Iterative Reconstruction Technique for Reducing Body Radiation Dose at Ct: Feasibility Study Hara Et Al. Ct Iterative Reconstruction Technique Gastrointestinal Imaging Original Research

OBJECTIVE The purpose of this study was to evaluate the image noise, low-contrast resolution, image quality, and spatial resolution of adaptive statistical iterative reconstruction in low-dose body CT. MATERIALS AND METHODS Adaptive statistical iterative reconstruction was used to scan the American College of Radiology phantom at the American College of Radiology reference value and at one-half that value (12.5 mGy). Test objects in low- and high-contrast and uniformity modules were evaluated. Low-dose CT with adaptive statistical iterative reconstruction was then tested on 12 patients (seven men, five women; average age, 67.5 years) who had previously undergone routine-dose CT. Two radiologists blinded to scanning technique evaluated images of the same patients obtained with routine-dose CT and low-dose CT with and without adaptive statistical iterative reconstruction. Image noise, low-contrast resolution, image quality, and spatial resolution were graded on a scale of 1 (best) to 4 (worst). Quantitative noise measurements were made on clinical images. RESULTS In the phantom, low- and high-contrast and uniformity assessments showed no significant difference between routine-dose imaging and low-dose CT with adaptive statistical iterative reconstruction. In patients, low-dose CT with adaptive statistical iterative reconstruction was associated with CT dose index reductions of 32-65% compared with routine imaging and had the least noise both quantitatively and qualitatively (p < 0.05). Low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT had identical results for low-contrast resolution and nearly identical results for overall image quality (grade 2.1-2.2). Spatial resolution was better with routine-dose CT (p = 0.004). CONCLUSION These preliminary results support body CT dose index reductions of 32-65% when adaptive statistical iterative reconstruction is used. Studies with larger statistical samples are needed to confirm these findings.

[1]  G. Hounsfield Computerized transverse axial scanning (tomography): Part I. Description of system. 1973. , 1973, The British journal of radiology.

[2]  Thomas L Toth,et al.  Low-dose CT of the abdomen: evaluation of image improvement with use of noise reduction filters pilot study. , 2003, Radiology.

[3]  O. Linton,et al.  National conference on dose reduction in CT, with an emphasis on pediatric patients. , 2003, AJR. American journal of roentgenology.

[4]  M. Kalra,et al.  Strategies for CT radiation dose optimization. , 2004, Radiology.

[5]  A. Beck,et al.  Feasibility of MDCT Colonography in ultra-low-dose technique in the detection of colorectal lesions: comparison with high-resolution video colonoscopy. , 2004, AJR. American journal of roentgenology.

[6]  Cynthia H McCollough,et al.  The phantom portion of the American College of Radiology (ACR) computed tomography (CT) accreditation program: practical tips, artifact examples, and pitfalls to avoid. , 2004, Medical physics.

[7]  Thomas L Toth,et al.  Radiation from "extra" images acquired with abdominal and/or pelvic CT: effect of automatic tube current modulation. , 2004, Radiology.

[8]  Thomas L Toth,et al.  Detection and characterization of lesions on low-radiation-dose abdominal CT images postprocessed with noise reduction filters. , 2004, Radiology.

[9]  T. Mulkens,et al.  Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. , 2005, Radiology.

[11]  C. McCollough,et al.  CT dose reduction and dose management tools: overview of available options. , 2006, Radiographics : a review publication of the Radiological Society of North America, Inc.

[12]  C. Becker,et al.  Dose reduction and image quality in MDCT colonography using tube current modulation. , 2006, AJR. American journal of roentgenology.

[13]  J. Valentin,et al.  Managing patient dose in multi-detector computed tomography(MDCT). ICRP Publication 102. , 2007, Annals of the ICRP.

[14]  Jean-Baptiste Thibault,et al.  A three-dimensional statistical approach to improved image quality for multislice helical CT. , 2007, Medical physics.

[15]  T. Trabold,et al.  Evaluation of dose exposure in 64-slice CT colonography , 2007, European Radiology.

[16]  Adrian K. Dixon,et al.  Benefits and costs, an eternal balance , 2007 .

[17]  Molly E.W. Thiessen Low-Dose Versus Standard-Dose CT Protocol in Patients With Clinically Suspected Renal Colic , 2007 .

[18]  Ali H. Mokdad,et al.  Epidemiology of obesity in the Western Hemisphere. , 2008, The Journal of clinical endocrinology and metabolism.

[19]  G. Bongartz,et al.  Diagnostic performance of low-dose CT for the detection of urolithiasis: a meta-analysis. , 2008, AJR. American journal of roentgenology.

