Determination of the optimal axial length of the electrode in an electrical capacitance tomography sensor

Abstract Electrical capacitance tomography (ECT) is one of the most successful process tomography techniques, which has been developed for more than a decade. Thus far, most of the research on ECT has been based on two-dimensional modeling, which simplifies the ECT sensor structure as two dimensional. However, certain key issues such as the three-dimensional effect of the ECT sensor are not considered at all. In this paper, the finite element method is used to solve the partial differential equation governing the ECT sensing zone, which is actually three dimensional. The sensitivity distributions between different electrode pairs are calculated and compared with those based on the two-dimensional modeling. The impacts of the length of both the measurement electrode and the guard electrode on the sensitivity distribution are investigated. The effect of the length of the measurement electrode on the normalized capacitance is also discussed. According to the numerical results presented in this paper, it is suggested that the optimal length of the measurement electrode for electrical capacitance tomography is the same as the width or diameter of the sensor.

[1]  Lihui Peng,et al.  Image reconstruction algorithms for electrical capacitance tomography , 2003 .

[2]  Wuqiang Yang,et al.  An image-reconstruction algorithm based on Landweber's iteration method for electrical-capacitance tomography , 1999 .

[3]  Richard A Williams,et al.  Process tomography: a European innovation and its applications , 1996 .

[4]  F T Kühn,et al.  Design of an active-differentiator-based capacitance transducer for electrical capacitance tomography , 1997 .

[5]  Brian S. Hoyle,et al.  Electrical capacitance tomography for flow imaging: system model for development of image reconstruction algorithms and design of primary sensors , 1992 .

[6]  Dieter Mewes,et al.  Recent developments and industrial/research applications of capacitance tomography , 1996 .

[7]  Wuqiang Yang,et al.  Hardware design of electrical capacitance tomography systems , 1996 .

[8]  Tomasz Dyakowski,et al.  Application of capacitance tomography to gas-solid flows , 1997 .

[9]  Tomasz Dyakowski,et al.  Process tomography - the state of the art , 1998 .

[10]  Yan Hua,et al.  Three-dimensional analysis of electrical capacitance tomography sensing fields , 1999 .

[11]  Wuqiang Yang,et al.  Principles and Industrial Applications of Electrical Capacitance Tomography , 1997 .

[12]  Ø. Isaksen,et al.  A review of reconstruction techniques for capacitance tomography , 1996 .

[13]  Yewang Su,et al.  Investigation of square fluidized beds using capacitance tomography: preliminary results , 2001 .

[14]  Roland Martin,et al.  Reconstruction of permittivity images from capacitance tomography data by using very fast simulated annealing , 2004 .

[15]  Maurice Beck,et al.  Tomographic imaging of two-component flow using capacitance sensors , 1989 .

[16]  M. S. Beck,et al.  Capacitance-based tomographic flow imaging system , 1988 .