Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator.

The wear of acetabular cups made from conventional gamma-sterilized, and electron-beam cross-linked ultra-high-molecular-weight polyethylene was studied with a biaxial hip wear simulator. The femoral heads were either polished or roughened so that they represented the type of roughening and the value of surface roughness (R(a) = 0.14-0.18 micro m) observed in the roughest explanted femoral heads. The lubricant was diluted calf serum, and the test length 3 million cycles. The mean wear rate and standard deviation of conventional polyethylene cups against polished and against roughened heads was 11.6 +/- 0.07 and 64.4 +/- 10.1 mg per 1 million cycles, respectively. The latter value closely corresponds to that measured from explanted Charnley prostheses. Against polished heads, cross-linked polyethylene cups showed net weight gain, whereas against roughened heads, their mean wear rate was 2.4 +/- 0.3 mg per 1 million cycles. The mean equivalent circle diameters of polyethylene wear particles produced in the above four categories were: conventional/roughened 0.32 micro m, cross-linked/roughened 0.29 micro m, conventional/polished 0.28 micro m, cross-linked/polished 0.23 micro m. The size ranges and shapes were close to those seen in particles isolated from tissue samples. In conclusion, the tests indicated that electron-beam irradiation effectively reduces the harmful polyethylene wear particle production in total hip arthroplasty.

[1]  A Unsworth,et al.  Wear in Retrieved Charnley Acetabular Sockets , 1996, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[2]  A Unsworth,et al.  The effect of surface topography of retrieved femoral heads on the wear of UHMWPE sockets. , 1997, Medical engineering & physics.

[3]  M. Freeman,et al.  The size and shape of particulate polyethylene wear debris in total joint replacements , 1997, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[4]  J H Dumbleton,et al.  Effect of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene acetabular cups. , 1998, The Journal of arthroplasty.

[5]  V. Saikko,et al.  Type of motion and lubricant in wear simulation of polyethylene acetabular cup , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[6]  D. McNulty,et al.  Analysis of prosthetic femoral heads retrieved at autopsy. , 1999, Clinical orthopaedics and related research.

[7]  F. Shen,et al.  Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls. , 1999, Clinical orthopaedics and related research.

[8]  J. Uvehammer,et al.  Knee joint kinematics, fixation and function related to joint area design in total knee arthroplasty , 2001, Acta orthopaedica Scandinavica. Supplementum.

[9]  V. Saikko,et al.  Effect of counterface roughness on the wear of conventional and crosslinked ultrahigh molecular weight polyethylene studied with a multi-directional motion pin-on-disk device. , 2001, Journal of biomedical materials research.

[10]  V. Saikko,et al.  Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon. , 2001, Biomaterials.

[11]  H. Oonishi,et al.  Gamma-irradiated cross-linked polyethylene in total hip replacements--analysis of retrieved sockets after long-term implantation. , 2001, Journal of biomedical materials research.