Transesophageal echocardiography and intracardiac echocardiography differently predict potential technical challenges or failures of interatrial shunts catheter-based closure.

We sought to prospectively assess the role of transesophageal (TEE) and intracardiac echocardiography (ICE) in detecting potential technical difficulties or failures in patients submitted to interatrial shunts percutaneous closure. We prospectively enrolled 46 consecutive patients (mean age 35+/-28, 8 years, 30 female) referred to our center for catheter-based closure of interatrial shunts. All patients were screened with TEE before the intervention. Patients who met the inclusion criteria underwent ICE study before the closure attempt (40 patients). TEE detected potential technical difficulties in 22.5% (9/40) patients, whereas ICE detected technical difficulties in 32.5% (13/40 patients). In patients with positive TEE/ICE the procedural success (92.4% versus 100% and, P = ns) and follow-up failure rate (7.7% versus 0%, P = ns) were similar to patients with negative TEE/ICE, whereas the fluoroscopy time (7 +/- 1.2 versus 5 +/- 0.7 minutes, P < 0.03), the procedural time (41 +/- 4.1 versus 30 +/- 8.2 minutes, P +/- 0.03), and technical difficulties rate (23.1% versus 0%, P = 0.013) were higher. Differences between ICE and TEE in the evaluation of rims, measurement of ASD or fossa ovalis, and detection of venous valve and embryonic septal membrane remnants impacted on technical challenges and on procedural and fluoroscopy times but did not influence the success rate and follow-up failure rate.

[1]  Gianluca Rigatelli,et al.  Intracardiac echocardiography in cardiovascular catheter-based interventions: different devices for different purposes. , 2006, The Journal of invasive cardiology.

[2]  William Stewart,et al.  Recommendations for chamber quantification. , 2006, European journal of echocardiography : the journal of the Working Group on Echocardiography of the European Society of Cardiology.

[3]  R. Hein,et al.  Atrial and ventricular septal defects can safely be closed by percutaneous intervention. , 2005, Journal of interventional cardiology.

[4]  A. Angelini,et al.  Intracardiac echocardiography: gross anatomy and magnetic resonance correlations and validations , 2005, The International Journal of Cardiovascular Imaging.

[5]  E. Valsangiacomo-Buechel,et al.  Challenges Encountered During Closure of Atrial Septal Defects , 2005, Pediatric Cardiology.

[6]  V. Zadnik,et al.  Echocardiographic characteristics of secundum-type atrial septal defects in adult patients: implications for percutaneous closure using Amplatzer septal occluders. , 2004, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[7]  M. Zennaro,et al.  Intracardiac echocardiography-guided transcatheter closure of secundum atrial septal defect: a new efficient device selection method. , 2003, Journal of the American College of Cardiology.

[8]  M. Zennaro,et al.  Intracardiac Echocardiography Evaluation in Secundum Atrial Septal Defect Transcatheter Closure , 2003, CardioVascular and Interventional Radiology.

[9]  Q. Cao,et al.  Comparison of transcatheter closure of secundum atrial septal defect using the Amplatzer septal occluder associated with deficient versus sufficient rims. , 2002, The American journal of cardiology.

[10]  G. Rigatelli,et al.  Persistent redundant Eustachian valve interfering with Amplatzer PFO occluder placement: Anatomico‐clinical and technical implications , 2002, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[11]  Q. Cao,et al.  Transcatheter closure of atrial septal defects and patent foramen ovale under intracardiac echocardiographic guidance: Feasibility and comparison with transesophageal echocardiography , 2001, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[12]  T. Cheng Potential source of cerebral embolism in migraine with aura: a transcranial Doppler study , 1999 .

[13]  T. Meinertz,et al.  Chiari's network: normal anatomic variant or risk factor for arterial embolic events? , 1995, Journal of the American College of Cardiology.

[14]  C. Boutin,et al.  Echocardiographic Follow‐up of Atrial Septal Defect After Catheter Closure by Double‐Umbrella Device , 1993, Circulation.