Previous work has shown that an automated build-up method, called the Reference Step Method (RSM), can measure the gain ratios of a dc voltage source with an uncertainty of 0.2 μV/V up to 1000 V. We have extended this method to also establish the gain ratio of the digital voltmeter (DVM) used in the build-up method. This only requires a few additional measurements so that the gain ratios of the source and meter can be measured simultaneously. Initial measurements suggest that comparable uncertainties can also be achieved for the DVM.
[1] M. Early,et al. The reference step method for DC voltage at 1 mV and 10 mV , 2014, 29th Conference on Precision Electromagnetic Measurements (CPEM 2014).
[2] L.A. Christian. A new method for measuring ratios of dc voltages from 10 V to 1100 V , 2000, Conference on Precision Electromagnetic Measurements. Conference Digest. CPEM 2000 (Cat. No.00CH37031).