On the inherent precision of mcDESPOT

A statistical analysis of the mcDESPOT protocol for characterizing two exchanging water proton pools—a seven‐dimensional problem that fits to multiple flip angle measurements of both spoiled and refocused gradient echoes—is presented. Theoretical calculations of the Cramér‐Rao lower bounds of the variance of fitted model parameters were made using a variety of model system parameters, meant to mimic those expected in human white matter. The results, validated by Monte Carlo simulations, indicated that mcDESPOT signals acquired at feasibly attainable signal‐to‐noise ratios cannot provide parameter estimates with useful levels of precision. Precision can be greatly improved by constraining solutions with a priori model information, although this will generally lead to biased parameter estimates with less specificity. These results indicate that previous, apparently successful applications of mcDESPOT to human white matter may have used data fitting methods that implicitly constrained parameter solutions, or that the two‐pool model of white matter may not be sufficient to describe the observed water proton signal in mcDESPOT acquisitions. In either case, mcDESPOT‐derived estimates of two‐pool model parameters cannot yet be unambiguously related to specific tissue characteristics. Magn Reson Med, 2013. © 2012 Wiley Periodicals, Inc.

[1]  H. Mcconnell Reaction Rates by Nuclear Magnetic Resonance , 1958 .

[2]  R. Henkelman Measurement of signal intensities in the presence of noise in MR images. , 1985, Medical physics.

[3]  Alex L. MacKay,et al.  Quantitative interpretation of NMR relaxation data , 1989 .

[4]  J. G. Gander,et al.  An introduction to signal detection and estimation , 1990 .

[5]  S. H. Koenig,et al.  Cholesterol of myelin is the determinant of gray‐white contrast in MRI of brain , 1991, Magnetic resonance in medicine.

[6]  C. S. Poon,et al.  Practical T2 quantitation for clinical applications , 1992, Journal of magnetic resonance imaging : JMRI.

[7]  P. Allen,et al.  Proton Relaxation Studies of Water Compartmentalization in a Model Neurological System , 1992, Magnetic resonance in medicine.

[8]  R M Henkelman,et al.  Relaxivity and magnetization transfer of white matter lipids at MR imaging: importance of cerebrosides and pH. , 1994, Radiology.

[9]  A. MacKay,et al.  In vivo visualization of myelin water in brain by magnetic resonance , 1994, Magnetic resonance in medicine.

[10]  H. Gudbjartsson,et al.  The rician distribution of noisy mri data , 1995, Magnetic resonance in medicine.

[11]  Gene H. Golub,et al.  Matrix computations (3rd ed.) , 1996 .

[12]  R E Snyder,et al.  Multiexponential T2 relaxation in degenerating peripheral nerve , 1996, Magnetic resonance in medicine.

[13]  A. Mackay,et al.  In vivo measurement of T2 distributions and water contents in normal human brain , 1997, Magnetic resonance in medicine.

[14]  R E Snyder,et al.  Multiecho imaging with suboptimal spoiler gradients. , 1998, Journal of magnetic resonance.

[15]  M. Bronskill,et al.  Characterizing white matter with magnetization transfer and T2 , 1999, Magnetic resonance in medicine.

[16]  J C Gore,et al.  Rapid acquisition transverse relaxometric imaging. , 2000, Journal of magnetic resonance.

[17]  V. Simplaceanu,et al.  Water compartments in the myelinated nerve. III. Pulsed NMR result , 1978, Experientia.

[18]  Fuzhen Zhang The Schur complement and its applications , 2005 .

[19]  T A Bjarnason,et al.  Characterization of the NMR behavior of white matter in bovine brain , 2005, Magnetic resonance in medicine.

[20]  K. Scheffler,et al.  On the origin of apparent low tissue signals in balanced SSFP , 2006, Magnetic resonance in medicine.

[21]  S. Nelson,et al.  Measurement of in vivo multi-component T2 relaxation times for brain tissue using multi-slice T2 prep at 1.5 and 3 T. , 2006, Magnetic resonance imaging.

[22]  A. MacKay,et al.  The effect of varying echo spacing within a multiecho acquisition: better characterization of long T2 components. , 2007, Magnetic resonance imaging.

[23]  K. Scheffler,et al.  Quantitative magnetization transfer imaging using balanced SSFP , 2008, Magnetic resonance in medicine.

[24]  D. Gochberg,et al.  MT effects and T1 quantification in single‐slice spoiled gradient echo imaging , 2008, Magnetic resonance in medicine.

[25]  Derek K. Jones,et al.  Gleaning multicomponent T1 and T2 information from steady‐state imaging data , 2008, Magnetic resonance in medicine.

[26]  D. Gochberg,et al.  Optimal echo spacing for multi-echo imaging measurements of Bi-exponential T2 relaxation. , 2009, Journal of magnetic resonance.

[27]  D. Gochberg,et al.  Multiexponential T2, magnetization transfer, and quantitative histology in white matter tracts of rat spinal cord , 2010, Magnetic resonance in medicine.

[28]  S. Deoni,et al.  Correction of main and transmit magnetic field (B0 and B1) inhomogeneity effects in multicomponent‐driven equilibrium single‐pulse observation of T1 and T2 , 2011, Magnetic resonance in medicine.

[29]  Oliver Bieri,et al.  Influence of MT effects on T2 quantification with 3D balanced steady‐state free precession imaging , 2011, Magnetic resonance in medicine.

[30]  Mark H. Johnson,et al.  Mapping Infant Brain Myelination with Magnetic Resonance Imaging , 2011, The Journal of Neuroscience.

[31]  Shannon H Kolind,et al.  Rapid three‐dimensional multicomponent relaxation imaging of the cervical spinal cord , 2011, Magnetic resonance in medicine.

[32]  K. D. Harkins,et al.  Effect of intercompartmental water exchange on the apparent myelin water fraction in multiexponential T2 measurements of rat spinal cord , 2012, Magnetic resonance in medicine.

[33]  Craig K. Jones,et al.  Applications of stimulated echo correction to multicomponent T2 analysis , 2012, Magnetic resonance in medicine.