Validation of a Clinical Shack-Hartmann Aberrometer

Purpose. To validate the accuracy, tolerance, and repeatability of the complete ophthalmic analysis system aberrometer (COAS, Wavefront Sciences Inc.) with model eyes and normal human eyes. Method. Model eyes were constructed from six polymethyl methacrylate, single-surface lenses with known characteristics. Accuracy of second-order aberrations was verified by measuring defocus and astigmatism induced by series of spherical and cylindrical trial lenses. Accuracy of higher-order aberrations was evaluated by comparing ray-tracing predictions with measured spherical aberration and coma of the aspheric model eyes. Tolerance to axial and lateral misalignment was measured by controlled displacements of the model eye relative to the aberrometer. Repeatability was tested on the same model eyes with repeated measurements taken within 1 s or within half an hour with realignment between each trial. Analyses were based on a 5-mm pupil diameter. Results. Defocus and astigmatism were accurately measured within the working range of the instrument automatic focus adjustment (e.g., measured defocus was within ±0.25 diopters over a −6.50 to +3.00 D range of refractive error). Accuracy of spherical aberration and coma agreed closely with theoretical predictions (e.g., for all six aspheric models, the mean absolute difference between predicted and measured Z40 was 0.007 &mgr;m). Axial displacements over the range ±2.5 mm had little effect on measurements for myopic and emmetropic model eyes. Also, lateral displacements over the range ±1.5 mm did not produce significant coma. The standard deviations of repeated measurements of higher-order root mean square on model eyes were <1% of the mean with repeated measures within 1 s and 10% of the mean for five individual measurements with realignment in between each. Tolerance to small lateral displacements was also observed for human eyes. Conclusion. The complete ophthalmic analysis system aberrometer can measure second-, third-, and fourth-order aberrations accurately and repeatedly on model eyes.

[1]  A. Bradley,et al.  Relationship between Refractive Error and Monochromatic Aberrations of the Eye , 2003, Optometry and vision science : official publication of the American Academy of Optometry.

[2]  Comparison of Monochromatic Ocular Aberrations Measured with an Objective Cross-Cylinder Aberroscope and a Shack-Hartmann Aberrometer , 2003, Optometry and vision science : official publication of the American Academy of Optometry.

[3]  Junzhong Liang,et al.  Aberrations and retinal image quality of the normal human eye. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[4]  Habib Hamam,et al.  Objective Measurement of Optical Aberrations in Myopic Eyes , 2002, Optometry and vision science : official publication of the American Academy of Optometry.

[5]  L N Thibos,et al.  Principles of Hartmann-Shack aberrometry. , 2000, Journal of refractive surgery.

[6]  Z. Nagy,et al.  First results with wavefront-guided photorefractive keratectomy for hyperopia. , 2002, Journal of refractive surgery.

[7]  D A Goss,et al.  Reliability of refraction--a literature review. , 1996, Journal of the American Optometric Association.

[8]  A. Bradley,et al.  Statistical variation of aberration structure and image quality in a normal population of healthy eyes. , 2002, Journal of the Optical Society of America. A, Optics, image science, and vision.

[9]  Norberto López-Gil,et al.  Aberration generation by contact lenses with aspheric and asymmetric surfaces. , 2002, Journal of refractive surgery.

[10]  Takashi Fujikado,et al.  Wavefront analysis of higher‐order aberrations in patients with cataract , 2002, Journal of cataract and refractive surgery.

[11]  Jochen Straub,et al.  Higher order aberrations in normal, dilated, intraocular lens, and laser in situ keratomileusis corneas. , 2002, Journal of refractive surgery.

[12]  A Bradley,et al.  Comparison of the eye's wave-front aberration measured psychophysically and with the Shack-Hartmann wave-front sensor. , 1998, Journal of the Optical Society of America. A, Optics, image science, and vision.

[13]  Arthur Bradley,et al.  Test-retest reliability of clinical Shack-Hartmann measurements. , 2004, Investigative ophthalmology & visual science.

[14]  D. Williams,et al.  Monochromatic aberrations of the human eye in a large population. , 2001, Journal of the Optical Society of America. A, Optics, image science, and vision.

[15]  L. Thibos,et al.  Standards for reporting the optical aberrations of eyes. , 2002, Journal of refractive surgery.

[16]  W. Wesemann,et al.  Automatic infrared refractors--a comparative study. , 1987, American journal of optometry and physiological optics.

[17]  Longitudinal evaluation of optical aberrations following laser in situ keratomileusis surgery. , 2000, Journal of refractive surgery.

[18]  L N Thibos,et al.  Clinical applications of the Shack-Hartmann aberrometer. , 1999, Optometry and vision science : official publication of the American Academy of Optometry.

[19]  Takashi Fujikado,et al.  Effect of tear film break-up on higher-order aberrations measured with wavefront sensor. , 2002, American journal of ophthalmology.

[20]  Austin Roorda,et al.  Applying wavefront sensors and corneal topography to keratoconus. , 2002, Biomedical sciences instrumentation.

[21]  D R Williams,et al.  Effect of rotation and translation on the expected benefit of an ideal method to correct the eye's higher-order aberrations. , 2001, Journal of the Optical Society of America. A, Optics, image science, and vision.

[22]  X. Hong,et al.  Use of a Shack-Hartmann Aberrometer to Assess the Optical Outcome of Corneal Transplantation in a Keratoconic Eye , 2001, Optometry and vision science : official publication of the American Academy of Optometry.

[23]  R E Fusaro,et al.  The Repeatability of Automated and Clinician Refraction , 1998, Optometry and vision science : official publication of the American Academy of Optometry.

[24]  Z. Nagy,et al.  Wavefront-guided photorefractive keratectomy for myopia and myopic astigmatism. , 2002, Journal of refractive surgery.

[25]  Junzhong Liang,et al.  Objective measurement of wave aberrations of the human eye with the use of a Hartmann-Shack wave-front sensor. , 1994, Journal of the Optical Society of America. A, Optics, image science, and vision.

[26]  Norberto López-Gil,et al.  Ocular wave-front aberration statistics in a normal young population , 2002, Vision Research.

[27]  S A Klein,et al.  Optimal corneal ablation for eyes with arbitrary Hartmann-Shack aberrations. , 1998, Journal of the Optical Society of America. A, Optics, image science, and vision.

[28]  David R Williams,et al.  Calculated impact of higher-order monochromatic aberrations on retinal image quality in a population of human eyes. , 2002, Journal of the Optical Society of America. A, Optics, image science, and vision.

[29]  Nikole L. Himebaugh,et al.  On-Eye Evaluation of Optical Performance of Rigid and Soft Contact Lenses , 2001, Optometry and vision science : official publication of the American Academy of Optometry.