[20]  Max Wintermark,et al.  Radiation dose reduction strategy for CT protocols: successful implementation in neuroradiology section. , 2008, Radiology.

[21]  Jin Mo Goo,et al.  Radiation dose modulation techniques in the multidetector CT era: from basics to practice. , 2008, Radiographics : a review publication of the Radiological Society of North America, Inc.

[22]  Zhao-qi Zhang,et al.  Coronary CT angiography with low radiation dose , 2010, The International Journal of Cardiovascular Imaging.

[23]  Alvin C. Silva,et al.  Preliminary estimate of triphasic CT enterography performance in hemodynamically stable patients with suspected gastrointestinal bleeding. , 2009, AJR. American journal of roentgenology.

[24]  L. Tanoue Computed Tomography — An Increasing Source of Radiation Exposure , 2009 .

[25]  Joel G Fletcher,et al.  Appropriate patient selection at abdominal dual-energy CT using 80 kV: relationship between patient size, image noise, and image quality. , 2010, Radiology.

[26]  Radiation redux for coronary CT angiography: how low can we go? , 2010, AJR. American journal of roentgenology.

[27]  Noriyuki Tomiyama,et al.  Adaptive statistical iterative reconstruction technique for pulmonary CT: image quality of the cadaveric lung on standard- and reduced-dose CT. , 2010, Academic radiology.

[28]  M Wintermark,et al.  FDA Investigates the Safety of Brain Perfusion CT , 2010, American Journal of Neuroradiology.

[29]  Jiang Hsieh,et al.  Diffuse lung disease: CT of the chest with adaptive statistical iterative reconstruction technique. , 2010, Radiology.

[30]  Alvin C. Silva,et al.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. , 2010, AJR. American journal of roentgenology.

[31]  William Pavlicek,et al.  Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. , 2010, AJR. American journal of roentgenology.

[32]  Jiang Hsieh,et al.  Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. , 2010, Radiology.

[33]  K. Ohtomo,et al.  Adaptive statistical iterative reconstruction for volume-rendered computed tomography portovenography: improvement of image quality , 2010, Japanese Journal of Radiology.

[34]  Joseph J Budovec,et al.  Update on multidetector computed tomography angiography of the abdominal aorta. , 2010, Radiologic clinics of North America.

[35]  A. Hlavacek Imaging of congenital cardiovascular disease: the case for computed tomography. , 2010, Journal of thoracic imaging.

[36]  D. Sahani,et al.  Reducing Abdominal CT Radiation Dose With Adaptive Statistical Iterative Reconstruction Technique , 2010, Investigative radiology.

[37]  Alvin C. Silva,et al.  Reducing the radiation dose for CT colonography using adaptive statistical iterative reconstruction: A pilot study. , 2010, AJR. American journal of roentgenology.

[38]  James K Min,et al.  Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study. , 2010, AJR. American journal of roentgenology.

[39]  G. Raff Radiation dose from coronary CT angiography: five years of progress. , 2010, Journal of cardiovascular computed tomography.

[40]  J. Leipsic,et al.  Cardiac computed tomography technology and dose-reduction strategies. , 2010, Radiologic clinics of North America.

[41]  Ting-Yim Lee,et al.  Impact of new technologies on dose reduction in CT. , 2010, European journal of radiology.

[42]  E. S. Amis,et al.  Radiation exposure from medical imaging in patients with chronic and recurrent conditions. , 2010, Journal of the American College of Radiology : JACR.

[43]  F. Rybicki,et al.  CT Coronary Angiography: 256-Slice and 320-Detector Row Scanners , 2010, Current cardiology reports.

[44]  J. Leipsic,et al.  Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. , 2010, AJR. American journal of roentgenology.

[45]  Toshinori Hirai,et al.  Comparison of standard- and low-tube voltage MDCT angiography in patients with peripheral arterial disease , 2010, European Radiology.

[46]  Giang Nguyen,et al.  A prospective evaluation of dose reduction and image quality in chest CT using adaptive statistical iterative reconstruction. , 2010, AJR. American journal of roentgenology.

[47]  Strategien zur Reduktion der CT-Strahlendosis , 2010, Der Radiologe.

[48]  R. Jain,et al.  Perfusion CT Imaging of Brain Tumors: An Overview , 2011, American Journal of Neuroradiology.

[49]  C. McCollough,et al.  Validation of a lower radiation computed tomography enterography imaging protocol to detect Crohn's disease in the small bowel , 2011, Inflammatory bowel diseases.