A numerical model to reproduce squeaking of ceramic-on-ceramic total hip arthroplasty. Influence of design and material.

BACKGROUND Modern ceramic (CoC) bearings for hip arthroplasty (THA) have been used in younger patients who expect improved survivorship. However, audible squeaking produced by the implant is an annoying complication. Previous numerical simulations were not able to accurately reproduce in vitro and in vivo observations. Therefore, we developed a finite element model to: (1) reproduce in vitro squeaking and validate the model by comparing it with in vivo recordings, (2) determine why there are differences between in vivo and in vitro squeaking frequencies, (3) identify the stem's role in this squeaking, (4) predict which designs and materials are more likely to produce squeaking. HYPOTHESIS A CoC THA numerical model can be developed that reproduces the squeaking frequencies observed in vivo. MATERIAL AND METHODS Numerical methods (finite element analysis [ANSYS]) and experimental methods (using a non-lubricated simulated hip with a cementless 32mm CoC THA) were developed to reproduce squeaking. Numerical analysis was performed to identify the frequencies that cause vibrations perceived as an acoustic emission. The finite element analysis (FEA) model was enhanced by adjusting periprosthetic bone and soft tissue elements in order to reproduce the squeaking frequencies recorded in vivo. A numerical method (complex eigenvalue analysis) was used to find the acoustic frequencies of the squeaking noise. The frequencies obtained from the model and the hip simulator were compared to those recorded in vivo. RESULTS The numerical results were validated by experiments with the laboratory hip simulator. The frequencies obtained (mean 2790Hz with FEA, 2755Hz with simulator, decreasing to 1759Hz when bone and soft tissue were included in the FEA) were consistent with those of squeaking hips recorded in vivo (1521Hz). The cup and ceramic insert were the source of the vibration, but had little influence on the diffusion of the noise required to make the squeaking audible to the human ear. The FEA showed that diffusion of squeaking was due to an unstable vibration of the stem during frictional contact. The FEA predicted a higher rate of squeaking (at a lower coefficient of friction) when TZMF™ alloy is used instead of Ti6Al4V and when an anatomic press-fit stem is used instead of straight self-locking designs. DISCUSSION The current FEA model is reliable; it can be used to assess various stem designs and alloys to predict the different rates of squeaking that certain stems will likely produce. LEVEL OF EVIDENCE Level IV in vitro study.

[1]  C. Colwell,et al.  Incidence of ‘Squeaking’ After Ceramic-on-Ceramic Total Hip Arthroplasty , 2010, Clinical orthopaedics and related research.

[2]  John Fisher,et al.  In vitro investigation of friction under edge‐loading conditions for ceramic‐on‐ceramic total hip prosthesis , 2010, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[3]  O. Mahoney,et al.  Can squeaking with ceramic-on-ceramic hip articulations in total hip arthroplasty be avoided? , 2013, The Journal of arthroplasty.

[4]  J. Parvizi,et al.  A Squeaky Reputation: The Problem May Be Design-dependent , 2011, Clinical orthopaedics and related research.

[5]  G Willmann,et al.  Mismatched wear couple zirconium oxide and aluminum oxide in total hip arthroplasty. , 2001, The Journal of arthroplasty.

[6]  J. Capozzi,et al.  Squeaking in third- and fourth-generation ceramic-on-ceramic total hip arthroplasty: meta-analysis and systematic review. , 2012, The Journal of arthroplasty.

[7]  Matteo Bruzzone,et al.  The squeaking hip: a phenomenon of ceramic-on-ceramic total hip arthroplasty. , 2009, The Journal of bone and joint surgery. American volume.

[8]  Camilo Restrepo,et al.  The effect of stem design on the prevalence of squeaking following ceramic-on-ceramic bearing total hip arthroplasty. , 2010, The Journal of bone and joint surgery. American volume.

[9]  W. Walter,et al.  Mid-term results of third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty: a ten-year minimum follow-up. , 2012, The Journal of bone and joint surgery. American volume.

