Impact of blood collection devices on clinical chemistry assays.

Blood collection devices interact with blood to alter blood composition, serum, or plasma fractions and in some cases adversely affect laboratory tests. Vascular access devices may release coating substances and exert shear forces that lyse cells. Blood-dissolving tube additives can affect blood constituent stability and analytical systems. Blood tube stoppers, stopper lubricants, tube walls, surfactants, clot activators, and separator gels may add materials, adsorb blood components, or interact with protein and cellular components. Thus, collection devices can be a major source of preanalytical error in laboratory testing. Device manufacturers, laboratory test vendors, and clinical laboratory personnel must understand these interactions as potential sources of error during preanalytical laboratory testing. Although the effects of endogenous blood substances have received attention, the effects of exogenous substances on assay results have not been well described. This review will identify sources of exogenous substances in blood specimens and propose methods to minimize their impact on clinical chemistry assays.

[1]  R. Langer,et al.  Biodegradable photo-crosslinked poly(ether-ester) networks for lubricious coatings. , 2000, Biomaterials.

[2]  M. Sharp,et al.  Scaling of Hemolysis in Needles and Catheters , 1998, Annals of Biomedical Engineering.

[3]  宮崎 泰彦 High shear stress can initiate both platelet aggregation and shedding of procoagulant containing microparticles , 1999 .

[4]  R. Dean,et al.  Absorption of therapeutic drugs by barrier gels in serum separator blood collection tubes. Volume- and time-dependent reduction in total and free drug concentrations. , 1994, American journal of clinical pathology.

[5]  T. Buclin,et al.  Misleading tacrolimus concentration value in blood taken from a catheter used for tacrolimus administration. , 2008, American journal of health-system pharmacy : AJHP : official journal of the American Society of Health-System Pharmacists.

[6]  J. Mahoney,et al.  Changes in oxygen measurements when whole blood is stored in iced plastic or glass syringes. , 1991, Clinical chemistry.

[7]  J. D. Janssen,et al.  Effects of four different methods of sampling arterial blood and storage time on gas tensions and shunt calculation in the 100% oxygen test. , 1997, The European respiratory journal.

[8]  J. Hicks,et al.  Evaluation of a new blood-collecting device ("microtainer") that is suited for pediatric use. , 1976, Clinical chemistry.

[9]  J. D. Cook,et al.  Benzalkonium interference with test methods for potassium and sodium. , 1990, Clinical chemistry.

[10]  M. Morris,et al.  Incompatibility of diazepam injection in plastic intravenous bags. , 1979, American journal of hospital pharmacy.

[11]  M. Lambert,et al.  Percutaneous venepuncture practice in a large urban teaching hospital. , 2007, Clinical medicine.

[12]  M. Evenson,et al.  Effects of contaminants in blood-collection devices on measurements of therapeutic drugs. , 1983, Clinical chemistry.

[13]  G. Hortin,et al.  Immunoassay interference by a commonly used blood collection tube additive, the organosilicone surfactant silwet L-720. , 2005, Clinical chemistry.

[14]  J. Butler Negative interference in immunoassays. , 1995, Clinical chemistry.

[15]  C. L. Bird,et al.  Measurement of organochlorine levels in postprandial serum or in blood collected in serum separator tubes. , 1996, Cancer epidemiology, biomarkers & prevention : a publication of the American Association for Cancer Research, cosponsored by the American Society of Preventive Oncology.

[16]  J. Krahn,et al.  Lithium heparinised blood-collection tubes give falsely low albumin results with an automated bromcresol green method in haemodialysis patients , 2008, Clinical chemistry and laboratory medicine.

[17]  J. Honour,et al.  Unusual results from immunoassays and the role of the clinical endocrinologist , 2006, Clinical endocrinology.

[18]  J. Egan,et al.  The Role of Incretins in Glucose Homeostasis and Diabetes Treatment , 2008, Pharmacological Reviews.

[19]  C. Niemeyer,et al.  Spuriously raised cyclosporin concentrations drawn through polyurethane central venous catheter , 1998, The Lancet.

[20]  M. Perales,et al.  Radioisotopic assay of aspartate and alanine aminotransferase. , 1992, Clinical biochemistry.

[21]  K. Melber,et al.  Measurement of hematological, clinical chemistry, and infection parameters from hirudinized blood collected in universal blood sampling tubes. , 2001, Seminars in thrombosis and hemostasis.

[22]  J. D. Robinson,et al.  Effect of serum separator blood collection tubes on drug concentrations. , 1983, Therapeutic drug monitoring.

[23]  O. Müller-Plathe,et al.  Stability of Blood Gases, Electrolytes and Haemoglobin in Heparinized Whole Blood Samples: Influence of the Type of Syringe , 1992, European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies.

[24]  A. Lyon,et al.  Evaluation of argatroban and DUP 714 as anticoagulants for blood gas, electrolyte and ionized calcium analyses , 2000, Scandinavian Journal of Clinical and Laboratory Investigation.

