MR imaging and biomedical implants, materials, and devices: an updated review.

Certain ferromagnetic metallic implants, materials, and devices are regarded as contraindications for magnetic resonance imaging, primarily because of the risks associated with their movement or dislodgment. More than 40 publications have reported the ferromagnetic qualities of 261 different metallic objects (aneurysm and hemostatic clips, 32; carotid artery vascular clamps, five; dental devices or materials, 16; heart valve prostheses, 29; intravascular coils, filters, and stents, 14; ocular implants, 12; orthopedic implants, materials, and devices, 15; otologic implants, 56; pellets and bullets, 23; penile implants, nine; vascular access ports, 33; and miscellaneous, 17) on the basis of measurements of deflection forces or attraction during exposure to static magnetic fields at strengths of 0.147-4.7 T. The results of these studies are listed with respect to the specific object tested, the material used to construct the object (if known), whether or not the object was deflected or moved during exposure to the static magnetic field, and the highest static magnetic field strength used for testing the object.

[1]  R. Lufkin,et al.  Magnetic resonance imaging artifacts: Mechanism and clinical significance , 1986 .

[2]  W. Bradley,et al.  Ferromagnetism and MR imaging: safety of carotid vascular clamps. , 1990, AJNR. American journal of neuroradiology.

[3]  F G Shellock,et al.  Biological effects and safety aspects of magnetic resonance imaging. , 1989, Magnetic resonance quarterly.

[4]  K. Go,et al.  Interaction of metallic neurosurgical implants with magnetic resonance imaging at 1.5 Tesla as a cause of image distortion and of hazardous movement of the implant , 1989, Clinical Neurology and Neurosurgery.

[5]  E M Haacke,et al.  MR artifacts: a review. , 1986, AJR. American journal of roentgenology.

[6]  A. Gomes,et al.  MR imaging of the bird's nest filter. , 1990, Radiology.

[7]  T F Budinger,et al.  Magnetic resonance imaging of prosthetic heart valves. , 1985, Radiology.

[8]  D Revel,et al.  Dental materials and magnetic resonance imaging. , 1991, Investigative radiology.

[9]  T. McCowan,et al.  Effect of a 1.5 Tesla Magnetic Field on Greenfield Filters in Vitro and in Dogs , 1988, Angiology.

[10]  C. Price,et al.  MR imaging artifacts produced by dental materials. , 1987, AJNR. American journal of neuroradiology.

[11]  W G Bradley,et al.  MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents, and coils. , 1988, Radiology.

[12]  M Dujovny,et al.  Aneurysm clip motion during magnetic resonance imaging: in vivo experimental study with metallurgical factor analysis. , 1985, Neurosurgery.

[13]  E Kanal,et al.  Safety considerations in MR imaging. , 1990, Radiology.

[14]  J. Breeze,et al.  Implants. , 2019, RDH.

[15]  E. Applebaum,et al.  Effects of magnetic resonance imaging fields on stapedectomy prostheses. , 1985, Archives of otolaryngology.

[16]  J. Gold,et al.  Safety of metallic surgical clips in patients undergoing high-field-strength magnetic resonance imaging. , 1989, Annals of Thoracic Surgery.

[17]  F. Shellock MR imaging of metallic implants and materials: a compilation of the literature. , 1988, AJR. American journal of roentgenology.

[18]  J. T. Thompson,et al.  The safety of magnetic resonance imaging with intraorbital metallic objects after retinal reattachment or trauma. , 1987, American journal of ophthalmology.

[19]  W S Hinshaw,et al.  Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. , 1983, Radiology.

[20]  G. Teitelbaum Metallic ballistic fragments: MR imaging safety and artifacts. , 1990, Radiology.

[21]  B. Haik,et al.  Magnetic resonance imaging of metal loop intraocular lenses. , 1987, American journal of ophthalmology.

[22]  F. Shellock,et al.  Policies, guidelines, and recommendations for MR imaging safety and patient management , 1991, Journal of magnetic resonance imaging : JMRI.

[23]  L. Kaufman,et al.  Potential hazards in NMR imaging: heating effects of changing magnetic fields and RF fields on small metallic implants. , 1981, AJR. American journal of roentgenology.

[24]  D. Levin,et al.  MR imaging of inferior vena caval filters: safety and artifacts. , 1988, AJR. American journal of roentgenology.

[25]  Technical artifacts in magnetic resonance imaging. , 1984, Physiological chemistry and physics and medical NMR.

[26]  F G Shellock,et al.  High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces. , 1988, AJR. American journal of roentgenology.

[27]  R McKinney,et al.  Magnetic field effects on surgical ligation clips. , 1987, Magnetic resonance imaging.

[28]  P. Joseph,et al.  Magnetic resonance imaging of postoperative patients with metallic implants. , 1984, AJR. American journal of roentgenology.

[29]  E P Strecker,et al.  Tantalum vascular stents: in vivo evaluation with MR imaging. , 1989, Radiology.

[30]  P. Paglen,et al.  Ferromagnetism of intraocular foreign body causes unilateral blindness after MR study. , 1986, AJNR. American journal of neuroradiology.

[31]  R. Henkelman,et al.  Magnetic resonance imaging and surgical clips. , 1984, Canadian journal of surgery. Journal canadien de chirurgie.

[32]  G. Teitelbaum,et al.  Metallic ballistic fragments: MR imaging safety and artifacts. , 1990, Radiology.

[33]  J. Tsuruda,et al.  Low-artifact intravascular devices: MR imaging evaluation. , 1988, Radiology.

[34]  M. Setzen,et al.  The Effect of Nuclear Magnetic Resonance Imaging on Metallic Middle Ear Prostheses , 1986, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[35]  H. Hricak,et al.  Intrauterine contraceptive devices: MR imaging. , 1987, Radiology.

[36]  R. Alfidi,et al.  MR imaging in patients with metallic implants. , 1985, Radiology.

[37]  W G Bradley,et al.  MR imaging in patients with intracranial aneurysm clips. , 1988, AJNR. American journal of neuroradiology.

[38]  E. Kanal,et al.  Magnetic ports in tissue expanders--a caution for MRI. , 1989, Magnetic resonance imaging.

[39]  R. Clark,et al.  Optimization of gradient-echo imaging parameters for intracaval filters and trapped thromboemboli. , 1990, Radiology.