Volumetric imaging with homogenised excitation and static field at 9.4 T
暂无分享,去创建一个
Klaus Scheffler | Dimo Ivanov | Kâmil Uludağ | Benedikt A. Poser | Daniel Brenner | Christopher J. Wiggins | Jens Hoffmann | Desmond H. Y. Tse | Gunamony Shajan | Christian Mirkes | K. Scheffler | K. Uludağ | C. Wiggins | D. Ivanov | B. Poser | C. Mirkes | J. Hoffmann | G. Shajan | D. Tse | D. Brenner
[1] Lawrence L. Wald,et al. Three dimensional echo-planar imaging at 7 Tesla , 2010, NeuroImage.
[2] Steen Moeller,et al. B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil , 2005, Magnetic resonance in medicine.
[3] D. Le Bihan,et al. Orientation Dependence of White Matter T 2 * Contrast at 7 T : A Direct Demonstration , 2007 .
[4] Ravi S. Menon,et al. Slice‐by‐slice B1+ shimming at 7 T , 2012, Magnetic resonance in medicine.
[5] Michael B. Smith,et al. Exploring the limits of RF shimming for high‐field MRI of the human head , 2006, Magnetic resonance in medicine.
[6] W. Edelstein,et al. The intrinsic signal‐to‐noise ratio in NMR imaging , 1986, Magnetic resonance in medicine.
[7] Natalia Petridou,et al. Fast high resolution whole brain T2* weighted imaging using echo planar imaging at 7T , 2011, NeuroImage.
[8] J. Duyn,et al. Magnetic susceptibility mapping of brain tissue in vivo using MRI phase data , 2009, Magnetic resonance in medicine.
[9] Kay Nehrke,et al. On the steady‐state properties of actual flip angle imaging (AFI) , 2009, Magnetic resonance in medicine.
[10] Benjamin Zahneisen,et al. SENSE and simultaneous multislice imaging , 2015, Magnetic resonance in medicine.
[11] Lawrence L. Wald,et al. Fast quantitative susceptibility mapping with L1‐regularization and automatic parameter selection , 2013, Magnetic resonance in medicine.
[12] J. Frahm,et al. The influence of the slice-selection gradient on functional MRI of human brain activation. , 1994, Journal of magnetic resonance. Series B.
[13] Tobias Kober,et al. Dielectric pads and low‐ B1+ adiabatic pulses: Complementary techniques to optimize structural T1w whole‐brain MP2RAGE scans at 7 tesla , 2014, Journal of magnetic resonance imaging : JMRI.
[14] Alessandro Sbrizzi,et al. Fast design of local N‐gram‐specific absorption rate–optimized radiofrequency pulses for parallel transmit systems , 2012, Magnetic resonance in medicine.
[15] Douglas C Noll,et al. Spatial domain method for the design of RF pulses in multicoil parallel excitation , 2006, Magnetic resonance in medicine.
[16] Alexis Amadon,et al. Parallel-transmission-enabled magnetization-prepared rapid gradient-echo T1-weighted imaging of the human brain at 7T , 2012, NeuroImage.
[17] Klaus Scheffler,et al. Functional MRI in human subjects with gradient‐echo and spin‐echo EPI at 9.4 T , 2014, Magnetic resonance in medicine.
[18] G. Glover,et al. Correction of physiologically induced global off‐resonance effects in dynamic echo‐planar and spiral functional imaging , 2002, Magnetic resonance in medicine.
[19] Hoby P Hetherington,et al. Role of very high order and degree B0 shimming for spectroscopic imaging of the human brain at 7 tesla , 2012, Magnetic resonance in medicine.
[20] Klaus Scheffler,et al. Safety testing and operational procedures for self‐developed radiofrequency coils , 2016, NMR in biomedicine.
[21] Yudong Zhu,et al. Parallel excitation with an array of transmit coils , 2004, Magnetic resonance in medicine.