[25]  J. Barrett,et al.  Influences of blood sample processing on low-molecular-weight proteome identified by surface-enhanced laser desorption/ionization mass spectrometry. , 2005, Clinical chemistry.

[26]  Jason Y. Park,et al.  Processing controls in blood collection tubes reveals interference. , 2005, Clinical chemistry.

[27]  B. Polack,et al.  Impact of the Blood Collection Tube on the Activation of Coagulation , 1997, Thrombosis and Haemostasis.

[28]  R. Nation,et al.  Leaching of 2-(2-hydroxyethylmercapto) benzothiazole into contents of disposable syringes. , 1981, Journal of pharmaceutical sciences.

[29]  M Plebani,et al.  Hemolyzed specimens: a reason for rejection or a clinical challenge? , 2000, Clinical chemistry.

[30]  Y Komiyama,et al.  High shear stress can initiate both platelet aggregation and shedding of procoagulant containing microparticles. , 1996, Blood.

[31]  K. Adeli,et al.  Heparin interference in whole blood sodium measurements in a pediatric setting. , 2006, Clinical biochemistry.

[32]  W. Mapleson,et al.  Leakage of Oxygen from Blood and Water Samples Stored in Plastic and Glass Syringes , 1971, British medical journal.

[33]  D. O'Kane,et al.  Plastic versus glass tubes: effects on analytical performance of selected serum and plasma hormone assays. , 2004, Clinical chemistry.

[34]  J. Borg,et al.  Multicenter evaluation of a bilayer polymer blood collection tube for coagulation testing: effect on routine hemostasis test results and on plasma levels of coagulation activation markers , 2006, Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis.

[35]  A. Dasgupta,et al.  Time-dependent absorption of therapeutic drugs by the gel of the Greiner Vacuette blood collection tube. , 2000, Therapeutic drug monitoring.

[36]  W. Withold,et al.  Storage of Serum or Whole Blood Samples? Effects of Time and Temperature on 22 Serum Analytes , 1995, European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies.

[37]  P. Tanabe,et al.  Factors affecting the risk of blood bank specimen hemolysis. , 2003, Academic emergency medicine : official journal of the Society for Academic Emergency Medicine.

[38]  G. D'Amico,et al.  High binding of immunoglobulin M kappa rheumatoid factor from type II cryoglobulins to cellular fibronectin: a mechanism for induction of in situ immune complex glomerulonephritis? , 1996, American journal of kidney diseases : the official journal of the National Kidney Foundation.

[39]  J. Smith,et al.  The influence on platelet aggregation of drugs that affect the accumulation of adenosine 3':5'-cyclic monophosphate in platelets. , 1971, The Biochemical journal.

[40]  N. Tietz Fundamentals of Clinical Chemistry , 1970 .

[41]  R. Shulman,et al.  Central venous catheters versus peripheral veins for sampling blood levels of commonly used drugs. , 1998, JPEN. Journal of parenteral and enteral nutrition.

[42]  F. Apple,et al.  Prevention of analytical false-positive increases of cardiac troponin I on the Stratus II analyzer. , 1997, Clinical chemistry.

[43]  H. Liss,et al.  Stability of blood gases in ice and at room temperature. , 1993, Chest.

[44]  J. Toffaletti Use of novel preparations of heparin to eliminate interference in ionized calcium measurements: have all the problems been solved? , 1994, Clinical chemistry.

[45]  W. Waring,et al.  Glycolysis inhibitors negatively bias blood glucose measurements: potential impact on the reported prevalence of diabetes mellitus , 2006, Journal of Clinical Pathology.

[46]  M. Landry,et al.  Use of Plastic Vacutainer Tubes for Quantification of Human Immunodeficiency Virus Type 1 in Blood Specimens , 2001, Journal of Clinical Microbiology.

[47]  V. Shah,et al.  Drug assay interference caused by plasticizer in Vacutainers. , 1982, American journal of hospital pharmacy.

[48]  R. Janknegt,et al.  Do evacuated blood collection tubes interfere with therapeutic drug monitoring? , 1983, Pharmaceutisch Weekblad.

[49]  A. Stanković,et al.  The role of in vitro diagnostic companies in reducing laboratory error , 2007, Clinical chemistry and laboratory medicine.

[50]  D. Shand,et al.  Spuriously low plasma propranolol concentrations resulting from blood collection methods , 1975, Clinical pharmacology and therapeutics.

[51]  G. Lippi,et al.  Influence of the needle bore size used for collecting venous blood samples on routine clinical chemistry testing , 2006, Clinical chemistry and laboratory medicine.

[52]  S. Meites Skin-puncture and blood-collecting technique for infants: update and problems. , 1988, Clinical chemistry.

[53]  Yoon-Sik Lee,et al.  Surface-grafted polystyrene beads with comb-like poly(ethylene glycol) chains: preparation and biological application. , 2004, Macromolecular bioscience.

