BACKGROUND
We were concerned about the risk of inadequate humidification during high-frequency percussive ventilation (HFPV).
METHODS
We studied 5 humidifiers during HFPV with a lung model, at bias gas flows of 10 L/min, 30 L/min, and 50 L/min, and compared the results to those from a comparator ventilator/humidifier setup and to the minimum temperature (30 degrees C) and humidity (30 mg/L) [corrected] recommended by the American Association for Respiratory Care, at both regular room temperature and a high ambient temperature. Temperature was measured at the humidifier outflow point and at the artificial carina. Humidity was measured at the artificial carina.
RESULTS
Of the 7 HFPV/humidifier combinations, 2 (the MR850 at a bias flow of 50 L/min, and the ConchaTherm Hi-Flow with VDR nebulizer) provided a carinal temperature equivalent to the comparator setup at room temperature, whereas one HFPV/humidifier combination (the ConchaTherm Hi-Flow with modified programming, at bias flows of 30 L/min and 50 L/min) provided a higher carinal temperature. At high ambient temperature, all of the setups delivered lower carinal temperature than the comparator setup. Only 2 setups (the ConchaTherm with modified programming at a bias flow of 50 L/min, and the ConchaTherm Hi-Flow with VDR nebulizer) provided carinal humidification equivalent to the comparator setup, without regard to ambient temperature; the other humidifiers were less effective. The ConchaTherm with modified programming, and the ConchaTherm with the VDR nebulizer provided the most consistent humidification.
CONCLUSION
HFPV's distinctive gas-flow mechanism may impair gas heating and humidification, so all humidification systems should be tested with HFPV prior to clinical use.
[1]
P. Allan,et al.
Corrective measures for compromised oxygen delivery during endotracheal tube cuff deflation with high-frequency percussive ventilation.
,
2007,
Respiratory care.
[2]
François Lellouche,et al.
Influence of ambient and ventilator output temperatures on performance of heated-wire humidifiers.
,
2004,
American journal of respiratory and critical care medicine.
[3]
R. Branson.
The effects of inadequate humidity.
,
1998,
Respiratory care clinics of North America.
[4]
R. Branson,et al.
Humidification in the intensive care unit.
,
1998,
Respiratory care clinics of North America.
[5]
AARC clinical practice guideline. Humidification during mechanical ventilation. American Association for Respiratory Care.
,
1992,
Respiratory care.