Automated on-line L-edge measurement of SNM concentration for near-real-time accounting
暂无分享,去创建一个
The L-edge densitometer developed at Los Alamos National Laboratory has been modified, tested, and demonstrated for on-line assay of special nuclear material concentration in flowing solution streams. The demonstration was part of a larger demonstration of near-real-time nuclear materials accounting during a continuous, week-long, cold operation of the Allied General Nuclear Services facility in Barnwell, South Carolina. The L-edge data were automatically analyzed and the results were transmitted to the materials accounting computer once every 5.5 min for the duration of the cold run. This report compares the results of the L-edge analyses with the delayed results obtained from destructive analysis of samples withdrawn from the same process line. Comparisons are also made with the results obtained in near real time from an automated process control instrument installed in series with the L-edge densitometer. The performance of the L-edge instrument was reliable throughout the continous operation. The assay precision was consistent with that predicted by the counting statistics of the measurement. The results of the L-edge assays show good agreement with those of the destructive assays. A gradually varying discrepancy (of a few percent) between the L-edge and the process control results remains unexplained. 9 figures.