Use of an axisymmetric finite element method to compare maxillary bone variables for a loaded implant.

PURPOSE This study determined the effect of various bone models on the stresses and strains generated under occlusal loading of a dental implant. MATERIALS AND METHODS A two-dimensional finite-element model was created for stress analysis. The geometric and elastic properties of a 3.8 x 10-mm Steri-Oss implant embedded in a segment of premaxilla were modeled. Computed tomography scanning of a dried maxilla half was used to determine representative geometry and density of this region. Material properties for bone were varied to simulate the following: all-cancellous bone, cancellous bone with a thin (1.5-mm) crestal isotropic cortical layer, cancellous bone with a thick (3-mm) crestal isotropic cortical layer, and cancellous bone with a thick (3-mm) layer of transversely isotropic (orthotropic) cortical bone. RESULTS Low stresses and high strains surrounded the fixture apex for the all-cancellous bone model. When a layer of cortical bone was added, higher crestal stresses and lower apical strains were observed. The thicker layer of isotropic cortical bone produced stresses at least 50% less than the thinner layer. The assumption of transverse isotropy (orthotropy) increased stresses and strains by approximately 25% compared with isotropic bone. CONCLUSIONS Crestal cortical layer thickness and bone isotropy have a substantial impact on resultant stresses and strains. Clinical assessment of these parameters is recommended.

[1]  T Kitagami,et al.  Analysis of stress on a fixed partial denture with a blade-vent implant abutment. , 1978, The Journal of prosthetic dentistry.

[2]  T Suetsugu,et al.  Two-dimensional FEM analysis of hydroxyapatite implants: diameter effects on stress distribution. , 1990, The Journal of oral implantology.

[3]  W. Hayes,et al.  The compressive behavior of bone as a two-phase porous structure. , 1977, The Journal of bone and joint surgery. American volume.

[4]  L. Borchers,et al.  Three-dimensional Stress Distribution Around a Dental Implant at Different Stages of Interface Development , 1983, Journal of dental research.

[5]  G L Kinzel,et al.  Alternative materials for three endosseous implants. , 1989, The Journal of prosthetic dentistry.

[6]  G. Hellsing On the regulation of interincisor bite force in man. , 1980, Journal of oral rehabilitation.

[7]  D Pipko,et al.  Three-dimensional finite element stress analysis in and around the Screw-Vent implant. , 1991, The International journal of oral & maxillofacial implants.

[8]  G E Carlsson,et al.  Bite force and state of dentition. , 1977, Acta odontologica Scandinavica.

[9]  W C Van Buskirk,et al.  The Elastic Properties of a Human Mandible , 1987, Advances in dental research.

[10]  G L Kinzel,et al.  Finite element analysis of bone-adapted and bone-bonded endosseous implants. , 1989, The Journal of prosthetic dentistry.

[11]  F. Takeshita,et al.  An experimental study on the tissue compatibility of the titanium blade-vent implant coated with HAP-alumina in the semi-functional state. , 1987, The Journal of oral implantology.