Erbium: yttrium-aluminum-garnet laser stapedotomy—A safe technique

Objective To standardize the technical parameters of the erbium: yttrium-aluminum-garnet (Er:YAG) laser stapedotomy. Study Design Retrospective study of all patients with otosclerosis who underwent stapedotomy from January 2002 to January 2006. Subjects and Methods The charts of 152 consecutive patients who underwent stapedotomy were reviewed. The patients were stratified into two groups, according to the instrument used. Stapedotomies were performed in group A, with the OPMI®TwinEr: YAG laser; and in group B with manual microperforators. Results No statistically significant differences were found over all measured frequencies, between pre- and postoperative bone conduction thresholds, in each group. At the last postoperative follow-up, vertigo and nystagmus were not detected; two patients in group A and one patient in group B showed persistent tinnitus. Conclusion Er:YAG laser stapedotomy is a safe and effective procedure, with no damage of the inner ear when strict adherence to the safety parameters is observed. The Er:YAG laser is definitively suitable for stapes surgery, and represents a useful and safe tool in the armamentarium of otological microsurgery.

[1]  J. Dibartolomeo,et al.  The argon laser in otology , 1980, The Laryngoscope.

[2]  S. G. Lesinski Revision stapedectomy , 2003, Current opinion in otolaryngology & head and neck surgery.

[3]  J. Werner,et al.  Experimental and clinical results of Er:YAG laser stapedotomy , 2001, Lasers in surgery and medicine.

[4]  G. Paludetti,et al.  Erbium: Yttrium-Aluminum-Garnet Laser Application in Stapedotomy , 2005, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[5]  D. Anft,et al.  Experimental studies on the suitability of the erbium laser for stapedotomy in an animal model , 2004, European Archives of Oto-Rhino-Laryngology.

[6]  I. Siddiqui American Academy of Otolaryngology—Head and Neck Surgery , 1995 .

[7]  L Reinisch,et al.  Bone ablation with Er:YAG and CO2 laser: Study of thermal and acoustic effects , 1992, Lasers in surgery and medicine.

[8]  H P Weber,et al.  Advantages and dangers of erbium laser application in stapedotomy. , 1999, Acta oto-laryngologica.

[9]  D. Vernick A comparison of the results of KTP and CO2 laser stapedotomy. , 1996, American Journal of Otology.

[10]  G. Rettinger,et al.  Safety of the Erbium:Yttrium-Aluminum-Garnet Laser in Stapes Surgery in Otosclerosis , 2002, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[11]  C. Buchman,et al.  Comparison of argon and CO2 laser stapedotomy in primary otosclerosis surgery. , 2000, American journal of otolaryngology.

[12]  S. G. Lesinski,et al.  CO2 laser for otosclerosis: safe energy parameters. , 1989, The Laryngoscope.

[13]  Rodney C. Perkins,et al.  Laser stapedotomy for otosclerosis , 1980 .

[14]  T. Palva Argon laser in otosclerosis surgery. , 1987, Acta oto-laryngologica.

[15]  A Huber,et al.  Is the Er:YAG Laser Damaging to Inner Ear Function? , 2001, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[16]  H. Scherer,et al.  CO2 Laser Stapedotomy with the “One-Shot” Technique— Clinical Results , 2004, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[17]  R Hibst,et al.  Mechanical effects of erbium:YAG laser bone ablation , 1992, Lasers in surgery and medicine.

[18]  J. W. House,et al.  Committee on Hearing and Equilibrium Guidelines for the Evaluation of Results of Treatment of Conductive Hearing Loss ∗ , 1995, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[19]  G. Rettinger,et al.  Prospective Clinical Study on Cochlear Function after Erbium:Yttrium‐Aluminum‐Garnet Laser Stapedotomy , 2005, The Laryngoscope.

[20]  H P Weber,et al.  Temperature and pressure effects during erbium laser stapedotomy , 1996, Lasers in surgery and medicine.

[21]  T. Lenarz,et al.  Clinical Results After Stapedotomy: A Comparison Between the Erbium: Yttrium-Aluminum-Garnet Laser and the Conventional Technique , 2006, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.