Separation and identification of picogram levels of dioxins and PCBs by GC/cryogenic trapping FTIR
暂无分享,去创建一个
Capillary gas chromatography/mass spectrometry (GC/MS) has routinely been used by the analytical chemist to separate and identify low levels of environmentally important compounds. A GC/Cryogenic Trapping Fourier Transform Infrared Spectrometer (Tracer) provides the sensitivity of the GC/MS with the added capability of differentiating between compounds of the same mass. In this work, the Tracer was utilized to study low levels of six Polychlorinated Biphenyls (PCBs), eight Chlorinated Dibenzo-p-Doxins and Norflurazon. In all cases, picogram levels of these compounds were easily detected from `on the fly' generated IR chromatograms. Since the separated compounds eluting from the capillary column are cryogenically trapped onto a moving liquid nitrogen cooled ZnSe crystal, excellent signal-to- noise spectra of these same compounds may be collected after the run by returning to the same areas of deposition and signal averaging.