On the implementation of a recently proposed dosimetric formalism to a robotic radiosurgery system.

PURPOSE The aim of this work is to implement a recently proposed dosimetric formalism for nonstandard fields to the calibration and small field output factor measurement of a robotic stereotactic radiosurgery system. METHODS Reference dosimetry measurements were performed in the nonstandard, 60 mm diameter machine specific reference (msr) field using a Farmer ion chamber, five other cylindrical chambers with cavity lengths ranging from 16.25 down to 2.7 mm, and alanine dosimeters. Output factor measurements were performed for the 5, 7.5, 10, and 15 mm field sizes using microchambers, diode detectors, alanine dosimeters, TLD microcubes, and EBT Gafchromic films. Measurement correction factors as described in the proposed formalism were calculated for the ion chamber and diode detector output factor measurements based on published Monte Carlo data. Corresponding volume averaging correction factors were calculated for the alanine output factor measurements using 3D dose distributions, measured with polymer gel dosimeters. RESULTS Farmer chamber and alanine reference dosimetry results were found in close agreement, yielding a correction factor of k(Q(msr),Q)(f(msr),f(ref)) = 0.999 +/- 0.016 for the chamber readings. These results were also found to be in agreement within experimental uncertainties with corresponding results obtained using the shorter cavity length ionization chambers. The mean measured dose values of the latter, however, were found to be consistently greater than that of the Farmer chamber. This finding, combined with an observed inverse relationship between the mean measured dose and chamber cavity length that follows the trend predicted by theoretical volume averaging calculations in the msr field, implies that the Farmer k(Q(msr),Q)(f(msr),f(ref)) correction is greater than unity. Regarding the output factor results, deviations as large as 33% were observed between the different dosimeters used. These deviations were substantially decreased when appropriate correction factors were applied to the measured microchamber, diode, and alanine values. After correction, all diode and microchamber measured output factors agreed within 1.6% with the corresponding alanine measurements, and within 3.1% with the TLD measurements. The weighted mean output factors were 0.681 +/- 0.001, 0.824 +/- 0.001, 0.875 +/- 0.001, and 0.954 +/- 0.001 for the 5, 7.5, 10, and 15 mm beams, respectively. CONCLUSIONS The comparison of Farmer chamber measurements versus alanine reference dosimetry validates the use of the former for dosimetry in the msr field of this treatment delivery system. The corresponding results of this work obtained using chambers with different cavity lengths, combined with previous literature findings, suggest that a k(Q(msr),Q)(f(msr),f(ref)) Farmer chamber dose response correction factor of 1.01 may improve calibration measurement accuracy when using the proposed dosimetric formalism. The k(Q(msr),Q)(f(msr),f(ref)) correction factor is within 0.5% from unity for ion chambers with cavity lengths less than 10 mm. Substantial improvements in small field output factor measurement accuracy can be obtained when using microchambers and diodes by applying appropriately calculated correction factors to the detector measurements according to the proposed dosimetric formalism, and their routine use is therefore recommended.

[1]  Carlo Cavedon,et al.  Application of a Monte Carlo‐based method for total scatter factors of small beams to new solid state micro‐detectors , 2009, Journal of applied clinical medical physics.

[2]  M Stasi,et al.  The behavior of several microionization chambers in small intensity modulated radiotherapy fields. , 2004, Medical physics.

[3]  J. Seuntjens,et al.  A new formalism for reference dosimetry of small and nonstandard fields. , 2008, Medical physics.

[4]  Robert Jeraj,et al.  Dose calibration of nonconventional treatment systems applied to helical tomotherapy. , 2005, Medical physics.

[5]  Carlo Cavedon,et al.  Total scatter factors of small beams: a multidetector and Monte Carlo study. , 2008, Medical physics.

[6]  P. Andreo,et al.  Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water , 2001 .

[7]  R. Drzymala,et al.  Calibration of the Gamma Knife using a new phantom following the AAPM TG51 and TG21 protocols. , 2008, Medical physics.

[8]  Christos Antypas,et al.  Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system , 2008, Physics in medicine and biology.

[9]  Fujio Araki,et al.  Monte Carlo study of a Cyberknife stereotactic radiosurgery system. , 2006, Medical physics.

[10]  C Antypas,et al.  Dosimetric characterization of CyberKnife radiosurgical photon beams using polymer gels. , 2008, Medical physics.

[11]  The determination of beam quality correction factors: Monte Carlo simulations and measurements. , 2009, Physics in medicine and biology.

[12]  W. Simon,et al.  Total scatter factors and tissue maximum ratios for small radiosurgery fields: comparison of diode detectors, a parallel-plate ion chamber, and radiographic film. , 2000, Medical physics.

[13]  A. Brahme,et al.  Comparative dosimetry in narrow high-energy photon beams. , 2000, Physics in medicine and biology.

[14]  Determination of output factors for stereotactic radiosurgery beams. , 2009, Medical physics.

[15]  W. Laub,et al.  The volume effect of detectors in the dosimetry of small fields used in IMRT. , 2003, Medical physics.

[16]  Otto A Sauer,et al.  Measurement of output factors for small photon beams. , 2007, Medical physics.

[17]  Juan J. Torres,et al.  Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration , 2008, Journal of applied clinical medical physics.

[18]  A. Sethi,et al.  Comparison of ionization chambers of various volumes for IMRT absolute dose verification. , 2003, Medical physics.

[19]  E. Sham,et al.  Precise radiochromic film dosimetry using a flat-bed document scanner. , 2005, Medical physics.

[20]  Relative output factor measurements of a 5 mm diameter radiosurgical photon beam using polymer gel dosimetry. , 2005, Medical physics.

[21]  Carlo Cavedon,et al.  Erratum: “Total scatter factors of small beams: A multidetector and Monte Carlo study” [Med. Phys.35, 504–513 (2008)] , 2010 .

[22]  M. Torrens,et al.  Gamma knife output factor measurements using VIP polymer gel dosimetry. , 2009, Medical physics.

[23]  John D Fenwick,et al.  Using a Monte Carlo model to predict dosimetric properties of small radiotherapy photon fields. , 2008, Medical physics.

[24]  I. Das,et al.  Small fields: nonequilibrium radiation dosimetry. , 2007, Medical physics.

[25]  Michael Krystek,et al.  Response of the alanine/ESR dosimetry system to MV x-rays relative to 60Co radiation , 2008, Physics in medicine and biology.

[26]  C Ross Schmidtlein,et al.  Precise radiochromic film dosimetry using a flat-bed document scanner. , 2005, Medical physics.

[27]  E. Pappas,et al.  Relative output factor measurements of a 5 mm diameter radiosurgical photon beam using polymer gel dosimetry. , 2005, Medical physics.

[28]  G. Zeng,et al.  An experimental and Monte Carlo investigation of the energy dependence of alanine/EPR dosimetry: I. Clinical x-ray beams , 2004, Physics in medicine and biology.

[29]  Kazuo Hatano,et al.  Reference dosimetry condition and beam quality correction factor for CyberKnife beam. , 2008, Medical physics.

[30]  F. Foppiano,et al.  Response to high-energy photons of PTW31014 PinPoint ion chamber with a central aluminum electrode. , 2008, Medical physics.

[31]  Otto A Sauer,et al.  Determination of the quality index (Q) for photon beams at arbitrary field sizes. , 2009, Medical physics.

[32]  L Paelinck,et al.  The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambers , 2009, Physics in medicine and biology.

[33]  D W O Rogers,et al.  Relationship between %dd(10)x and stopping-power ratios for flattening filter free accelerators: a Monte Carlo study. , 2008, Medical physics.