[54]  P. D'Orazio,et al.  Recommendation for measuring and reporting chloride by ISEs in undiluted serum, plasma or blood: International Federation of Clinical Chemistry and Laboratory Medicine (IFCC): IFCC Scientific Division, Committee on Point of Care Testing and Working Group on Selective Electrodes , 2006, Clinical Chemistry and Laboratory Medicine.

[55]  G. Contant,et al.  Heparin inactivation during blood storage: its prevention by blood collection in citric acid, theophylline, adenosine, dipyridamole-C.T.A.D. mixture. , 1983, Thrombosis research.

[56]  Graham B. I. Scott,et al.  HUPO Plasma Proteome Project specimen collection and handling: Towards the standardization of parameters for plasma proteome samples , 2005, Proteomics.

[57]  Sol F. Green Improving the Preanalytical Process: The Focus on Specimen Quality , 2008 .

[58]  N. Blanckaert,et al.  Evaluation of the BD Vacutainer® PST™ II Blood Collection Tube for special chemistry analytes , 2009, Clinical chemistry and laboratory medicine.

[59]  B. Vinet,et al.  Stability of PO2, PCO2, and pH in fresh blood samples stored in a plastic syringe with low heparin in relation to various blood-gas and hematological parameters. , 1999, Clinical biochemistry.

[60]  V. Wiwanitkit Glass syringes are better than plastic for preserving arterial blood gas for oxygen partial pressure determination: an explanation based on nanomaterial composition , 2006, International journal of nanomedicine.

[61]  F. Eldrige,et al.  Change in oxygen tension of shed blood at various temperature. , 1965, Journal of applied physiology.

[62]  M. Vidaud,et al.  Comparison of single and repeat centrifugation of blood specimens collected in BD evacuated blood collection tubes containing a clot activator for cardiac troponin I assay on the ACCESS analyzer. , 2000, Clinical chemistry.

[63]  M. Cooper,et al.  Recommended "order of draw" for collecting blood specimens into additive-containing tubes. , 1982, Clinical chemistry.

[64]  E. Bruck,et al.  National Committee for Clinical Laboratory Standards. , 1980, Pediatrics.

[65]  A. Stanković,et al.  Elevated serum potassium values: the role of preanalytic variables. , 2004, American journal of clinical pathology.

[66]  P. Mannucci,et al.  Monoclonal Antibodies Directed to the Calcium-Free Conformation of Human Protein S , 1989, Thrombosis and Haemostasis.

[67]  W. Rutten,et al.  Effect of magnesium contamination in evacuated blood collection tubes on the prothrombin time test and ISI calibration using recombinant human thromboplastin and different types of coagulometer. , 2005, Thrombosis research.

[68]  Brenda Eskenazi,et al.  Biological sample collection and processing for molecular epidemiological studies. , 2003, Mutation research.

[69]  Roger L. Lundblad,et al.  Considerations for the Use of Blood Plasma and Serum for Proteomic Analysis , 2003 .

[70]  A. Jungbauer,et al.  Complex formation of a calcium-dependent antibody: a thermodynamical consideration. , 2003, Journal of chromatography. A.

[71]  G. Hortin,et al.  Potential interferences from blood collection tubes in mass spectrometric analyses of serum polypeptides. , 2004, Clinical chemistry.

[72]  C. Kennedy,et al.  A comparison of hemolysis rates using intravenous catheters versus venipuncture tubes for obtaining blood samples. , 1996, Journal of emergency nursing: JEN : official publication of the Emergency Department Nurses Association.

[73]  W. Desmet,et al.  Iodinated contrast media interfere with gel barrier formation in plasma and serum separator tubes. , 2003, Clinical chemistry.

[74]  Raymond Gambino,et al.  Acidification of blood is superior to sodium fluoride alone as an inhibitor of glycolysis. , 2009, Clinical chemistry.

[75]  M. Daves,et al.  Different collection tubes in cardiac biomarkers detection , 2008, Journal of clinical laboratory analysis.

[76]  L. Hause,et al.  Differences between values for plasma and serum in tests performed in the Ektachem 700 XR Analyzer, and evaluation of "plasma separator tubes (PST)". , 1989, Clinical chemistry.

[77]  G. Hortin,et al.  Interference in ionized calcium measurements by heparin salts. , 1994, Clinical chemistry.

[78]  R. Bertina,et al.  Prothrombin Time Ratio Is Reduced by Magnesium Contamination in Evacuated Blood Collection Tubes , 2001, Thrombosis and Haemostasis.

[79]  M. Déchaux,et al.  Anticoagulant-induced preanalytical errors in ionized calcium determination on blood. , 1989, Scandinavian journal of clinical and laboratory investigation.

[80]  A. Lilienfeld,et al.  Observations on an outbreak of infectious hepatitis in Baltimore during 1951. , 1953, American journal of public health and the nation's health.

