A metric for testing the accuracy of cross-modality image registration: validation and application.

PURPOSE Our goals were to (a) develop and validate a sensitive measure of image registration, applicable between as well as within imaging modalities; and (b) quantify the accuracy of the automated image registration (AIR) algorithm in retrospectively registering MR and PET images of baboon brain. METHOD We studied five monkeys, each with a surgically implanted "cap" that was bolted to an acrylic headholder containing fiducial markers. Anatomic MRI and PET H2 15O blood flow images were aligned using AIR (ignoring the fiducials). RESULTS (a) Fiducial points were localized to about one-tenth the voxel size. Distances computed from the fiducial points were correct to within 0.2% (MRI) and 1.4% (PET). (b) The mean error remaining after AIR, measured at the five fiducial points, varied from 2.9 to 5.3 mm. The average registration error within the brain was calculated to be 2.20 mm (maximum, 3.57 mm). This error was not primarily rotational about the center of the brain and was worsened by heavy smoothing of the PET images. CONCLUSION (a) We have defined a reliable, sensitive metric that can stringently test the accuracy of various image registration techniques. (b) The AIR algorithm leaves modest error after aligning PET blood flow and anatomic MR images of baboon brain.

[1]  H. Damasio,et al.  A new technique for pet slice orientation and MRI‐PET coregistration , 1994 .

[2]  Richard S. J. Frackowiak,et al.  Area V5 of the human brain: evidence from a combined study using positron emission tomography and magnetic resonance imaging. , 1993, Cerebral cortex.

[3]  T J Spinks,et al.  Physical performance of a positron tomograph for brain imaging with retractable septa. , 1992, Physics in medicine and biology.

[4]  I Kapouleas,et al.  Registration of three-dimensional MR and PET images of the human brain without markers. , 1991, Radiology.

[5]  W T Yuh,et al.  Neoplastic vs Inflammatory Meningeal Enhancement with Gd‐DTPA , 1990, Journal of computer assisted tomography.

[6]  W. A. Hanson,et al.  Interactive 3D segmentation of MRI and CT volumes using morphological operations. , 1992, Journal of computer assisted tomography.

[7]  Max A. Viergever,et al.  Matching and integrated display of brain images from multiple modalities , 1995, Medical Imaging.

[8]  C. Tamminga,et al.  Positron emission tomography evaluation of dopaminergic pathways , 1995 .

[9]  J. Mazziotta,et al.  MRI‐PET Registration with Automated Algorithm , 1993, Journal of computer assisted tomography.

[10]  S. Buchholz,et al.  PET Measured Evoked Cerebral Blood Flow Responses in an Awake Monkey , 1991, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[11]  I. Perkash,et al.  Registration error quantification of a surface-based multimodality image fusion system. , 1995, Medical physics.

[12]  R A Robb,et al.  Interactive display and analysis of 3-D medical images. , 1989, IEEE transactions on medical imaging.

[13]  S. Strother,et al.  Quantitative Comparisons of Image Registration Techniques Based on High‐Resolution MRI of the Brain , 1994, Journal of computer assisted tomography.

[14]  J. Mazziotta,et al.  Rapid Automated Algorithm for Aligning and Reslicing PET Images , 1992, Journal of computer assisted tomography.