Virtual scatter modulation for X-ray CT scatter correction using primary modulator.

A new scatter estimation algorithm with a concept of virtual scatter modulation for X-ray scatter correction using primary modulator is proposed to reduce the aliasing errors in the estimated scatter. Virtual scatter modulation can be realized through dividing the measured primary-modulated image by the measured modulation function. After the division, the aggravation of the aliasing of primary due to the non-uniformity of the modulation function is largely transferred to that of scatter. Since scatter in general has less high frequencies than primary does, the aggravation of its aliasing is expected to be weaker, and therefore the overall aliasing can be reduced. A CatPhan©600 phantom and an anthropomorphic thorax phantom are scanned on a tabletop X-ray cone-beam computed tomography system to validate our proposed algorithm. On the Catphan phantom, the oscillations that are clearly observed in the central region of the Catphan scatter profile estimated using the original primary-modulation algorithm, are mostly eliminated with the proposed scatter modulation algorithm, leading to less residual artifacts and better CT number uniformity in the reconstructed image. Compared with 38.9 HU of CT nonuniformity in a selected uniform region when the primary-modulation algorithm is used, the new algorithm significantly reduces it to 4.5 HU, reaching the same level of uniformity as the ground truth reference. On the thorax phantom, overall better CT number uniformity is also achieved.

[1]  J A Seibert,et al.  X-ray scatter removal by deconvolution. , 1988, Medical physics.

[2]  Lei Zhu,et al.  Modulator design for x-ray scatter correction using primary modulation: material selection. , 2010, Medical physics.

[3]  Rebecca Fahrig,et al.  Empirical cupping correction for CT scanners with primary modulation (ECCP) , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[4]  N. R. Bennett,et al.  Scatter correction method for x-ray CT using primary modulation: phantom studies. , 2010, Medical physics.

[5]  Lei Zhu,et al.  Evaluation of an erbium modulator in x-ray scatter correction using primary modulation , 2011, Medical Imaging.

[6]  Ruola Ning,et al.  X-ray scatter correction algorithm for cone beam CT imaging. , 2004, Medical physics.

[7]  Lei Zhu,et al.  Optimization of system parameters for modulator design in x-ray scatter correction using primary modulation , 2010, Medical Imaging.

[8]  J E Bowsher,et al.  Scatter Compensation for Digital Chest Radiography Using Maximum Likelihood Expectation Maximization , 1993, Investigative radiology.

[9]  J M Boone Scatter correction algorithm for digitally acquired radiographs: theory and results. , 1986, Medical physics.

[10]  Lei Zhu,et al.  Noise suppression in scatter correction for cone-beam CT. , 2009, Medical physics.

[11]  Lei Zhu,et al.  Scatter correction for x-ray conebeam CT using one-dimensional primary modulation , 2009, Medical Imaging.

[12]  Lei Zhu,et al.  Local filtration based scatter correction for cone-beam CT using primary modulation. , 2016, Medical physics.

[13]  John M Boone,et al.  Grid and slot scan scatter reduction in mammography: comparison by using Monte Carlo techniques. , 2002, Radiology.

[14]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[15]  W. Kalender,et al.  Combining deterministic and Monte Carlo calculations for fast estimation of scatter intensities in CT , 2006, Physics in medicine and biology.

[16]  Matthias Bertram,et al.  Model based scatter correction for cone-beam computed tomography , 2005, SPIE Medical Imaging.

[17]  Jingchen Ma,et al.  Optimization-based scatter estimation using primary modulation for computed tomography. , 2016, Medical physics.

[18]  Lei Zhu,et al.  Scatter Correction Method for X-Ray CT Using Primary Modulation: Theory and Preliminary Results , 2006, IEEE Transactions on Medical Imaging.

[19]  Rebecca Fahrig,et al.  Scatter correction using a primary modulator on a clinical angiography C‐arm CT system , 2017, Medical physics.

[20]  Josh Star-Lack,et al.  SU‐FF‐I‐04: A Fast Variable‐Intensity Ring Suppression Algorithm , 2006 .

[21]  S. Richard,et al.  A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT. , 2005, Medical physics.