The importance of assessing the fidelity of electrocardiographs by using a simple calibration signal

ECG distortion, mimicking changes seen in acute myocardial infarction, occurred in traces transmitted by the South East Coast NHS Ambulance Trust. This was due to low frequency phase nonlinearity produced by baseline filtering. The very narrow calibration pulses made the detection of this distortion difficult. A 1 mV square wave signal revealed the cause. Adequate calibration pulses 0.3 s wide should be recorded with all ECGs, to show that the signal is being reproduced reliably.

[1]  E. W. Hancock,et al.  AHA/ACCF/HRS recommendations for the standardization and interpretation of the electrocardiogram: part IV: the ST segment, T and U waves, and the QT interval: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the , 2009, Journal of the American College of Cardiology.

[2]  D I Tayler,et al.  Artefactual ST segment abnormalities due to electrocardiograph design. , 1985, British heart journal.

[3]  W. Einthoven THE DIFFERENT FORMS OF THE HUMAN ELECTROCARDIOGRAM AND THEIR SIGNIFICATION. , 1912 .

[4]  James W. Wagner A Simple Test for Phase Characteristics of Chart Recorders , 1979, IEEE Transactions on Biomedical Engineering.

[5]  A. Kerwin,et al.  The Effect of the Frequency Response of Electrocardiographs on the Form of Electrocardiograms and Vectorcardiograms , 1953, Circulation.

[6]  Longmore Db ADRENAL FUNCTION AND ASTHMA. , 1965 .

[7]  O. H. Schmitt,et al.  Report of committee on electrocardiography, American Heart Association. Recommendations for standardization of leads and of specifications for instruments in electrocardiography and vectorcardiography. , 1967, Circulation.

[8]  E. Heinonen Frequency response testing of ECG devices , 2006, Medical and Biological Engineering and Computing.

[9]  A. Sykes A D Waller and the electrocardiogram, 1887. , 1987, British medical journal.

[10]  P. Macfarlane,et al.  Recommendations for standardization and specifications in automated electrocardiography: bandwidth and digital signal processing. A report for health professionals by an ad hoc writing group of the Committee on Electrocardiography and Cardiac Electrophysiology of the Council on Clinical Cardiology, , 1990, Circulation.

[11]  H V Pipberger,et al.  The low-frequency response of electrocardiographs, a frequent source of recording errors. , 1966, American heart journal.

[12]  D. Tayler,et al.  Signal Distortion in the Electrocardiogram Due to Inadequate Phase Response , 1983, IEEE Transactions on Biomedical Engineering.