67%-86%). In spite of its longer reconstruction time and the fact that clinical studies are still required to complete these results, VEO clearly confirms the tremendous potential of iterative reconstructions for dose reduction in CT and appears to be an important tool for patient follow-up, especially for pediatric patients where cumulative lifetime dose still remains high.

[1]  M. Gunn,et al.  State of the art: technologies for computed tomography dose reduction , 2010, Emergency Radiology.

[2]  Jean-Baptiste Thibault,et al.  TU‐A‐201B‐03: Dose Reduction and Image Quality Benefits Using Model Based Iterative Reconstruction (MBIR) Technique for Computed Tomography , 2010 .

[3]  Alvin C. Silva,et al.  Iterative Reconstruction Technique for Reducing Body Radiation Dose at Ct: Feasibility Study Hara Et Al. Ct Iterative Reconstruction Technique Gastrointestinal Imaging Original Research , 2022 .

[4]  P C Shrimpton,et al.  National survey of doses from CT in the UK: 2003. , 2006, The British journal of radiology.

[5]  A ROSE,et al.  Quantum effects in human vision. , 1957, Advances in biological and medical physics.

[6]  Dominik Fleischmann,et al.  Computed tomography—old ideas and new technology , 2011, European Radiology.

[7]  M. Galanski,et al.  Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey , 2003, European Radiology.

[8]  Willi A Kalender,et al.  Technical approaches to the optimisation of CT. , 2008, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[9]  C. McCollough,et al.  CT dose reduction and dose management tools: overview of available options. , 2006, Radiographics : a review publication of the Radiological Society of North America, Inc.

[10]  J. Remy,et al.  Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): evaluation of image noise reduction in 32 patients , 2011, European Radiology.

[11]  E. Samei,et al.  Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. , 2010, Radiology.

[12]  F R Verdun,et al.  EXPOSURE OF THE SWISS POPULATION BY RADIODIAGNOSTICS: 2003 REVIEW , 2007, Health physics.

[13]  J. Remy,et al.  Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients , 2011, European Radiology.

[14]  F. Gudinchet,et al.  Paediatric cardiac CT examinations: impact of the iterative reconstruction method ASIR on image quality – preliminary findings , 2011, Pediatric Radiology.

[15]  T. Flohr,et al.  Computed tomography—patient dose and dose reduction technologies. , 2011, Health physics.

[16]  Katsuyuki Taguchi,et al.  Combination of a Low-Tube-Voltage Technique With Hybrid Iterative Reconstruction (iDose) Algorithm at Coronary Computed Tomographic Angiography , 2011, Journal of computer assisted tomography.

[17]  François Gudinchet,et al.  Potential benefit of the CT adaptive statistical iterative reconstruction method for pediatric cardiac diagnosis , 2010, Medical Imaging.

[18]  Kai Yang,et al.  Noise power properties of a cone-beam CT system for breast cancer detection. , 2008, Medical physics.

[19]  Mythreyi Bhargavan,et al.  MEDICAL RADIATION EXPOSURE IN THE U.S. IN 2006: PRELIMINARY RESULTS , 2008, Health physics.

[20]  岩崎 民子 SOURCES AND EFFECTS OF IONIZING RADIATION : United Nations Scientific Committee on the Effects of Atomic Radiation UNSCEAR 2000 Report to the General Assembly, with Scientific Annexes , 2002 .

[21]  J H Siewerdsen,et al.  Soft-tissue detectability in cone-beam CT: evaluation by 2AFC tests in relation to physical performance metrics. , 2007, Medical physics.

[22]  C Ghetti,et al.  CT iterative reconstruction in image space: a phantom study. , 2012, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[23]  A E Burgess,et al.  The Rose model, revisited. , 1999, Journal of the Optical Society of America. A, Optics, image science, and vision.

[24]  Alvin C. Silva,et al.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. , 2010, AJR. American journal of roentgenology.