Pseudohypertension and the measurement of blood pressure.

Riva-Rocci Korotkoff (RRK) blood pressure measurements may overestimate the intra-arterial pressure (IAP) in individual patients. To study pseudohypertension, defined as an overestimation larger than 10 mm Hg, we compared RRK and IAP measurements in 76 patients. These patients were considered to be at risk for pseudohypertension because of high age, hypertension, or vascular disease. RRK measurements underestimated simultaneously measured systolic IAP values by 6.0 +/- 6.5 (mean +/- SD) mm Hg, whereas simultaneous diastolic RRK readings overestimated the IAP by 1.9 +/- 5.6 mm Hg. Diastolic overestimation increased slightly with age. Vascular rigidity, as measured by counterpressure plethysmography, did not correlate with these errors. Systolic and diastolic pseudohypertension was observed in two and five patients, respectively. Pseudohypertension was only present in the group in which IAP was measured in the aorta. However, the number of patients with systolic and diastolic pseudohypertension could be increased to three and seven by using the average intra-arterial systolic and diastolic pressure during an RRK measurement or to six and 11 by using the IAP during a control period just before an RRK measurement rather than the IAP at the moment of appearance or disappearance of the Korotkoff sounds. The widely diverging prevalence of pseudohypertension described in the literature might be explained because of such different comparison techniques.

[1]  A. Sacks Indirect Blood Pressure Measurements: A Matter of Interpretation , 1979, Angiology.

[2]  A. Treloar,et al.  Circulation A Journal of the American-Heart Association Comparison of Indirect and Direct Methods of Measuring Arterial Blood Pressure , 2005 .

[3]  R M Gardner,et al.  Direct blood pressure measurement--dynamic response requirements. , 1981, Anesthesiology.

[4]  J. Spence,et al.  Blood pressure measurement in the elderly: correlation of arterial stiffness with difference between intra-arterial and cuff pressures. , 1985, Journal of hypertension.

[5]  E. O'Brien,et al.  Accuracy of indirect blood pressure measurement in the elderly. , 1983, British medical journal.

[6]  N. Schork,et al.  Artefacts in measurement of blood pressure and lack of target organ involvement in the assessment of patients with treatment-resistant hypertension. , 1990, Annals of internal medicine.

[7]  Comparison of direct and indirect arterial pressure measurements in hospitalized patients. , 1974, Australian and New Zealand journal of medicine.

[8]  J. Spence,et al.  Direct, indirect and mean blood pressures in hypertensive patients: the problem of cuff artefact due to arterial wall stiffness, and a partial solution. , 1979, Clinical and investigative medicine. Medecine clinique et experimentale.

[9]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[10]  H. L. Poulsen,et al.  Accuracy of auscultatory blood pressure measurements in hypertensive and obese subjects. , 1983, Hypertension.

[11]  Observer error in systolic blood pressure measurement in the elderly. A case for automatic recorders? , 1986, Archives of internal medicine.

[12]  A. Folsom,et al.  The random-zero versus the standard mercury sphygmomanometer: a systematic blood pressure difference. , 1985, American journal of epidemiology.

[13]  A. C. Burton,et al.  Recommendations for Human Blood Pressure Determination By Sphygmomanometers , 1981, Circulation.

[14]  J. Feussner,et al.  Screening for pseudohypertension. A quantitative, noninvasive approach. , 1988, Archives of internal medicine.

[15]  J. Laragh,et al.  Characterization of Auscultatory Gaps With Wideband External Pulse Recording , 1991, Hypertension.

[16]  K. H. Wesseling,et al.  Finapres, continuous noninvasive finger arterial pressure based on the method of Peñáz , 1990 .

[17]  J. Feussner,et al.  Observer Error in Systolic Blood Pressure Measurement in the Elderly , 1987 .

[18]  M. O'Rourke,et al.  Comparison of Direct and Indirect Arterial Pressure Measurements in Hospitalised Patients , 1974 .

[19]  S. London,et al.  Critique of indirect diastolic end point. "Muffling" vs "Last" Sound. , 1967, Archives of internal medicine.

[20]  J. Chalmers,et al.  Optimal size of cuff bladder for indirect measurement of arterial pressure in adults. , 1989, Journal of hypertension.

[21]  R. Grover,et al.  EFFECT OF SIZE OF CUFF BLADDER ON ACCURACY OF MEASUREMENT OF INDIRECT BLOOD PRESSURE. , 1965, American heart journal.

[22]  Pseudohypertension in the elderly. , 1990, Journal of hypertension.

[23]  A. J. Dunning,et al.  Accuracy of auscultatory blood pressure measurement with a long cuff. , 1987, British medical journal.

[24]  S N Hunyor,et al.  Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings. , 1978, British medical journal.

[25]  F. Mee,et al.  Inaccuracy of the Hawksley random zero sphygmomanometer , 1990, The Lancet.

[26]  J. Staessen,et al.  Isolated systolic hypertension in the elderly. , 1990, Journal of hypertension.

[27]  A. Dyer,et al.  A comparison of the random-zero and standard mercury sphygmomanometers. , 1988, Hypertension.

[28]  Manning,et al.  A Comparison of Direct and Indirect Blood‐Pressure Determinations , 1953, Circulation.

[29]  A. Simon,et al.  Vasodilatation of small and large arteries in hypertension. , 1985, Journal of cardiovascular pharmacology.

[30]  K. Wesseling,et al.  Forearm arterial pressure-volume relationships in man. , 1988, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.

[31]  H. Smulyan,et al.  Systolic hypertension. Direct and indirect BP measurements. , 1983, Archives of internal medicine.

[32]  M. O'Rourke,et al.  Arterial stiffness, systolic blood pressure, and logical treatment of arterial hypertension. , 1990, Hypertension.

[33]  R. Gifford An Algorithm for the Management of Resistant Hypertension , 1988, Hypertension.

[34]  I. Mackay,et al.  DEMONSTRATION OF THE IN VIVO COMPRESSIBILITY OF THE BRACHIAL ARTERY IN MAN. , 1963, American heart journal.

[35]  H. Ventura,et al.  Osler's maneuver and pseudohypertension. , 1985, The New England journal of medicine.

[36]  R. Collins,et al.  Blood pressure, stroke, and coronary heart disease Part 2, short-term reductions in blood pressure: overview of randomised drug trials in their epidemiological context , 1990, The Lancet.

[37]  R. Collins,et al.  Blood pressure, stroke, and coronary heart disease Part 1, prolonged differences in blood pressure: prospective observational studies corrected for the regression dilution bias , 1990, The Lancet.