Computer generation of cobalt-60 single beam dose distribution using an analytical beam model.
暂无分享,去创建一个
A beam dose calculation model based on evaluation of tissue air ratios (TAR) and scatter air ratios (SAR) for cobalt-60 beams of rectangular cross section has been developed. Off-central axis fall-off of primary radiation intensity is derived by an empirical formulation involving an arctangent function with the slope of the geometrical penumbra acting as an essential constant. Central axis TAR and SAR values are assessed by semi-empirical polynomial expressions employing the two sides of the rectangular field as the variables. The model utilises a minimum number of parametric constants and is useful for computer generation of isodose curves. The model is capable of accounting for situations where wedge filters or split field shielding blocks, are encountered. Further it could be widely applied with minor modifications to several makes of the currently available cobalt-60 units. The paper explains the model and shows examples of the results obtained in comparison with the corresponding experimentally determined dose distributions.