Activity of single ion channel proteins detected with a planar microstructure

We present recordings of currents mediated by single ion channel proteins obtained using planar, microstructured glass chips. In these chips, pores with diameters of 1–2 μm are produced by ion track etching and are used for patch clamping instead of using the classical micropipette. Our results represent success in using such devices to record from single channels in cell membranes. The planar chip greatly enhances the accessibility of the ion channel containing membrane and can serve as a workbench for experiments on single ion channels using combinations of patch clamp current recording with other single molecule techniques.

[1]  B. Sakmann,et al.  Single-channel currents recorded from membrane of denervated frog muscle fibres , 1976, Nature.

[2]  K. Magleby,et al.  Single channel recordings of Ca2+-activated K+ currents in rat muscle cell culture , 1981, Nature.

[3]  B. Hille Ionic channels of excitable membranes , 2001 .

[4]  F. Hofmann,et al.  Regulation of stably expressed and native BK channels from human myometrium by cGMP- and cAMP-dependent protein kinase , 1998, Pflügers Archiv.

[5]  N. Fertig,et al.  Stable integration of isolated cell membrane patches in a nanomachined aperture: a step towards a novel device for membrane physiology , 1999, cond-mat/9910217.

[6]  Horst Vogel,et al.  Chip based biosensor for functional analysis of single ion channels , 2000 .

[7]  Lei Wu,et al.  Ion-channel assay technologies: quo vadis? , 2001, Drug discovery today.

[8]  C. Trautmann,et al.  Microstructured glass chip for ion-channel electrophysiology. , 2001, Physical review. E, Statistical, nonlinear, and soft matter physics.

[9]  F Bezanilla,et al.  Bilayer reconstitution of voltage-dependent ion channels using a microfabricated silicon chip. , 2001, Biophysical journal.

[10]  Robert H Blick,et al.  Whole cell patch clamp recording performed on a planar glass chip. , 2002, Biophysical journal.

[11]  F. Sigworth,et al.  Patch clamp on a chip. , 2002, Biophysical journal.

[12]  M. Reed,et al.  Micromolded PDMS planar electrode allows patch clamp electrical recordings from cells. , 2002, Biosensors & bioelectronics.