[81]  B. Bilski Needlestick injuries in nurses--the Poznań study. , 2005, International journal of occupational medicine and environmental health.

[82]  N. Yoshikawa,et al.  Hemolysis in Serum Samples Drawn by Emergency Department Personnel versus Laboratory Phlebotomists , 2002 .

[83]  G. Hortin The MALDI-TOF mass spectrometric view of the plasma proteome and peptidome. , 2006, Clinical chemistry.

[84]  K. Miller,et al.  Availability of diazepam from plastic containers. , 1980, American journal of hospital pharmacy.

[85]  A. Kratz,et al.  A comparison of glass and plastic blood collection tubes for routine and specialized coagulation assays: a comprehensive study. , 2006, Archives of pathology & laboratory medicine.

[86]  P. Vermeersch,et al.  A critical appraisal of current practice in the detection, analysis, and reporting of cryoglobulins. , 2008, Clinical chemistry.

[87]  M. Landt,et al.  Monoject Samplette capillary blood container with serum separator evaluated for collection of specimens for therapeutic drug assays and common clinical-chemical tests. , 1986, Clinical chemistry.

[88]  G. Lippi,et al.  No influence of a butterfly device on routine coagulation assays and D‐dimer measurement , 2005, Journal of thrombosis and haemostasis : JTH.

[89]  J. Jansson,et al.  Comparison of soluble thrombomodulin, von Willebrand factor, tPA/PAI-1 complex, and high-sensitivity CRP concentrations in serum, EDTA plasma, citrated plasma, and acidified citrated plasma (Stabilyte) stored at -70 degrees C for 8-11 years. , 2005, Thrombosis research.

[90]  D. Williams Trace metal determinations in blood obtained in evacuated collection tubes. , 1979, Clinica chimica acta; international journal of clinical chemistry.

[91]  S. Shohet,et al.  Energy metabolism in human erythrocytes. I. Effects of sodium fluoride. , 1971, The Journal of clinical investigation.

[92]  M T Landi,et al.  Sample collection, processing and storage. , 1997, IARC scientific publications.

[93]  G. Hortin,et al.  Evaluation of evacuated blood-collection tubes: effects of three types of polymeric separators on therapeutic drug-monitoring specimens. , 1993, Clinical chemistry.

[94]  D. Bruns,et al.  Stabilization of glucose in blood specimens: mechanism of delay in fluoride inhibition of glycolysis. , 2008, Clinical chemistry.

[95]  W. L. Chiou,et al.  Interactions between drugs and plastic intravenous fluid bags. I. Sorption studies on 17 drugs. , 1974, American journal of hospital pharmacy.

[96]  Gd,et al.  High Total Protein Impairs Appropriate Gel Barrier Formation in Bd Vacutainer Blood Collection Tubes , 2007 .

[97]  A. Kratz,et al.  Comparison of glass K3EDTA versus plastic K2EDTA blood-drawing tubes for complete blood counts, reticulocyte counts, and white blood cell differentials. , 2003, Laboratory hematology : official publication of the International Society for Laboratory Hematology.

[98]  E. Cavalier,et al.  Stability of intact parathyroid hormone in samples from hemodialysis patients. , 2007, Kidney international.

[99]  W. G. Wood,et al.  Immobilisation of Antibodies and Antigens on Macro Solid Phases — A Comparison Between Adsorptive and Covalent Binding. A critical study of macro solid phases for use in immunoassay systems, Part I , 1983, Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie.

[100]  J. Pretto,et al.  Effects of sample storage time, temperature and syringe type on blood gas tensions in samples with high oxygen partial pressures. , 1994, Thorax.

[101]  I. Kupke Evaluation of the polyester gel barrier Microtainer. , 1987, Clinical chemistry.

[102]  R. Johnson,et al.  Source of error on A-aDO2 calculated from blood stored in plastic and glass syringes. , 1997, Journal of applied physiology.

[103]  P. Glendenning,et al.  Parathyroid hormone is more stable in EDTA plasma than in serum. , 2002, Clinical chemistry.

[104]  Yoshito Ikada,et al.  Surface lubrication of polymer films by photoinduced graft polymerization , 1990 .

[105]  S. Narayanan,et al.  The preanalytic phase. An important component of laboratory medicine. , 2000, American journal of clinical pathology.

[106]  A. Dasgupta,et al.  Stability of therapeutic drug measurement in specimens collected in VACUTAINER plastic blood-collection tubes. , 1996, Therapeutic drug monitoring.

[107]  M. Popovsky,et al.  Needlestick injuries in blood collection staff , 1989, Transfusion.

[108]  B. Polack Response to 'Multicenter evaluation of a bilayer polymer blood coagulation tube for coagulation testing: effect on routine hemostasis test results and on plasma levels of coagulation activation markers'. , 2007, Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis.

