Renal imaging at 7 Tesla: preliminary results

ObjectiveTo investigate the feasibility of 7T MR imaging of the kidneys utilising a custom-built 8-channel transmit/receive radiofrequency body coil.MethodsIn vivo unenhanced MR was performed in 8 healthy volunteers on a 7T whole-body MR system. After B0 shimming the following sequences were obtained: 1) 2D and 3D spoiled gradient-echo sequences (FLASH, VIBE), 2) T1-weighted 2D in and opposed phase 3) True-FISP imaging and 4) a T2-weighted turbo spin echo (TSE) sequence. Visual evaluation of the overall image quality was performed by two radiologists.ResultsRenal MRI at 7T was feasible in all eight subjects. Best image quality was found using T1-weighted gradient echo MRI, providing high anatomical details and excellent conspicuity of the non-enhanced vasculature. With successful shimming, B1 signal voids could be effectively reduced and/or shifted out of the region of interest in most sequence types. However, T2-weighted TSE imaging remained challenging and strongly impaired because of signal heterogeneities in three volunteers.ConclusionThe results demonstrate the feasibility and diagnostic potential of dedicated 7T renal imaging. Further optimisation of imaging sequences and dedicated RF coil concepts are expected to improve the acquisition quality and ultimately provide high clinical diagnostic value.

[1]  M. Schlamann,et al.  First Clinical Study on Ultra-High-Field MR Imaging in Patients with Multiple Sclerosis: Comparison of 1.5T and 7T , 2009, American Journal of Neuroradiology.

[2]  E. Merkle,et al.  A review of MR physics: 3T versus 1.5T. , 2007, Magnetic resonance imaging clinics of North America.

[3]  Diego R. Martín,et al.  MR colonography: 1.5T versus 3T. , 2007, Magnetic resonance imaging clinics of North America.

[4]  Niels Kuster,et al.  The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulations , 2010, Physics in medicine and biology.

[5]  Oliver Kraff,et al.  To TOF or not to TOF: strategies for non-contrast-enhanced intracranial MRA at 7 T , 2008, Magnetic Resonance Materials in Physics, Biology and Medicine.

[6]  R. Semelka,et al.  Liver MR imaging: 1.5T versus 3T. , 2007, Magnetic resonance imaging clinics of North America.

[7]  R. Lenkinski,et al.  Body MR imaging at 3.0 T: understanding the opportunities and challenges. , 2007, Radiographics : a review publication of the Radiological Society of North America, Inc.

[8]  Peter Andersen,et al.  Whole‐body imaging at 7T: Preliminary results , 2009, Magnetic resonance in medicine.

[9]  E. Merkle,et al.  MR cholangiopancreatography: 1.5T versus 3T. , 2007, Magnetic resonance imaging clinics of North America.

[10]  Jörg Stadler,et al.  MR imaging of the human hand and wrist at 7 T , 2009, Skeletal Radiology.

[11]  Ravinder R Regatte,et al.  Ultra‐high‐field MRI of the musculoskeletal system at 7.0T , 2007, Journal of magnetic resonance imaging : JMRI.

[12]  Matt A Bernstein,et al.  Imaging artifacts at 3.0T , 2006, Journal of magnetic resonance imaging : JMRI.

[13]  H H Quick,et al.  Evaluation of Intracranial Aneurysms with 7 T versus 1.5 T Time-of-Flight MR Angiography – Initial Experience , 2009, RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin.

[14]  C. Kuhl,et al.  Intra-individual comparison of image contrast in SPIO-enhanced liver MRI at 1.5T and 3.0T , 2007, European Radiology.

[15]  Sharmila Majumdar,et al.  Rapid in vivo musculoskeletal MR with parallel imaging at 7T , 2008, Magnetic resonance in medicine.

[16]  T. Bolz,et al.  An 8-channel add-on RF shimming system for whole-body 7 Tesla MRI including real-time SAR monitoring , 2008 .

[17]  R. Vikram,et al.  Imaging and staging of transitional cell carcinoma: part 2, upper urinary tract. , 2009, AJR. American journal of roentgenology.

[18]  Ying Lu,et al.  Detection of hepatic metastases from cancers of the gastrointestinal tract by using noninvasive imaging methods (US, CT, MR imaging, PET): a meta-analysis. , 2002, Radiology.

[19]  A. Aisen,et al.  Abdominal MR imaging at 3.0 T. , 2007, Radiographics : a review publication of the Radiological Society of North America, Inc.

[20]  M. E. Ladd,et al.  7 T Human in vivo Cardiac Imaging with an 8-Channel Transmit / Receive Array , 2008 .

[21]  G J Metzger,et al.  Initial results of cardiac imaging at 7 tesla , 2009, Magnetic resonance in medicine.

[22]  E. Merkle,et al.  Abdominal MR imaging at 3T. , 2006, Magnetic resonance imaging clinics of North America.

[23]  D. Tudorascu,et al.  Nephrogenic systemic fibrosis: Center case review , 2007, Journal of magnetic resonance imaging : JMRI.

[24]  R. Semelka,et al.  Quantitative and qualitative comparison of 1.5 and 3.0 tesla MRI in patients with chronic liver diseases , 2009, Journal of magnetic resonance imaging : JMRI.