Comparative study of three semiautomated specular microscopes

Abstract We compared two clinical video‐assisted specular microscopes (Zeiss, noncontact, and the wide‐field Keeler Konan sp 3300, contact) with an autofocus microscope (Konan noncon Robo‐ca sp 8000, noncontact) with built‐in analyzing software by studying the morphometry of the central corneal endothelium of 12 eyes in 12 patients. The mean coefficients of variation of the cell size analysis of three successive images per cornea for the three methods were 3.9%, 4.0%, and 2.2%, respectively. One‐way analysis of variance showed a significant difference in the mean cell area measured with the three methods; there was no significant difference in the results of hexagonality (pleomorphism) and coefficient of variation (polymegathism). Although each microscope showed a high reproducibility, we do not recommend using the three interchangeably. The Konan noncon Robo‐ca sp 8000 specular microscope provides a rapid morphometric endothelial analysis. The Keeler Konan sp 3300 is better for more detailed studies and photography of the corneal endothelium.

[1]  S. Macrae,et al.  The long-term effects of polymethylmethacrylate contact lens wear on the corneal endothelium. , 1994, Ophthalmology.

[2]  H. Cheng,et al.  Contact and noncontact specular microscopy. , 1981, The British journal of ophthalmology.

[3]  O Nishi,et al.  Automated morphometry of corneal endothelial cell: use of video camera and video tape recorder. , 1988, The British journal of ophthalmology.

[4]  R A Laing,et al.  In vivo photomicrography of the corneal endothelium. , 1975, Archives of ophthalmology.

[5]  O Nishi,et al.  Automated Determination of Polygonality of Corneal Endothelial Cells , 1989, Cornea.

[6]  R. Buckley,et al.  Clinical specular microscopy of the corneal endothelium. , 1981, Transactions of the ophthalmological societies of the United Kingdom.

[7]  W Köditz,et al.  Automated Morphometric Endothelial Analysis Combined with Video Specular Microscopy , 1984, Cornea.

[8]  C. Koester,et al.  Wide field specular microscopy. Clinical and research applications. , 1980, Ophthalmology.

[9]  T. Olsen OPTICAL PRINCIPLES FOR ESTIMATION OF ENDOTHELIAL CELL DENSITY WITH THE NON‐CONTACT SPECULAR MICROSCOPE , 1979, Acta ophthalmologica.

[10]  Buckley Rj,et al.  Clinical specular microscopy of the corneal endothelium. , 1981 .

[11]  K R Kenyon,et al.  The Corneal Endothelium , 1982 .

[12]  W Köditz,et al.  [Morphometry of the endothelium using automated television image analysis]. , 1983, Klinische Monatsblatter fur Augenheilkunde.

[13]  D. G. Grant,et al.  Automated Analysis of Wide‐Field Specular Photomicrographs , 1984, Cornea.

[14]  H J Gundersen,et al.  Estimators of the number of objects per area unbiased by edge effects. , 1978, Microscopica acta.

[15]  G O Waring,et al.  Four methods of measuring human corneal endothelial cells from specular photomicrographs. , 1980, Archives of ophthalmology.

[16]  H E Kaufman,et al.  Specular microscopy of human corneal endothelium in vivo. , 1976, American journal of ophthalmology.

[17]  H E Kaufman,et al.  Clinical specular microscopy. , 1976, Transactions. Section on Ophthalmology. American Academy of Ophthalmology and Otolaryngology.

[18]  K. T. Nguyen,et al.  Assessment of the Reliability of Calculations of the Coefficient of Variation for Normal and Polymegethous Human Corneal Endothelium , 1993, Optometry and vision science : official publication of the American Academy of Optometry.

[19]  J. F. Brenner,et al.  Automated morphometric analysis of corneal endothelial cells. , 1981, Investigative ophthalmology & visual science.

[20]  Charles J. Koester,et al.  Wide-Field Specular Microscopy , 1980 .

[21]  C. Hanna,et al.  Endothelial cell population changes of human cornea during life. , 1978, Archives of ophthalmology.

[22]  N. Boot,et al.  Long-term changes after toxic endothelial cell destruction of the human cornea , 1994 .

[23]  D. Hodge,et al.  Continued endothelial cell loss ten years after lens implantation. , 1994, Ophthalmology.

[24]  G. N. Rao Morphometry of corneal endothelium. , 1984, Cornea.