[109]  B. Polack,et al.  Preanalytical Recommendations of the ‘Groupe d’Etude sur l’Hémostase et la Thrombose’ (GEHT) for Venous Blood Testing in Hemostasis Laboratories , 2001, Pathophysiology of Haemostasis and Thrombosis.

[110]  M. Landt,et al.  A new plastic evacuated tube with plasma separator , 1995, Journal of clinical laboratory analysis.

[111]  J. E. Devine Drug-protein binding interferences caused by the plasticizer TBEP. , 1984, Clinical biochemistry.

[112]  N. Howard-Jones,et al.  A critical study of the origins and early development of hypodermic medication. , 1947, Journal of the history of medicine and allied sciences.

[113]  J. Tanus-Santos,et al.  Differences in both matrix metalloproteinase 9 concentration and zymographic profile between plasma and serum with clot activators are due to the presence of amorphous silica or silicate salts in blood collection devices. , 2008, Analytical biochemistry.

[114]  J. Nosanchuk,et al.  False increases of troponin I attributable to incomplete separation of serum. , 1999, Clinical chemistry.

[115]  B. Altura,et al.  Measurement of ionized magnesium in whole blood, plasma and serum with a new ion-selective electrode in healthy and diseased human subjects. , 1991, Magnesium and trace elements.

[116]  M. Sharp,et al.  Hemolysis in Needleless Connectors for Phlebotomy , 2003, ASAIO journal.

[117]  A. House,et al.  Cardiac troponin T in hemodialyzed patients. , 1998, Clinical chemistry.

[118]  A. Dasgupta,et al.  Stability of Therapeutic Drugs in Serum Collected in Vacutainer Serum Separator Tubes Containing a New Gel (SST II) , 2001, Therapeutic drug monitoring.

[119]  R. Swaminathan,et al.  Effectiveness of sodium fluoride as a preservative of glucose in blood. , 1989, Clinical chemistry.

[120]  S. Hopfer,et al.  Rapid analysis of nickel in serum and whole blood by electrothermal atomic absorption spectrophotometry. , 1984, Annals of clinical and laboratory science.

[121]  S. Bryson,et al.  Therapeutic Monitoring of Cyclosporine following Pediatric Bone Marrow Transplantation: Problems with Sampling from Silicone Central Venous Lines , 1989, DICP : the annals of pharmacotherapy.

[122]  G. Hortin,et al.  Effect of blood collection tubes on total triiodothyronine and other laboratory assays. , 2005, Clinical chemistry.

[123]  C. Schmidt,et al.  Stability of pO 2 , pCO 2 and pH in Heparinized Whole Blood Samples: Influence of Storage Temperature with Regard to Leukocyte Count and Syringe Material , 1992, European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies.

[124]  R. Laessig,et al.  Comparison of plastic vs. glass evacuated serum-separator (SST) blood-drawing tubes for common clinical chemistry determinations. , 1992, Clinical chemistry.

[125]  M. Plebani,et al.  Quality specifications for biochemical markers of myocardial injury. , 2004, Clinica chimica acta; international journal of clinical chemistry.

[126]  T. Ashavaid,et al.  Influence of blood specimen collection method on various preanalytical sample quality indicators , 2008, Indian Journal of Clinical Biochemistry.

[127]  M. Grant,et al.  The effect of blood drawing techniques and equipment on the hemolysis of ED laboratory blood samples. , 2003, Journal of emergency nursing: JEN : official publication of the Emergency Department Nurses Association.

[128]  J. Kao,et al.  Interference caused by the contents of serum separator tubes in the Vitros CRP assay , 2003, Annals of clinical biochemistry.

[129]  E. Salzman,et al.  Cyclic 3′,5′-Adenosine Monophosphate in Human Blood Platelets , 1969, Nature.

[130]  Joseph B Lopez,et al.  Can Fluoride-Oxalate and Sodium Citrate Stabilise Homocysteine Levels after Blood Collection? , 2003, Clinical chemistry and laboratory medicine.

[131]  J. Wiseman,et al.  Evaluation of polyethylene terephthalate for ABO and Rh typing and alloantibody screening , 2000, Transfusion.

[132]  P Carroll Lab reports--only as good as the sample. , 1997, RN.

[133]  C. Gelfand,et al.  Inhibition of intrinsic proteolytic activities moderates preanalytical variability and instability of human plasma. , 2007, Journal of proteome research.

[134]  D. M. Lutomski,et al.  The Effect of Thrombocytosis on Serum Potassium and Phosphorus Concentrations , 1994, The American journal of the medical sciences.

[135]  Mario Plebani,et al.  Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories , 2008, Clinical chemistry and laboratory medicine.

[136]  María Fatás,et al.  Anomalous flotation of separator gel: density or viscosity? , 2008, Clinical chemistry.

[137]  D. Ernst Plastic collection tubes decrease risk of employee injury. , 2001, MLO: medical laboratory observer.

