Modeling current density distributions during transcutaneous cardiac pacing

The authors developed a two-dimensional finite element model of a cross-section of the human thorax to study the current density distribution during transcutaneous cardiac pacing. The model comprises 964 nodes and 1,842 elements and accounted for the electrical properties of eight different tissues or organs and also simulated the anisotropies of the intercostal muscles. The finite element software employed was a version for electrokinetics problems of Finite Element for Heat Transfer (FEHT) and the authors assessed the effects upon the efficacy of transcutaneous cardiac pacing of several electrode placements and sizes. To minimize pain in the chest wall and still be able to capture the heart, the authors minimized the ratio, R, between the current density in the thoracic wall (which causes pain) and the current density in the heart wall (which captures the heart). The best placement of the negative electrode was over the cardiac apex. The best placement of the positive electrode was under the right scapula, although other placements mere nearly as good. The efficiency of pacing increased as electrode size increased up to 70 cm and showed little improvement for larger areas. Between different configurations of the precordial electrodes V1, V2, /spl middot//spl middot//spl middot/, V6 the most efficient configuration to pace with was V1 and V2 positive and V5 and V6 negative. A more efficient configuration uses an auxiliary electrode located at the right subscapular region.<<ETX>>

[1]  J. D. Bourland,et al.  Fundamental criteria underlying the efficacy and safety of defibrillating current waveforms , 2006, Medical and Biological Engineering and Computing.

[2]  N. G. Sepulveda,et al.  Finite element analysis of cardiac defibrillation current distributions , 1990, IEEE Transactions on Biomedical Engineering.

[3]  L A Geddes,et al.  Precordial Pacing Windows , 1984, Pacing and clinical electrophysiology : PACE.

[4]  J J Bocka,et al.  External transcutaneous pacemakers. , 1989, Annals of emergency medicine.

[5]  J. G. Webster,et al.  Analysis and Control of the Current Distribution under Circular Dispersive Electrodes , 1982, IEEE Transactions on Biomedical Engineering.

[6]  P. Zoll,et al.  Resuscitation of the heart in ventricular standstill by external electric stimulation. , 1952, The New England journal of medicine.

[7]  R. Falk,et al.  External cardiac pacing: Influence of electrode placement on pacing threshold , 1986, Critical care medicine.

[8]  L. Baker,et al.  Origins of the impedance change in impedance cardiography by a three-dimensional finite element model , 1988, IEEE Transactions on Biomedical Engineering.

[9]  K Davies Prevention of pressure sores. , 1994, British journal of nursing.

[10]  L A Geddes,et al.  Choice of the Optimum Pulse Duration for Precordial Cardiac Pacing: A Theoretical Study , 1985, Pacing and clinical electrophysiology : PACE.

[11]  Mary N. Morrow,et al.  A Comparison of Finite Element and Integral Equation Formulations for the Calculation of Electrocardiographic Potentials , 1985, IEEE Transactions on Biomedical Engineering.

[12]  S J Walker,et al.  Forward and Inverse Electrocardiographic Calculations Using Resistor Network Models of the Human Torso , 1987, Circulation research.

[13]  Osama A. Mohammed,et al.  Implanted Electrodes Potential Distribution In A Man Model , 1990, [1990] Proceedings of the Twelfth Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[14]  P. Dimbylow,et al.  Finite difference calculations of current densities in a homogeneous model of a man exposed to extremely low frequency electric fields. , 1987, Bioelectromagnetics.

[15]  F. G. Stremler,et al.  Numerical solutions and mapping of electrostatic fields using the Apple MacIntosh computer , 1990 .

[16]  A. van Oosterom,et al.  The potential distribution generated by surface electrodes in inhomogeneous volume conductors of arbitrary shape , 1991, IEEE Transactions on Biomedical Engineering.

[17]  R. P. Patterson,et al.  Sources of the thoracic cardiogenic electrical impedance signal as determined by a model , 1985, Medical and Biological Engineering and Computing.

[18]  Yongmin Kim,et al.  Optimal Electrode Configurations for External Cardiac Pacing and Defibrillation: An Inhomogeneous Study , 1987, IEEE Transactions on Biomedical Engineering.

[19]  John G. Webster,et al.  A finite element model with node renumbering for adaptive impedance imaging , 1988, Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society.