Application of laser shock peening for spinal implant rods

Purpose – The current industry standard rigid spinal implants suffer fatigue failures due to bending and torsion loads. The purpose of this program was to design novel prototype flexible titanium alloy spinal implant rod with machined features, and then apply the laser shock peening (LSP) process to restore the fatigue strength debit due to these features.Design/methodology/approach – A flexible prototype rod was designed with flat section at the center of the rod. The flat section was laser shock peened. Static compression tests were conducted as per American Society of Testing Materials standards for three‐ and four‐point bending tests and “vertebrectomy” constructs. Finite element models were developed to aid in the design of LSP and also to guide the experiments.Findings – The test results indicated a ∼3X improvement in flexibility and a reduction in fatigue load ratio, defined as applied load divided by the yield load; from 72 to 68 percent. This rod was LSP's on the flat sections, and tested again. ...