Comparing Biomechanical Properties, Repair Times, and Value of Common Core Flexor Tendon Repairs

Background: The aim of the study was to compare biomechanical strength, repair times, and repair values for zone II core flexor tendon repairs. Methods: A total of 75 fresh-frozen human cadaveric flexor tendons were harvested from the index through small finger and randomized into one of 5 repair groups: 4-stranded cross-stitch cruciate (4-0 polyester and 4-0 braided suture), 4-stranded double Pennington (2-0 knotless barbed suture), 4-stranded Pennington (4-0 double-stranded braided suture), and 6-stranded modified Lim-Tsai (4-0 looped braided suture). Repairs were measured in situ and their repair times were measured. Tendons were linearly loaded to failure and multiple biomechanical values were measured. The repair value was calculated based on operating room costs, repair times, and suture costs. Analysis of variance (ANOVA) and Tukey post hoc statistical analysis were used to compare repair data. Results: The braided cruciate was the strongest repair (P > .05) but the slowest (P > .05), and the 4-stranded Pennington using double-stranded suture was the fastest (P > .05) to perform. The total repair value was the highest for braided cruciate (P > .05) compared with all other repairs. Barbed suture did not outperform any repairs in any categories. Conclusions: The braided cruciate was the strongest of the tested flexor tendon repairs. The 2-mm gapping and maximum load to failure for this repair approached similar historical strength of other 6- and 8-stranded repairs. In this study, suture cost was negligible in the overall repair cost and should be not a determining factor in choosing a repair.

[1]  Aniruddh N. Nayak,et al.  A Mechanical Evaluation of Zone II Flexor Tendon Repair Using a Knotless Barbed Suture Versus a Traditional Braided Suture. , 2015, The Journal of hand surgery.

[2]  Oded Ben-Amotz,et al.  The effect of barbed suture tendon repair on work of flexion. , 2015, The Journal of hand surgery.

[3]  G. Siegal,et al.  Flexor tendon repair with a knotless, bidirectional barbed suture: an in vivo biomechanical analysis. , 2015, The Journal of hand surgery.

[4]  M. Jordan,et al.  Biomechanical analysis of flexor tendon repair using knotted Kessler and Bunnell techniques and the knotless Bunnell technique. , 2015, The Journal of hand surgery.

[5]  Aakash Chauhan,et al.  Flexor tendon repairs: techniques, eponyms, and evidence. , 2014, The Journal of hand surgery.

[6]  R. Gelberman,et al.  The effect of suture caliber and number of core suture strands on zone II flexor tendon repair: a study in human cadavers. , 2014, The Journal of hand surgery.

[7]  F. O'Brien,et al.  Flexor tendon repair: a comparative study between a knotless barbed suture repair and a traditional four-strand monofilament suture repair , 2014, The Journal of hand surgery, European volume.

[8]  J. Gordon,et al.  Comparison of a multifilament stainless steel suture with FiberWire for flexor tendon repairs — an in vitro biomechanical study , 2013, The Journal of hand surgery, European volume.

[9]  J. Gordon,et al.  Analysis of a knotless flexor tendon repair using a multifilament stainless steel cable-crimp system. , 2013, The Journal of hand surgery.

[10]  K. Chung,et al.  History and evolution of the Kessler repair. , 2013, The Journal of hand surgery.

[11]  Matthew J. Silva,et al.  Intrasynovial flexor tendon repair: A biomechanical study of variations in suture application in human cadavera , 2012, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[12]  J. B. Tang,et al.  Flexor Tendon Repair With a Knotless Barbed Suture: A Comparative Biomechanical Study , 2012 .

[13]  D. Ruch,et al.  A Knotless Flexor Tendon Repair Technique Using a Bidirectional Barbed Suture: An Ex Vivo Comparison of Three Methods , 2011, Plastic and reconstructive surgery.

[14]  R. Meffert,et al.  Biomechanical evaluation of flexor tendon repair using barbed suture material: a comparative ex vivo study. , 2011, The Journal of hand surgery.

[15]  S. Davison,et al.  Barbed Suture Tenorrhaphy: An Ex Vivo Biomechanical Analysis , 2009, Plastic and reconstructive surgery.

[16]  I. Trail,et al.  Forces Transmitted Along Human Flexor Tendons During Passive and Active Movements of the Fingers , 2004, Journal of hand surgery.

[17]  J. W. Strickland Development of flexor tendon surgery: twenty-five years of progress. , 2000, The Journal of hand surgery.

[18]  S. Wolfe,et al.  Biomechanical analysis of the cruciate four-strand flexor tendon repair. , 1999, The Journal of hand surgery.

[19]  S. Woo,et al.  The effects of multiple-strand suture methods on the strength and excursion of repaired intrasynovial flexor tendons: a biomechanical study in dogs. , 1998, The Journal of hand surgery.