Gantry-angle resolved VMAT pretreatment verification using EPID image prediction.

PURPOSE Pretreatment verification of volumetric modulated arc therapy (VMAT) dose delivery with electronic portal imaging device (EPID) uses images integrated over the entire delivery or over large subarcs. This work aims to develop a new method for gantry-angle-resolved verification of VMAT dose delivery using EPID. METHODS An EPID dose prediction model was used to calculate EPID images as a function of gantry angle for eight prostate patient deliveries. EPID image frames at 7.5 frames per second were acquired during delivery via a frame-grabber system. The gantry angle for each image was encoded in kV frames which were synchronized to the MV frames. Gamma analysis results as a function of gantry angle were assessed by integrating the frames over 2° subarcs with an angle-to-agreement tolerance of 0.5° about the measured image angle. RESULTS The model agreed with EPID images integrated over the entire delivery with average Gamma pass-rates at 2%, 2 mm of 99.7% (10% threshold). The accuracy of the kV derived gantry angle for each image was found to be 0.1° (1 SD) using a phantom test. For the gantry-resolved analysis all Gamma pass-rates were greater than 90% at 3%, 3 mm criteria (with only two exceptions), and more than 90% had a 95% pass-rate, with an average of 97.3%. The measured gantry angle lagged behind the predicted angle by a mean of 0.3°±0.3°, with a maximum lag of 1.3°. CONCLUSIONS The method provides a comprehensive and highly efficient pretreatment verification of VMAT delivery using EPID. Dose delivery accuracy is assessed as a function of gantry angle to ensure accurate treatment.

[1]  P. B. Greer,et al.  Long-term two-dimensional pixel stability of EPIDs used for regular linear accelerator quality assurance , 2011, Australasian Physical & Engineering Sciences in Medicine.

[2]  Peter B. Greer,et al.  Gantry angle determination during arc IMRT: evaluation of a simple EPID‐based technique and two commercial inclinometers , 2012, Journal of applied clinical medical physics.

[3]  Harish K. Malhotra,et al.  EPID dosimetry for pretreatment quality assurance with two commercial systems , 2012, Journal of applied clinical medical physics.

[4]  Suresh Senan,et al.  Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. , 2009, International journal of radiation oncology, biology, physics.

[5]  C. Ling,et al.  Commissioning and quality assurance of RapidArc radiotherapy delivery system. , 2008, International journal of radiation oncology, biology, physics.

[6]  J. Bedford,et al.  Commissioning of volumetric modulated arc therapy (VMAT). , 2009, International journal of radiation oncology, biology, physics.

[7]  M C Kirby,et al.  Verification of dynamic multileaf collimation using an electronic portal imaging device. , 2000, Physics in medicine and biology.

[8]  P. Lambin,et al.  A literature review of electronic portal imaging for radiotherapy dosimetry. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[9]  Daniel A Low,et al.  Linac mechanic QA using a cylindrical phantom , 2008, Physics in medicine and biology.

[10]  K. Chytyk,et al.  Comprehensive fluence model for absolute portal dose image prediction. , 2009, Medical physics.

[11]  Luca Cozzi,et al.  Intensity modulation with photons for benign intracranial tumours: a planning comparison of volumetric single arc, helical arc and fixed gantry techniques. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[12]  Stine Korreman,et al.  The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy , 2008, Radiation oncology.

[13]  Manuel Guizar-Sicairos,et al.  Efficient subpixel image registration algorithms. , 2008, Optics letters.

[14]  J. Hendry,et al.  International perspectives on quality assurance and new techniques in radiation medicine: outcomes of an IAEA conference. , 2008, International Journal of Radiation Oncology, Biology, Physics.

[15]  Peter B. Greer,et al.  3D Dose reconstruction: Banding artefacts in cine mode EPID images during VMAT delivery , 2013, Journal of Physics: Conference Series.

[16]  Luca Cozzi,et al.  A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  Karl Otto,et al.  Volumetric modulated arc therapy: IMRT in a single gantry arc. , 2007, Medical physics.

[18]  Luca Cozzi,et al.  Volumetric modulated arc radiotherapy for carcinomas of the oro-pharynx, hypo-pharynx and larynx: a treatment planning comparison with fixed field IMRT. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[19]  Qiuwen Wu,et al.  Independent verification of gantry angle for pre-treatment VMAT QA using EPID , 2012, Physics in medicine and biology.

[20]  Toshikazu Imae,et al.  Quality assurance of volumetric modulated arc therapy using Elekta Synergy , 2009, Acta oncologica.

[21]  Peter B. Greer,et al.  3D EPID based dosimetry for pre-treatment verification of VMAT – methods and challenges , 2013 .

[22]  B M C McCurdy,et al.  Model-based prediction of portal dose images during patient treatment. , 2013, Medical physics.

[23]  David Sjostrom,et al.  Implementing RapidArc into clinical routine: a comprehensive program from machine QA to TPS validation and patient QA. , 2011, Medical physics.

[24]  Fang-Fang Yin,et al.  Volumetric-modulated arc therapy: effective and efficient end-to-end patient-specific quality assurance. , 2010, International journal of radiation oncology, biology, physics.

[25]  David A Jaffray,et al.  Accurate technique for complete geometric calibration of cone-beam computed tomography systems. , 2005, Medical physics.

[26]  Cedric X. Yu,et al.  Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy. , 1995, Physics in medicine and biology.

[27]  M van Herk,et al.  Catching errors with in vivo EPID dosimetry. , 2010, Medical physics.

[28]  Peter B. Greer,et al.  An evaluation of cine-mode 3D portal image dosimetry for Volumetric Modulated Arc Therapy , 2010 .

[29]  Pejman Rowshanfarzad,et al.  Measurement and modeling of the effect of support arm backscatter on dosimetry with a varian EPID. , 2010, Medical physics.

[30]  Luca Cozzi,et al.  Volumetric-modulated arc radiotherapy for carcinomas of the anal canal: A treatment planning comparison with fixed field IMRT. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[31]  Peter B. Greer,et al.  SU‐E‐T‐210: Precise Gantry Angle Determination for EPID Images during Rotational IMRT , 2011 .

[32]  Suresh Senan,et al.  Volumetric modulated arc radiotherapy for vestibular schwannomas. , 2009, International journal of radiation oncology, biology, physics.

[33]  A Fogliata,et al.  Quality assurance of RapidArc in clinical practice using portal dosimetry. , 2011, The British journal of radiology.

[34]  L Kumaraswamy,et al.  Using an EPID for patient-specific VMAT quality assurance. , 2011, Medical physics.

[35]  Stine Korreman,et al.  RapidArc volumetric modulated therapy planning for prostate cancer patients , 2009, Acta oncologica.

[36]  Xizeng Wu,et al.  SU‐DD‐A1‐06: RapidArc Patient Specific Quality Assurance: Comparison with DMLC IMRT , 2009 .

[37]  E. Schreibmann,et al.  Patient-specific quality assurance method for VMAT treatment delivery. , 2009, Medical physics.

[38]  Jan-Jakob Sonke,et al.  3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[39]  Stine Korreman,et al.  Dosimetric verification of RapidArc treatment delivery , 2009, Acta oncologica.

[40]  Suresh Senan,et al.  Whole-brain radiotherapy with simultaneous integrated boost to multiple brain metastases using volumetric modulated arc therapy. , 2009, International journal of radiation oncology, biology, physics.

[41]  Pejman Rowshanfarzad,et al.  Detection and correction for EPID and gantry sag during arc delivery using cine EPID imaging. , 2012, Medical physics.