Continuous Systolic Blood Pressure Monitoring By the Difference in Electrocardiogram and Pulse Oximetry in Near-Term, Exteriorized Goat Fetuses

Objective: Our purpose was to determine whether systolic arerial pressure, as predicted from the delay time from the electrocardiogram to pulse oximeter waveform, correlates with directly measured systolic blood pressure. Methods: Six Japanese Saanen goats at 120-125 days' gestation were used. Experiments were conducted in acutely operated animals under anesthesia. We placed catheters in the caroti artery and jugular vein, electrodes on the chest, and a pulse oximeter probe on the tongue. Systolic blood pressure (Psm), fetal heart rate (FHR), delay time (Td), and pulse oximeter waveform were monitored continuously, and the data were stored. Systolic blood pressure was estimated (Pse) by the simlpe linear equation of Td. Fetal blood pressure was decreased 10% by hydralazine and blood withdrawal. Correlation coefficients and the agreement between Psm and Pse were determined. Results: A significant correlation (P < .01) with good bias values (< 8 mmHg) and precision vlaues (< 5 mmHg) was obtanied between Psm and Pse in all animals. There was also a significant correlation between Psam and Pse with high coefficients for hydralazine (r2 = 0.82 ± 0.12) and for withdrawal (r2 = 0.85 ± 0.15). The correlations became slightly scattered in acidemia (pH < 7.25). Conclusion: This technique has the potential to provide information on fetal blood pressure during labor and delivery on a beat-by-beat basis, which will aid in the interpretation of electronic FHR recordings.

[1]  C. Martin,et al.  Cardiac systolic time intervals in fetal monkeys: pre-ejection period. , 1978, American journal of obstetrics and gynecology.

[2]  W Lu,et al.  Research on the main elements influencing blood pressure measurement by pulse wave velocity. , 1992, Frontiers of medical and biological engineering : the international journal of the Japan Society of Medical Electronics and Biological Engineering.

[3]  Kirk H. Shelley,et al.  How Does the Plethysmogram Derived from the Pulse Oximeter Relate to Arterial Blood Pressure in Coronary Artery Bypass Graft Patients? , 2001, Anesthesia and analgesia.

[4]  I. Chalmers,et al.  The Dublin randomized controlled trial of intrapartum fetal heart rate monitoring. , 1985, American journal of obstetrics and gynecology.

[5]  T. Ikeda,et al.  Sequence of neuronal responses assessed by immunohistochemistry in the newborn rat brain after hypoxia-ischemia. , 1997, American journal of obstetrics and gynecology.

[6]  J. Parer,et al.  Physiologic and histologic changes in near-term fetal lambs exposed to asphyxia by partial umbilical cord occlusion. , 1998, American journal of obstetrics and gynecology.

[7]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[8]  L W Organ,et al.  The pre-ejection period as an antepartum indicator of fetal well-being. , 1980, American journal of obstetrics and gynecology.

[9]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[10]  B. Pannier,et al.  Assessment of arterial distensibility by automatic pulse wave velocity measurement. Validation and clinical application studies. , 1995, Hypertension.

[11]  D B Swedlow,et al.  A multicenter controlled trial of fetal pulse oximetry in the intrapartum management of nonreassuring fetal heart rate patterns. , 2000, American journal of obstetrics and gynecology.

[12]  C. Cetrulo,et al.  The evaluation of continuous fetal heart rate monitoring in high-risk pregnancy. , 1976, American journal of obstetrics and gynecology.

[13]  A Steptoe,et al.  Pulse wave velocity and blood pressure change: calibration and applications. , 1976, Psychophysiology.

[14]  D Shapiro,et al.  Pulse transit time and blood pressure: an intensive analysis. , 1983, Psychophysiology.

[15]  L A Geddes,et al.  Pulse transit time as an indicator of arterial blood pressure. , 1981, Psychophysiology.

[16]  A. Steptoe,et al.  Pulse wave velocity as a measure of blood pressure change. , 1976, Psychophysiology.

[17]  A. Lisbon,et al.  An evaluation of a new continuous blood pressure monitoring system in critically ill patients. , 2000, Journal of clinical anesthesia.