[10]  Laurent Sedel,et al.  Alumina-on-Alumina Total Hip Arthroplasty: A Minimum 18.5-Year Follow-up Study , 2002, The Journal of bone and joint surgery. American volume.

[11]  Laurent Sedel,et al.  Ceramics in total hip replacement. , 2005, Clinical orthopaedics and related research.

[12]  S. Kurtz,et al.  The noisy ceramic hip: is component malpositioning the cause? , 2008, The Journal of arthroplasty.

[13]  John Fisher,et al.  Three-dimensional modeling of in vitro hip kinematics under micro-separation regime for ceramic on ceramic total hip prosthesis: an analysis of vibration and noise. , 2010, Journal of biomechanics.

[14]  D. Dennis,et al.  Correlation of Visual Images and Audible Sound for Total Hip Arthroplasties having Variable Bearing Surfaces , 2008 .

[15]  W. Capello,et al.  Alumina Ceramic Bearings for Total Hip Arthroplasty: Five-year Results of a Prospective Randomized Study , 2005, Clinical orthopaedics and related research.

[16]  T. Wright,et al.  Retrieval analysis of squeaking ceramic implants: are there related specific features? , 2012, Orthopaedics & traumatology, surgery & research : OTSR.

[17]  L. Sedel,et al.  Evolution of alumina-on-alumina implants: a review. , 2000, Clinical orthopaedics and related research.

[18]  A. Akay Acoustics of friction. , 2002, The Journal of the Acoustical Society of America.

[19]  B. W. Schreurs,et al.  High Incidence of Squeaking in THAs with Alumina Ceramic-on-ceramic Bearings , 2008, Clinical orthopaedics and related research.

[20]  Annalisa Fregolent,et al.  Instability scenarios between elastic media under frictional contact , 2013 .

[21]  Aldo Sestieri,et al.  Brake squeal: Linear and nonlinear numerical approaches , 2007 .

[22]  J. Fisher,et al.  Spectral characterization of squeaking in ceramic‐on‐ceramic total hip arthroplasty: Comparison of in vitro and in vivo values , 2012, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[23]  Linmao Qian,et al.  Analysis of squeaking on ceramic hip endoprosthesis using the complex eigenvalue method , 2011 .

[24]  L. Sedel,et al.  Occurrence of noise in alumina-on-alumina total hip arthroplasty. A survey on 284 consecutive hips. , 2011, Orthopaedics & traumatology, surgery & research : OTSR.

[25]  Arne Hothan,et al.  The influence of component design, bearing clearance and axial load on the squeaking characteristics of ceramic hip articulations. , 2011, Journal of biomechanics.

[26]  Laurent Baillet,et al.  Brake squeal as dynamic instability: An experimental investigation , 2006 .

[27]  J. Parvizi,et al.  Natural History of Squeaking after Total Hip Arthroplasty , 2010, Clinical orthopaedics and related research.

[28]  O. Guyen,et al.  Nine years follow-up of 100 ceramic-on-ceramic total hip arthroplasty , 2010, International Orthopaedics.

[29]  J. Currier,et al.  A proposed mechanism for squeaking of ceramic-on-ceramic hips , 2010 .

[30]  Michael M Morlock,et al.  Squeak in hip endoprosthesis systems: An experimental study and a numerical technique to analyze design variants. , 2010, Medical engineering & physics.

[31]  P. Boyer,et al.  Is alumina-on-alumina ceramic bearings total hip replacement the right choice in patients younger than 50 years of age? A 7- to 15-year follow-up study. , 2010, Orthopaedics & traumatology, surgery & research : OTSR.

[32]  W. Walter,et al.  Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. , 2007, The Journal of arthroplasty.

[33]  Kate Sutton,et al.  The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings. , 2007, The Journal of arthroplasty.

[34]  Michael M. Morlock,et al.  DECREASING THE SUSCEPTIBILITY OF CERAMIC BEARINGS TO SQUEAKING BY INCREASE OF THE CRITICAL FRICTION COEFFICIENT , 2012 .