Total Power Radiometer for Microwave Noise Measurement
暂无分享,去创建一个
Two types of radiometers are generally used for microwave noise measurements. One is a total power type and the other is the Dicke type. The former is theoretically the most sensitive and more broadband in frequency, but it is slightly unstable and worse in the linearity of the receiver gain. From the viewpoint of realizing a broadband radiometer, the total power type has been adopted for precision noise measurements. In order to reduce the influence of non-linearity, a differential detection scheme has been employed in the detection section. And also in order to reduce the instability of the gain, the temperature of the front end section and the IF and detection section has been stabilized with circulated warmed water. Based on these ideas, a new microwave radiometer has been built. In addition a new equation of the output indication for the total power radiometer has been derived. By using this equation, an equation has been derived for making noise calibration and several equations have been derived for measuring parameters indicating the characteristics of the radiometer, which are important to perform precision noise measurements. And the effectiveness of the radiometer has been proven experimentally using these equations.
[1] S. P. Pucic. Derivation of the system equation for null-balanced total-power radiometer system NCS1 , 1994 .
[2] Ichiro Yokoshima,et al. A Generalized Performance Analysis of One-Port Measuring Instruments with Signal Generators , 1972 .
[3] I. Yokoshima,et al. A 4-GHz band low-noise measurement system , 1987, IEEE Transactions on Instrumentation and Measurement.