[138]  J. Sheffield,et al.  False positive results for the AUSZYME monoclonal test. , 2005, Obstetrics and gynecology.

[139]  Altheia Roquemore-Goins,et al.  Effects of blood collection tubes, including pediatric devices, on 16 common immunoassays. , 2006, Clinical chemistry.

[140]  Rita Cornelis,et al.  Trace elements in human plasma or serum , 1989 .

[141]  Gilbert S Omenn,et al.  Immunoassay and antibody microarray analysis of the HUPO Plasma Proteome Project reference specimens: Systematic variation between sample types and calibration of mass spectrometry data , 2005, Proteomics.

[142]  H. Neels,et al.  Effects of needle size and storage temperature on measurements of serum potassium. , 1986, Clinical chemistry.

[143]  David E Bruns,et al.  Stabilization of glucose in blood samples: why it matters. , 2009, Clinical chemistry.

[144]  G. Hortin,et al.  Differential effect of blood collection tubes on total free fatty acids (FFA) and total triiodothyronine (TT3) concentration: a model for studying interference from tube constituents. , 2007, Clinica chimica acta; international journal of clinical chemistry.

[145]  Nikolov,et al.  Role of Silicone Surfactant in Flexible Polyurethane Foam. , 1999, Journal of colloid and interface science.

[146]  G. Wickus,et al.  Interference in the Allégro immunoassay system when blood is collected in silicone-coated tubes. , 1992, Clinical chemistry.

[147]  J. Coyle Psychiatric drugs in medical practice. , 1977, The Medical clinics of North America.

[148]  Ana K Stankovic,et al.  Assay interferences from blood collection tubes: a cautionary note. , 2006, Clinical chemistry.

[149]  G. Bray,et al.  Calcium-specific immunoassays for factor IX: reduced levels of antigen in patients with vitamin K disorders. , 1986, The Journal of laboratory and clinical medicine.

[150]  M. Gales,et al.  Comparison of Bayer Advia Centaur® immunoassay results obtained on samples collected in four different Becton Dickinson Vacutainer® tubes , 2006, Annals of clinical biochemistry.

[151]  F. Mannello Serum or plasma samples? The "Cinderella" role of blood collection procedures: preanalytical methodological issues influence the release and activity of circulating matrix metalloproteinases and their tissue inhibitors, hampering diagnostic trueness and leading to misinterpretation. , 2008, Arteriosclerosis, thrombosis, and vascular biology.

[152]  Christina Wang,et al.  Validation of a testosterone and dihydrotestosterone liquid chromatography tandem mass spectrometry assay: Interference and comparison with established methods , 2008, Steroids.

[153]  P. Gorevic,et al.  Evaluation of cryoglobulins. , 1999, Archives of pathology & laboratory medicine.

[154]  P. Desjardins,et al.  Storage of specimens in "Microtainer Amber Tubes" protects bilirubin from degradation by light. , 1987, Clinical chemistry.

[155]  U. Moehrlen,et al.  Evaluation of an improved blood-conserving POCT sampling system. , 2004, Clinical biochemistry.

[156]  E. Sugita,et al.  Interference by butyl rubber stoppers in GLC analysis for theophylline. , 1976, Journal of pharmaceutical sciences.

[157]  Thomas A Neubert,et al.  Sample preparation for serum/plasma profiling and biomarker identification by mass spectrometry , 2006, Journal of Chromatography A.

[158]  J. Savory,et al.  Hemolysis of Specimens Drawn in the ER , 1996 .

[159]  G. Hortin,et al.  Macromolecular chromogenic substrates for measuring proteinase activity. , 2001, Clinical chemistry.

[160]  R. Swaminathan,et al.  Errors due to heparin in the estimation of plasma sodium and potassium concentrations , 2004, Intensive Care Medicine.

[161]  A. Tuinman,et al.  Release of benzalkonium chloride from a heparin-bonded umbilical catheter with resultant factitious hypernatremia and hyperkalemia. , 1991, Pediatrics.

[162]  N Leigh Anderson,et al.  High-abundance polypeptides of the human plasma proteome comprising the top 4 logs of polypeptide abundance. , 2008, Clinical chemistry.

[163]  Edward R. Ashwood,et al.  Tietz Textbook of Clinical Chemistry and Molecular Diagnostics , 2005 .

[164]  H. Katus,et al.  Troponin T and I assays show decreased concentrations in heparin plasma compared with serum: lower recoveries in early than in late phases of myocardial injury. , 2000, Clinical chemistry.

[165]  C. Gauthier,et al.  Haemolyzed samples: responsibility of short catheters. , 1994, Annales de biologie clinique.

[166]  R. Gambino Glucose: a simple molecule that is not simple to quantify. , 2007, Clinical chemistry.

[167]  D. Ernst,et al.  Phlebotomy tools of the trade. , 2002, Home healthcare nurse.

[168]  P. Urban,et al.  Facilitated determination of ionized calcium. , 1985, Clinical chemistry.

