Design consideration of a multipinhole collimator with septa for ultra high-resolution silicon drift detector modules

Abstract The aim of this study was to design a multipinhole (MP) collimator with lead vertical septa coupled to a high-resolution detector module containing silicon drift detectors (SDDs) with an intrinsic resolution approaching the sub-millimeter level. Monte Carlo simulations were performed to determine pinhole parameters such as pinhole diameter, focal length, and number of pinholes. Effects of parallax error and collimator penetration were investigated for the new MP collimator design. The MP detector module was evaluated using reconstructed images of resolution and mathematical cardiac torso (MCAT) phantoms. In addition, the reduced angular sampling effect was investigated over 180°. The images were reconstructed using dedicated maximum likelihood expectation maximization (MLEM) algorithm. An MP collimator with 81-pinhole was designed with a 2-mm-diameter pinhole and a focal length of 40 mm . Planar sensitivity and resolution obtained using the devised MP collimator were 3.9 cps/μCi and 6 mm full-width at half-maximum (FWHM) at a 10 cm distance. The parallax error and penetration ratio were significantly improved using the proposed MP collimation design. The simulation results demonstrated that the proposed MP detector provided enlarged imaging field of view (FOV) and improved the angular sampling effect in resolution and MCAT phantom studies. Moreover, the novel design enables tomography images by simultaneously obtaining eight projections with eight-detector modules located along the 180° orbit surrounding a patient, which allows designing of a stationary cardiac SPECT. In conclusion, the MP collimator with lead vertical septa was designed to have comparable system resolution and sensitivity to those of the low-energy high-resolution (LEHR) collimator per detector. The system sensitivity with an eight-detector configuration would be four times higher than that with a standard dual-detector cardiac SPECT.

[1]  F. Beekman,et al.  Photon penetration and scatter in micro-pinhole imaging: a Monte Carlo investigation. , 2004, Physics in medicine and biology.

[2]  Donald W. Wilson,et al.  SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. , 2006, Medical physics.

[3]  Frans van der Have,et al.  The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging , 2007, European Journal of Nuclear Medicine and Molecular Imaging.

[4]  William D. Erwin,et al.  Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras , 2006, Journal of applied clinical medical physics.

[5]  Donald W. Wilson,et al.  Application of silicon strip detectors to small-animal imaging , 2003 .

[6]  F. Beekman,et al.  Design and simulation of a high-resolution stationary SPECT system for small animals. , 2004, Physics in medicine and biology.

[7]  Donald W. Wilson,et al.  FastSPECT II: a second-generation high-resolution dynamic SPECT imager , 2002, IEEE Transactions on Nuclear Science.

[8]  Jules S. Jaffe Limited angle reconstruction using stabilized algorithms , 1990 .

[9]  L. Strüder,et al.  Monolithic arrays of silicon drift detectors for medical imaging applications and related CMOS readout electronics , 2006 .

[10]  M. B. Williams,et al.  Investigation of efficiency and spatial resolution using pinholes with small pinhole angle , 2003 .

[11]  P. Steele,et al.  Comparison of Simultaneous Dual-Isotope Multipinhole SPECT with Rotational SPECT in a Group of Patients with Coronary Artery Disease , 2008, Journal of Nuclear Medicine.

[12]  Hiroshi Watabe,et al.  Evaluation of penetration and scattering components in conventional pinhole SPECT: phantom studies using Monte Carlo simulation. , 2003, Physics in medicine and biology.

[13]  W. Segars,et al.  Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT phantom , 2001 .

[14]  N. Schramm,et al.  High-resolution SPECT using multi-pinhole collimation , 2002, IEEE Nuclear Science Symposium Conference Record.

[15]  J. Terry,et al.  Three-dimensional iterative reconstruction algorithms with attenuation and geometric point response correction , 1990 .

[16]  Eric Sonnendrücker,et al.  Pinhole SPECT imaging: compact projection/backprojection operator for efficient algebraic reconstruction , 2006, IEEE Transactions on Medical Imaging.

[17]  Characterization of two deep-diffusion avalanche photodiode array prototypes with different optical coatings , 2004, IEEE Transactions on Nuclear Science.

[18]  이경한,et al.  Optimization of a Silicon Drift Detector Module for a Gamma Camera , 2009 .

[20]  Stefan Eberl,et al.  A prototype coded aperture detector for small animal SPECT , 2001 .

[21]  Stephen C. Moore,et al.  Collimator design for single photon emission tomography , 2004, European Journal of Nuclear Medicine.

[22]  J. Kemmer,et al.  Large area silicon drift detectors for new applications in nuclear medicine imaging , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[23]  Yong Choi,et al.  Optimization of pinhole collimator for small animal SPECT using Monte Carlo simulation , 2002 .

[24]  L. Shepp,et al.  Maximum Likelihood Reconstruction for Emission Tomography , 1983, IEEE Transactions on Medical Imaging.

[25]  C Lartizien,et al.  GATE: a simulation toolkit for PET and SPECT. , 2004, Physics in medicine and biology.

[26]  J. F. Briesmeister MCNP-A General Monte Carlo N-Particle Transport Code , 1993 .

[27]  G. Gullberg,et al.  A rotating and warping projector/backprojector for fan-beam and cone-beam iterative algorithm , 1994 .

[28]  B. Hasegawa,et al.  A novel approach to multipinhole SPECT for myocardial perfusion imaging. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[29]  Freek J. Beekman,et al.  Acceleration of Monte Carlo SPECT simulation using convolution-based forced detection , 1999 .

[30]  Yong Choi,et al.  Performance amelioration for small animal SPECT using optimized pinhole collimator and image correction technique , 2005 .