Color Doppler Determination of Regurgitant Flow: From Proximal Isovelocity Surface Areas to Proximal Velocity Profiles: An In Vitro Study

There is not yet a completely satisfactory Doppler method for determining the severity of valvular regurgitation. Recently, interest has focused on the so‐called “flow convergence region”. Color Doppler provides a longitudinal velocity profile of the flow convergence region proximal to restrictive orifices. With respect to determination of regurgitant flow rate, we studied the influence of orifice flow rate and orifice size on the proximal velocity profile. In a phantom model, flow across circular orifices was studied. The distance r(v) of discrete velocities detected proximal to the orifice was measured along the flow center line. Velocity profiles were established. (1) Proximal isovelocity surfaces: The radius r(v) also represented the central radius of the proximal isovelocity surfaces. Increases in flow resulted in larger central radii r(v). Application of the continuity equation and analysis of the resulting values yielded a radius and an orifice size dependence of the geometric isovelocity surface shape. Therefore, their surface area and flow rate could not be calculated from these axial measurements alone. (2) Proximal velocity profiles: The resulting hyperbolic velocity profile curves (x axis = velocity [v], y axis = distance [r(v)]) are shifted rightward by increases in flow. In contrast, increases in orifice size make the curve steeper, but large r(v) are not affected. This differential influence of flow rate and orifice size allows orifice size independent determination of flow rate. A nomogram is presented as one possible method for flow determination.

[1]  A. Weyman,et al.  The proximal flow convergence method for calculating orifice flow rate requires correction for surrounding leaflet geometry , 1991 .

[2]  A. Yoganathan,et al.  Accuracy of flow convergence methods for calculating mitral regurgitant flow: Validation studies in an animal model , 1990 .

[3]  A F Bolger,et al.  Computer analysis of Doppler color flow mapping images for quantitative assessment of in vitro fluid jets. , 1988, Journal of the American College of Cardiology.

[4]  A. Weyman,et al.  Validation of a new method for valve area calculation using the proximal isovelocity surface area in patients with mitral stenosis , 1990 .

[5]  I. Simpson,et al.  Doppler color flow mapping of simulated in vitro regurgitant jets: evaluation of the effects of orifice size and hemodynamic variables. , 1989, Journal of the American College of Cardiology.

[6]  M. Stauch,et al.  Color Doppler PISA-method in vitro: Non-circular orifices lead to under- and multiple orifices to overestimation of flow , 1991 .

[7]  D Patel,et al.  Doppler color flow "proximal isovelocity surface area" method for estimating volume flow rate: effects of orifice shape and machine factors. , 1991, Journal of the American College of Cardiology.

[8]  A. Yoganathan,et al.  Spatial velocity distribution and acceleration in serial subvalve tunnel and valvular obstructions: an in vitro study using Doppler color flow mapping. , 1989, Journal of the American College of Cardiology.

[9]  D. Sahn,et al.  Effects of orifice size and shape on flow rate estimated from flow convergence region imaged by color Doppler flow mapping proximal to restrictive orifices: An in vitro study , 1990 .

[10]  A. Weyman,et al.  Nyquist limit and orifice area independently affect the accuracy of proximal isovelocity surface area estimation of flow rate:An in vitro study , 1990 .

[11]  D. Patel,et al.  Calculation of orifice area using color doppler proximal isovelocity surface area method , 1990 .