[22] R. Ladebeck,et al. In vivo magnetic resonance imaging and spectroscopy of humans with a 4 t whole‐body magnet , 1990, NMR in biomedicine.
[23] Hoult,et al. Selective spin inversion in nuclear magnetic resonance and coherent optics through an exact solution of the Bloch-Riccati equation. , 1985, Physical review. A, General physics.
[24] A. Shmuel,et al. Imaging brain function in humans at 7 Tesla , 2001, Magnetic resonance in medicine.
[25] Klaus Scheffler,et al. Ultra-high resolution imaging of the human brain using acquisition-weighted imaging at 9.4T , 2014, NeuroImage.
[26] M. Zaitsev,et al. High resolution single-shot EPI at 7T , 2008, Magnetic Resonance Materials in Physics, Biology and Medicine.
[27] W. Barber,et al. Comparison of linear and circular polarization for magnetic resonance imaging , 1985 .
[28] Peter Jezzard,et al. Correction of geometric distortion in fMRI data , 2012, NeuroImage.
[29] P A Bottomley,et al. RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imaging. , 1978, Physics in medicine and biology.
[30] Stephen M Smith,et al. Fast robust automated brain extraction , 2002, Human brain mapping.
[31] A. MacKay,et al. The influence of white matter fibre orientation on MR signal phase and decay , 2011, NMR in biomedicine.
[32] P. Börnert,et al. Transmit SENSE , 2003, Magnetic resonance in medicine.
[33] Steen Moeller,et al. T 1 weighted brain images at 7 Tesla unbiased for Proton Density, T 2 ⁎ contrast and RF coil receive B 1 sensitivity with simultaneous vessel visualization , 2009, NeuroImage.
[34] Nikolaus Weiskopf,et al. High-Resolution Functional MRI at 3 T: 3D/2D Echo-Planar Imaging with Optimized Physiological Noise Correction , 2012, Magnetic resonance in medicine.
[35] W van der Zwaag,et al. Temporal SNR characteristics in segmented 3D‐EPI at 7T , 2012, Magnetic resonance in medicine.
[36] Peter Börnert,et al. Volumetric B1+ Mapping of the Brain at 7T using DREAM , 2014, Magnetic resonance in medicine.
[37] E. Adalsteinsson,et al. Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channels , 2008, Magnetic resonance in medicine.
[38] Cristina Granziera,et al. Cerebellar Cortical Layers : In Vivo Visualization with Structural High-Field-Strength MR Imaging 1 , 2010 .
[39] Jeff H Duyn,et al. A torque balance measurement of anisotropy of the magnetic susceptibility in white matter , 2015, Magnetic resonance in medicine.
[40] P. Jezzard,et al. Correction for geometric distortion in echo planar images from B0 field variations , 1995, Magnetic resonance in medicine.
[41] Steen Moeller,et al. A geometrically adjustable 16‐channel transmit/receive transmission line array for improved RF efficiency and parallel imaging performance at 7 Tesla , 2008, Magnetic resonance in medicine.
[42] Klaus Scheffler,et al. Numerical and experimental evaluation of RF shimming in the human brain at 9.4 T using a dual-row transmit array , 2014, Magnetic Resonance Materials in Physics, Biology and Medicine.
[43] Robin M Heidemann,et al. Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA) , 2006, Magnetic resonance in medicine.
[44] Richard Bowtell,et al. Whole-brain susceptibility mapping at high field: A comparison of multiple- and single-orientation methods , 2010, NeuroImage.
[45] Klaus Scheffler,et al. Signal‐to‐noise ratio and MR tissue parameters in human brain imaging at 3, 7, and 9.4 tesla using current receive coil arrays , 2016, Magnetic resonance in medicine.
[46] Peter Andersen,et al. 9.4T human MRI: Preliminary results , 2006, Magnetic resonance in medicine.
[47] Cornelis A T van den Berg,et al. Time efficient design of multi dimensional RF pulses: Application of a multi shift CGLS algorithm , 2011, Magnetic resonance in medicine.
