The IRAC point response function in the warm Spitzer mission

The Infrared Array Camera (IRAC) is now the only science instrument in operation on the Spitzer Space Telescope. The 3.6 and 4.5 µm channels are temperature-stabilized at ~28.7K, and the sensitivity of IRAC is nearly identical to what it was in the cryogenic mission. The instrument point response function (PRF) is a set of values from which one can determine the point spread function (PSF) for a source at any position in the field, and is dependent on the optical characteristics of the telescope and instrument as well as the detector sampling and pixel response. These data are necessary when performing PSF-fitting photometry of sources, for deconvolving an IRAC image, subtracting out a bright source in a field, or for estimating the flux of a source that saturates the detector. Since the telescope and instrument are operating at a higher temperature in the post-cryogenic mission, we re-derive the PRFs for IRAC from measurements obtained after the warm mission temperature set point and detector biases were finalized and compare them to the 3.6 and 4.5 µm PRFs determined during the cryogenic mission to assess any changes.