[169]  Y. Bergqvist,et al.  Effect of use of gel-barrier sampling tubes on determination of some antiepileptic drugs in serum. , 1984, Clinical chemistry.

[170]  Toshihiro Tanaka,et al.  A Pitfall of Measuring Lithium Levels in Neonates , 2008, Therapeutic drug monitoring.

[171]  H. Tammen,et al.  Peptidomic analysis of human blood specimens: Comparison between plasma specimens and serum by differential peptide display , 2005, Proteomics.

[172]  A. Storrow,et al.  Nalmefene and the urine opiate screen. , 1998, Clinical chemistry.

[173]  D. Yücel,et al.  Falsely increased free triiodothyronine in sera stored in serum separator tubes. , 2002, Clinical chemistry.

[174]  G. Rodgers,et al.  A Comparison of Blood Collection in Glass Versus Plastic Vacutainers on Results of Esoteric Coagulation Assays , 2003 .

[175]  M. Sampson,et al.  Positive interference in lithium determinations from clot activator in collection container. , 1997, Clinical chemistry.

[176]  B. Befeler,et al.  Influence of syringe materials on arterial blood gas determinations. , 1974, Chest.

[177]  G. Dimeski,et al.  Magnesium contamination from Terumo blood collection tubes. , 2006, Clinical chemistry.

[178]  J. Spann,et al.  Blood collection techniques, heparin and quinidine protein binding , 1979, Clinical pharmacology and therapeutics.

[179]  J. Ladenson,et al.  Serum versus heparinized plasma for eighteen common chemistry tests: is serum the appropriate specimen? , 1974, American journal of clinical pathology.

[180]  H. Busscher,et al.  On the influence of flow conditions and wettability on blood material interactions. , 2003, Biomaterials.

[181]  S. Meites,et al.  Skin-puncture and blood-collecting techniques for infants. , 1979, Clinical chemistry.

[182]  A. Grañena,et al.  Influence of different indwelling lines on the measurement of blood cyclosporin A levels. , 1988, Bone marrow transplantation.

[183]  D. Yücel,et al.  Effect of in vitro hemolysis on 25 common biochemical tests. , 1992, Clinical chemistry.

[184]  G. Lippi,et al.  Preanalytical variability in laboratory testing: influence of the blood drawing technique , 2005, Clinical chemistry and laboratory medicine.

[185]  Ernst Dj Plastic collection tubes decrease risk of employee injury. , 2001 .

[186]  G. J. Wilcox,et al.  Does transfusion using a syringe infusion pump and small‐gauge needle cause hemolysis? , 1981, Transfusion.

[187]  H. Offenbacher,et al.  Measurement of ionized magnesium with neutral carrier based ISE's. Progress and results with the AVL 988-4 magnesium analyzer. , 1994, Scandinavian journal of clinical and laboratory investigation. Supplementum.

[188]  M C Davies,et al.  Uptake of drugs by catheters: the influence of the drug molecule on sorption by polyurethane catheters. , 1996, Biomaterials.

[189]  F. Sedor,et al.  Stability of plasma lactate in vitro in the presence of antiglycolytic agents. , 1994, Clinical chemistry.

[190]  Graham B. I. Scott,et al.  Artifactual isoform profile modification following treatment of human plasma or serum with protease inhibitor, monitored by 2-dimensional electrophoresis and mass spectrometry. , 2005, BioTechniques.

[191]  V. Murthy Unusual interference from primary collection tube in a high‐performance liquid chromatography assay of amiodarone , 1997, Journal of clinical laboratory analysis.

[192]  D. Haverstick,et al.  Specimen Collection and Processing , 2012 .

[193]  A. Dasgupta,et al.  Comparison of BD Vacutainer SST Plus Tubes with BD SST II Plus Tubes for common analytes. , 2001, Clinica chimica acta; international journal of clinical chemistry.

[194]  E. van Meegen,et al.  Effect of Evacuated Blood Collection Tubes on Thromboplastin Calibration , 1998, Thrombosis and Haemostasis.

[195]  G. Rodgers,et al.  Comparison of five thrombin time reagents. , 2003, Clinical chemistry.

[196]  O. G. Nilsen,et al.  Plasma protein binding of basic drugs. I. Selective displacement from alpha 1-acid glycoprotein by tris(2-butoxyethyl) phosphate. , 1977, Clinical pharmacology and therapeutics.

[197]  T. Koch,et al.  Suitability of Collection Tubes with Separator Gels for Therapeutic Drug Monitoring , 1990, Therapeutic drug monitoring.

[198]  G. Racz,et al.  A Comparative Study of Plastic (Polypropylene) and Glass Syringes in Blood‐Gas Analysis , 1972, Anesthesia and Analgesia.

[199]  F. Douce,et al.  Effects of syringe material, sample storage time, and temperature on blood gases and oxygen saturation in arterialized human blood samples. , 2006, Respiratory care.

