Multi-frequency interpolation in spiral magnetic resonance fingerprinting for correction of off-resonance blurring.

Magnetic resonance fingerprinting (MRF) pulse sequences often employ spiral trajectories for data readout. Spiral k-space acquisitions are vulnerable to blurring in the spatial domain in the presence of static field off-resonance. This work describes a blurring correction algorithm for use in spiral MRF and demonstrates its effectiveness in phantom and in vivo experiments. Results show that image quality of T1 and T2 parametric maps is improved by application of this correction. This MRF correction has negligible effect on the concordance correlation coefficient and improves coefficient of variation in regions of off-resonance relative to uncorrected measurements.

[1]  N. Otsu A threshold selection method from gray level histograms , 1979 .

[2]  Vikas Gulani,et al.  Development of a Combined MR Fingerprinting and Diffusion Examination for Prostate Cancer. , 2017, Radiology.

[3]  J. Duerk,et al.  Magnetic Resonance Fingerprinting , 2013, Nature.

[4]  Nicholas R Zwart,et al.  Efficient sample density estimation by combining gridding and an optimized kernel , 2012, Magnetic resonance in medicine.

[5]  M. Griswold,et al.  Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom , 2017, Magnetic resonance in medicine.

[6]  Hermann Schomberg,et al.  Off-resonance correction of MR images , 1999, IEEE Transactions on Medical Imaging.

[7]  D G Nishimura,et al.  Efficient off‐resonance correction for spiral imaging , 2001, Magnetic resonance in medicine.

[8]  J A Frank,et al.  Simple correction method for k-space trajectory deviations in MRI. , 1998, Journal of magnetic resonance.

[9]  Yun Jiang,et al.  SVD Compression for Magnetic Resonance Fingerprinting in the Time Domain , 2014, IEEE Transactions on Medical Imaging.

[10]  Hung-Mo Lin,et al.  Computer programs for the concordance correlation coefficient , 2007, Comput. Methods Programs Biomed..

[11]  Vikas Gulani,et al.  MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. , 2015, Magnetic resonance in medicine.

[12]  Tobias Knopp,et al.  Field Inhomogeneity Correction Based on Gridding Reconstruction for Magnetic Resonance Imaging , 2007, IEEE Transactions on Medical Imaging.

[13]  John M. Pauly,et al.  Improved automatic off‐resonance correction without a field map in spiral imaging , 1997, Magnetic resonance in medicine.

[14]  Isaac Amidror,et al.  Scattered data interpolation methods for electronic imaging systems: a survey , 2002, J. Electronic Imaging.

[15]  Yong Chen,et al.  MR Fingerprinting for Rapid Quantitative Abdominal Imaging. , 2016, Radiology.

[16]  D. O. Walsh,et al.  Adaptive reconstruction of phased array MR imagery , 2000, Magnetic resonance in medicine.

[17]  Hugo R. Shi,et al.  Toeplitz-based iterative image reconstruction for MRI with correction for magnetic field inhomogeneity , 2005, IEEE Transactions on Signal Processing.

[18]  A Macovski,et al.  Multifrequency interpolation for fast off‐resonance correction , 1997, Magnetic resonance in medicine.

[19]  K. Scheffler,et al.  Principles and applications of balanced SSFP techniques , 2003, European Radiology.

[20]  Nicholas R Zwart,et al.  Spiral trajectory design: A flexible numerical algorithm and base analytical equations , 2014, Magnetic resonance in medicine.

[21]  Peng Hu,et al.  Comparison of T1 measurement using ISMRM/NIST system phantom | NIST , 2016 .