Testing of Anesthesia Machines and Defibrillators in Healthcare Institutions

To improve the quality of patient treatment by improving the functionality of medical devices in healthcare institutions. To present the results of the safety and performance inspection of patient-relevant output parameters of anesthesia machines and defibrillators defined by legal metrology. This study covered 130 anesthesia machines and 161 defibrillators used in public and private healthcare institutions, during a period of two years. Testing procedures were carried out according to international standards and legal metrology legislative procedures in Bosnia and Herzegovina. The results show that in 13.84% of tested anesthesia machine and 14.91% of defibrillators device performance is not in accordance with requirements and should either have its results be verified, or the device removed from use or scheduled for corrective maintenance. Research emphasizes importance of independent safety and performance inspections, and gives recommendations for the frequency of inspection based on measurements. Results offer implications for adequacy of preventive and corrective maintenance performed in healthcare institutions. Based on collected data, the first digital electronical database of anesthesia machines and defibrillators used in healthcare institutions in Bosnia and Herzegovina is created. This database is a useful tool for tracking each device’s performance over time.

[1]  Stephan Mieke,et al.  Metrology in Medicine , 2011 .

[2]  Zhao Wei,et al.  Energy calibration technique of defibrillator analyzer , 2015, 2015 12th IEEE International Conference on Electronic Measurement & Instruments (ICEMI).

[3]  C. J. Cassidy,et al.  Critical incident reports concerning anaesthetic equipment: analysis of the UK National Reporting and Learning System (NRLS) data from 2006–2008 * , 2011, Anaesthesia.

[4]  C. Ronsmans,et al.  Maternal mortality: who, when, where, and why , 2006, The Lancet.

[5]  Joseph F. Dyro,et al.  Clinical Engineering Handbook , 2012 .

[6]  M. Cecconi,et al.  Metrology in Medicine: From Measurements to Decision, with Specific Reference to Anesthesia and Intensive Care , 2014, Anesthesia and analgesia.

[7]  Binseng Wang,et al.  Medical equipment management strategies. , 2006, Biomedical instrumentation & technology.

[8]  Aaron Y. K. Yan,et al.  Calibration of Defibrillator Analyzers , 2015 .

[9]  H. B. Mann,et al.  On a Test of Whether one of Two Random Variables is Stochastically Larger than the Other , 1947 .

[10]  Ernesto Iadanza,et al.  Testing of mechanical ventilators and infant incubators in healthcare institutions. , 2017, Technology and health care : official journal of the European Society for Engineering and Medicine.

[11]  Janet B. Craig,et al.  A survey of anaesthetic misadventures , 1981, Anaesthesia.

[12]  A. Badnjević,et al.  Testing of dialysis machines in healthcare institutions in Bosnia and Herzegovina , 2017 .

[13]  Lejla Gurbeta,et al.  Development and perspectives of biomedical engineering in South East European countries , 2016, 2016 39th International Convention on Information and Communication Technology, Electronics and Microelectronics (MIPRO).

[14]  Dusanka Boskovic,et al.  Measurement in medicine – Past, present, future , 2015 .

[15]  M. do Céu Ferreira,et al.  The role of metrology in the field of medical devices , 2011 .

[16]  R. Campbell,et al.  Pressure-controlled versus volume-controlled ventilation: does it matter? , 2002, Respiratory care.

[17]  Mitchell D. Feldman,et al.  Who is Responsible for Evaluating the Safety and Effectiveness of Medical Devices? The Role of Independent Technology Assessment , 2007, Journal of General Internal Medicine.

[18]  M. Movahedi,et al.  A Case Study on Performance and Safety Test of Defibrillator Devices , 2016 .

[19]  Lejla Gurbeta,et al.  Inspection process of medical devices in healthcare institutions: software solution , 2017 .

[20]  K. Domino,et al.  Patient Injuries from Anesthesia Gas Delivery Equipment: A Closed Claims Update , 2013, Anesthesiology.

[21]  A. Badnjević,et al.  Testing of infusion pumps in healthcare institutions in Bosnia and Herzegovina , 2017 .

[22]  Dusanka Boskovic,et al.  Opportunities and challenges in biomedical engineering education: Focus on Bosnia and Herzegovina , 2015, 2015 4th Mediterranean Conference on Embedded Computing (MECO).

[23]  M. Movahedi,et al.  A Study on Performance and Safety Tests of Defibrillator Equipment , 2017, Journal of biomedical physics & engineering.

[24]  B Karaböce,et al.  Medical metrology studies at Tübitak UME , 2015 .

[25]  A. LAZO GUERRA Maternal mortality , 1951 .