Observations on the limitations of two apnea/respiration monitors commonly used in the home

The performance of two types of respiration monitor was studied with regard to detection of central apnea and individual breaths. Each monitor detected central apnea more frequently than routine intermittent nursing observation. Both false alarms, when the monitor failed to detect breathing movements (false positive for apnea), and false respiration signals, when the monitor failed to detect the absence of breathing movements (false negative for apnea) were found. The latter sometimes prevented alarms. It was found that a trial period of monitoring under the controlled conditions described permitted adjustments to be made which eliminated or minimized these sources of error. This should be considered when the monitors are issued for use in the home. It is important that persons issuing the monitors be aware of their limitations.

[1]  Y. Pirson,et al.  CYCLOSPORIN-DILTIAZEM INTERACTION , 1986, The Lancet.

[2]  M. Mackay,et al.  Home monitoring for central apnoea. , 1984, Archives of disease in childhood.

[3]  S. Proctor,et al.  Hypothermia during chemotherapy for Hodgkin's disease. , 1983, British medical journal.

[4]  E. P. Wright Zoography: the use of animal terms in medicine. , 1983, British medical journal.

[5]  M. H. Bryan,et al.  Home apnea monitoring in 'near-miss' sudden infant death syndrome (SIDS) and in siblings of SIDS victims. , 1982, Pediatrics.

[6]  S. Tonkin APNOSA AND UNEXPECTED CHILD DEATH , 1980, The Lancet.

[7]  D. Southall,et al.  24-hour tape recordings of ECG and respiration in the newborn infant with findings related to sudden death and unexplained brain damage in infancy , 1980, Archives of disease in childhood.

[8]  D. Southall,et al.  An explanation for failure of impedance apnoea alarm systems. , 1980, Archives of disease in childhood.

[9]  D. Shannon,et al.  Electrode belt: a new method for long-term monitoring. , 1979, Pediatrics.

[10]  D. Shannon,et al.  Care of infants with near-miss sudden infant death syndrome. , 1978, Pediatrics.

[11]  A. Stark,et al.  Apnea monitor failure in infants with upper airway obstruction. , 1977, Pediatrics.

[12]  A. Steinschneider,et al.  A reexamination of "the apnea monitor business". , 1976, Pediatrics.

[13]  P. Noone,et al.  Letter: Assay of gentamicin. , 1974, Lancet.

[14]  L. Victorin,et al.  Transthoracic Impedance: III. Methodological Studies in Newborn Infants , 1970 .

[15]  J. E. Lewin,et al.  AN APNŒA-ALARM MATTRESS , 1969 .

[16]  G. Grimby,et al.  Pulmonary Ventilation Measured from Body Surface Movements , 1967, Science.