[48] R. Cusack,et al. New Robust 3-D Phase Unwrapping Algorithms: Application to Magnetic Field Mapping and Undistorting Echoplanar Images , 2002, NeuroImage.
[49] Stephen M. Smith,et al. Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithm , 2001, IEEE Transactions on Medical Imaging.
[50] L. Axel,et al. Rapid B1+ mapping using a preconditioning RF pulse with TurboFLASH readout , 2010, Magnetic resonance in medicine.
[51] Jeff H. Duyn,et al. The future of ultra-high field MRI and fMRI for study of the human brain , 2012, NeuroImage.
[52] D Le Bihan,et al. kT‐points: Short three‐dimensional tailored RF pulses for flip‐angle homogenization over an extended volume , 2012, Magnetic resonance in medicine.
[53] P. Röschmann,et al. Spectroscopy and imaging with a 4 tesla whole‐body mr system , 1988, NMR in biomedicine.
[54] Rolf Pohmann,et al. Human imaging at 9.4 T using T2*‐, phase‐, and susceptibility‐weighted contrast , 2011, Magnetic resonance in medicine.
[55] J W Carlson,et al. Electromagnetic fields of surface coil in vivo NMR at high frequencies , 1991, Magnetic resonance in medicine.
[56] P. Batchelor,et al. International Society for Magnetic Resonance in Medicine , 1997 .
[57] Jeff H. Duyn,et al. High-field MRI of brain cortical substructure based on signal phase , 2007, Proceedings of the National Academy of Sciences.
[58] Jun Ma,et al. Joint design of large‐tip‐angle parallel RF pulses and blipped gradient trajectories , 2016, Magnetic resonance in medicine.
[59] Alexander Hammers,et al. Tailored excitation in 3D with spiral nonselective (SPINS) RF pulses , 2012, Magnetic resonance in medicine.
[60] Stephen P. Boyd,et al. Graph Implementations for Nonsmooth Convex Programs , 2008, Recent Advances in Learning and Control.
[61] Peter G. Morris,et al. fMRI at 1.5, 3 and 7 T: Characterising BOLD signal changes , 2009, NeuroImage.
[62] Gabriele Eichfelder,et al. Local specific absorption rate control for parallel transmission by virtual observation points , 2011, Magnetic resonance in medicine.
[63] Peter Börnert,et al. A specific absorption rate prediction concept for parallel transmission MR , 2012, Magnetic resonance in medicine.
[64] Johnpauly. A k-Space Analysis of Small-Tip-Angle Excitation , 2012 .
[65] Klaus Scheffler,et al. A 16‐channel dual‐row transmit array in combination with a 31‐element receive array for human brain imaging at 9.4 T , 2014, Magnetic resonance in medicine.
[66] Uwe Aickelin,et al. Tailored RF pulse for magnetization inversion at ultrahigh field , 2010, Magnetic resonance in medicine.
[67] Jörg Felder,et al. Encoding methods for B1(+) mapping in parallel transmit systems at ultra high field. , 2014, Journal of magnetic resonance.
[68] Shaihan J. Malik,et al. Initial experience with SPIral Non Selective ( SPINS ) RF pulses for homogeneous excitation at 7 T , 2013 .
[69] Andrew G Webb,et al. Quantitative assessment of the effects of high‐permittivity pads in 7 Tesla MRI of the brain , 2012, Magnetic resonance in medicine.
[70] J. Mugler,et al. Rapid three‐dimensional T1‐weighted MR imaging with the MP‐RAGE sequence , 1991, Journal of magnetic resonance imaging : JMRI.
[71] Elfar Adalsteinsson,et al. Online Local SAR Supervision for Transmit Arrays at 7 T , 2012 .
[72] Kristian Bredies,et al. Fast quantitative susceptibility mapping using 3D EPI and total generalized variation , 2015, NeuroImage.