RMDP: a dedicated package for 131I SPECT quantification, registration and patient-specific dosimetry.

The limitations of traditional targeted radionuclide therapy (TRT) dosimetry can be overcome by using voxel-based techniques. All dosimetry techniques are reliant on a sequence of quantitative emission and transmission data. The use of (131)I, for example, with NaI or mIBG, presents additional quantification challenges beyond those encountered in low-energy NM diagnostic imaging, including dead-time correction and additional photon scatter and penetration in the camera head. The Royal Marsden Dosimetry Package (RMDP) offers a complete package for the accurate processing and analysis of raw emission and transmission patient data. Quantitative SPECT reconstruction is possible using either FBP or OS-EM algorithms. Manual, marker- or voxel-based registration can be used to register images from different modalities and the sequence of SPECT studies required for 3-D dosimetry calculations. The 3-D patient-specific dosimetry routines, using either a beta-kernel or voxel S-factor, are included. Phase-fitting each voxel's activity series enables more robust maps to be generated in the presence of imaging noise, such as is encountered during late, low-count scans or when there is significant redistribution within the VOI between scans. Error analysis can be applied to each generated dose-map. Patients receiving (131)I-mIBG, (131)I-NaI, and (186)Re-HEDP therapies have been analyzed using RMDP. A Monte-Carlo package, developed specifically to address the problems of (131)I quantification by including full photon interactions in a hexagonal-hole collimator and the gamma camera crystal, has been included in the dosimetry package. It is hoped that the addition of this code will lead to improved (131)I image quantification and will contribute towards more accurate 3-D dosimetry.

[1]  Finbarr O'Sullivan,et al.  Reducing negativity artifacts in emission tomography: post-processing filtered backprojection solutions , 1993, IEEE Trans. Medical Imaging.

[2]  R Mohan,et al.  Treatment planning for internal radionuclide therapy: three-dimensional dosimetry for nonuniformly distributed radionuclides. , 1990, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[3]  I Clairand,et al.  DOSE3D: EGS4 Monte Carlo code-based software for internal radionuclide dosimetry. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[4]  G Sgouros,et al.  Implementation and evaluation of patient-specific three-dimensional internal dosimetry. , 1997, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  R. K. Wu,et al.  Absolute quantitation of radioactivity using the buildup factor. , 1984, Medical physics.

[6]  M. Flower,et al.  Estimation and implications of random errors in whole-body dosimetry for targeted radionuclide therapy. , 2002, Physics in medicine and biology.

[7]  P B Hoffer,et al.  Computerized three-dimensional segmented human anatomy. , 1994, Medical physics.

[8]  C E Floyd,et al.  A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT. , 1995, Physics in medicine and biology.

[9]  D A Podoloff,et al.  Development of a SPECT-based three-dimensional treatment planning system for radioimmunotherapy. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[10]  J. Chatal,et al.  Radionuclide therapy practice and facilities in Europe , 1999, European Journal of Nuclear Medicine.

[11]  J Y Wong,et al.  A radionuclide therapy treatment planning and dose estimation system. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[12]  Dimitris Visvikis,et al.  DETECT-dual energy transmission estimation CT-for improved attenuation correction in SPECT and PET , 1997 .

[13]  K. Ogawa,et al.  Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. , 1993, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[14]  C S Chui,et al.  A Monte Carlo approach to patient-specific dosimetry. , 1996, Medical physics.

[15]  M A Flower,et al.  An automated technique for SPECT marker-based image registration in radionuclide therapy , 2001, Physics in medicine and biology.

[16]  G J Hademenos,et al.  Comparison of four scatter correction methods using Monte Carlo simulated source distributions. , 1994, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[17]  I. Kanno,et al.  Automated interstudy image registration technique for SPECT and PET. , 1996, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[18]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[19]  D Front,et al.  SPECT quantitation of iodine-131 concentration in phantoms and human tumors. , 1990, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[20]  R E Carson,et al.  Three-dimensional registration of cardiac positron emission tomography attenuation scans. , 1993, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[21]  Guy Marchal,et al.  Multimodality image registration by maximization of mutual information , 1997, IEEE Transactions on Medical Imaging.

[22]  R. Vessella,et al.  Dosimetry of solid tumors. , 1993, Medical physics.