[200]  R. Vyzula,et al.  Surface-enhanced laser desorption ionization/time-of-flight mass spectrometry reveals significant artifacts in serum obtained from clot activator-containing collection devices. , 2006, Clinical chemistry.

[201]  J. Livesey,et al.  Effect of anticoagulants and storage temperatures on stability of plasma and serum hormones. , 2001, Clinical biochemistry.

[202]  R. Elin,et al.  Evaluation of AVL988/4 analyzer for measurement of ionized magnesium and ionized calcium. , 2001, Scandinavian journal of clinical and laboratory investigation.

[203]  M. Parviainen,et al.  Suitability of Collection Tubes with Separator Gels for Collecting and Storing Blood Samples for Therapeutic Drug Monitoring (TDM) , 2000, Clinical chemistry and laboratory medicine.

[204]  James J. Miller,et al.  Interference in immunoassays: avoiding erroneous results , 2004 .

[205]  J. Dale,et al.  Specimen collection volumes for laboratory tests. , 2009, Archives of pathology & laboratory medicine.

[206]  Greg Ward,et al.  Interferences in immunoassay. , 2004, The Clinical biochemist. Reviews.

[207]  R. Miniero,et al.  Monitoring of Cyclosporine Blood Levels From Central Venous Lines: A Misleading Assay? , 1994, Therapeutic drug monitoring.

[208]  P. Kierulf,et al.  Binding and Displacement of Basic, Acidic and Neutral Drugs in Normal and Orosomucoid-deficient Plasma , 1981, Clinical pharmacokinetics.

[209]  Francesco M Marincola,et al.  Effects of storage time and exogenous protease inhibitors on plasma protein levels. , 2006, American journal of clinical pathology.

[210]  R. Laessig,et al.  The effects of 0.1 and 1.0 per cent erythrocytes and hemolysis on serum chemistry values. , 1976, American journal of clinical pathology.

[211]  G. Lippi,et al.  Quality and reliability of routine coagulation testing: can we trust that sample? , 2006, Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis.

[212]  F. Pazzucconi,et al.  Measurement of ionized magnesium with AVL 988/4 electrolyte analyzer: preliminary analytical and clinical results. , 1996, Scandinavian journal of clinical and laboratory investigation. Supplementum.

[213]  P. Houillier,et al.  Variation in serum and plasma PTH levels in second-generation assays in hemodialysis patients: a cross-sectional study. , 2008, American journal of kidney diseases : the official journal of the National Kidney Foundation.

[214]  A. Sheel,et al.  Performance of evacuated blood collection tubes at high altitude. , 2008, High Altitude Medicine & Biology.

[215]  Kenneth E Blick,et al.  Stability studies of twenty-four analytes in human plasma and serum. , 2002, Clinical chemistry.

[216]  V. Shah,et al.  Interference with measurements of certain drugs in plasma by a plasticizer in vacutainer tubes. , 1982, Clinical chemistry.

[217]  D. Wild The Immunoassay Handbook , 2001 .

[218]  Hirudin as an anticoagulant for both haematology and chemistry tests , 2000, Journal of automated methods & management in chemistry.

[219]  C. Feng,et al.  The Plastic Tube Compared with the Glass Tube in Blood Bank Tests , 1988, Pathology.

[220]  L. Valentino,et al.  Needles and needleless devices for infusion of anti‐haemophilic factor concentrate: impact on protein structure and function , 2006, Haemophilia : the official journal of the World Federation of Hemophilia.

[221]  G. Lippi,et al.  The role of ethylenediamine tetraacetic acid (EDTA) as in vitro anticoagulant for diagnostic purposes , 2007, Clinical chemistry and laboratory medicine.

[222]  O. Borgå,et al.  Plasma protein binding of basic drugs , 1977 .

[223]  D. Grucker,et al.  Insulin measurements in haemolysed serum: influence of insulinase inhibitors. , 1998, Clinica chimica acta; international journal of clinical chemistry.

[224]  Michael V Sefton,et al.  Biomaterial-associated thrombosis: roles of coagulation factors, complement, platelets and leukocytes. , 2004, Biomaterials.

[225]  R. Flanagan,et al.  Ethylbenzene and xylene from Sarstedt Monovette serum gel blood-collection tubes. , 1993, Clinical Chemistry.

[226]  G. Papatheodoridis,et al.  Pseudohyperkalemia in serum: the phenomenon and its clinical magnitude. , 2006, The Journal of laboratory and clinical medicine.

[227]  C. Roe,et al.  Importance of blood-collection tubes in plasma lidocaine determinations. , 1979, Clinical chemistry.

[228]  A. D. Leener,et al.  Evaluation of a new generation of plastic evacuated blood-collection tubes in clinical chemistry, therapeutic drug monitoring, hormone and trace metal analysis , 2004, Clinical chemistry and laboratory medicine.