Intercomparison of calibration procedures for 192Ir HDR sources in Brazil.

The lack of well established dosimetry protocols for HDR sources is a point of great concern regarding the uniformity of procedures within a particular country and worldwide. The main objective of this paper is to report the results from ten institutions of an intercomparison of calibration procedures for 192Ir HDR sources currently in use in Brazil. The treatment irradiator of one institution was calibrated by a reference system and used by all participants with their own measuring electrometers and ionization chambers under the same experimental conditions. Two methods were used: the calibration jig and the well-type ionization chamber. Each participant was allowed to use their own method and formalism. The results of this exercise were very positive since this was the first time in Brazil that a group of users gathered to share their experience and openly discuss the physical concepts behind the calibration procedures. The results were all within +/-3.0%, except one case where -4.6% was observed and later identified as a problem with the Nk value for x-rays. Though the magnitude of the deviations found was generally acceptable considering the diversity of formalisms currently in use, a proposal is now being prepared to be adopted as a national protocol. The identification of the institutions was left out for the sake of confidentiality.